Instruction Manual - PDA-8000

Technical Reference Summary

Covered Models & Document Scope

PDA-8000 optical emission spectrometer (PDA photoelectric-photometric type) for simultaneous multi-element analysis of metals (iron/steel, aluminum, copper, etc.). The manual covers the main body, data processing units (PC, monitor, printer, computer rack; may not be included depending on specification), standard accessories, and optional items such as the stand cooling kit (211-78899-92) and oil mist separator. No other model variants or ordering codes are described.

Section-by-Section Technical Breakdown

pp. 4-6
System Components and Emission Table

Describes the front and rear system configuration: spectrometer main body, excitation unit, read-out unit, emission table, argon gas control unit, vacuum pump, switch panel, and connectors. Details the emission table with sample holder, interlocked cover, cover lock lamp, start/stop switches, electrode, dust cup, and maintenance door.

  • Emission table cover has an interlock switch; spark discharge is disabled when the cover is open.
  • Cover is locked during optical emission; cover lock lamp lights during emission.
  • Do not disable the cover switch during maintenance or inspection due to electric shock risk.
  • Rear panel: AC 200/220/230/240 V, 12.5 A, 50/60 Hz power input, LAN, SOURCE and MAIN switches, argon gas connector, vacuum pump connector, argon gas bubbler.
pp. 7-8
Checking the Argon Gas Flow

Procedure for checking argon gas flow using the flow meter on the right side of the spectrometer. Specifies flow rates for Sleep, Standby, and Analysis modes and the required argon supply pressure range.

  • Flow rates: Sleep 0.1 L/min, Standby 0.5 L/min, Analysis 5 L/min (differs depending on analyzed sample).
  • Argon gas supply pressure must be 140 kPa to 200 kPa and stable; fluctuation affects analysis values.
  • Read flow at the center of the ball.
  • Check residual argon cylinder pressure daily; replace cylinder if 2 MPa or less.
pp. 14-18
Startup

Startup procedure after long-term shutdown. Requires 5 hours from power-on until the spectrometer is stable enough for use. Includes checks of room temperature, argon pressure, dummy sample, power-on sequence, software login, argon bubbler water level, waste discharge, recalibration, master curve correction, and check analysis.

  • Room temperature must be 15 °C to 30 °C.
  • Argon supply pressure must be 140 kPa to 200 kPa.
  • Wait 5 hours after turning ON MAIN before analysis.
  • Argon gas bubbler water level must be 8 cm to 10 cm; install bubbler at height of 60 cm or less from floor.
  • Turn ON MAIN, then PC and peripherals, log in to Windows, start PDA-R software with password.
  • After startup: waste discharge, recalibration, master curve correction as necessary, check analysis.
pp. 19-21
Shutdown

Procedure for long-term shutdown of the system. Covers setting instrument status to OFF, shutting down software and Windows, turning OFF SOURCE then MAIN, cleaning the emission table, setting a dummy sample, and removing the argon bubbler cap.

  • Set [Instrument Status] window to OFF.
  • Shut down software and Windows, turn OFF PC and peripherals.
  • Turn OFF excitation unit power switch (SOURCE) first, then main power switch (MAIN).
  • Clean emission table and set a dummy sample on the sample plate.
  • Remove cap from argon gas bubbler and remove tip of argon emission pipe from water.
pp. 114-121
Troubleshooting

Provides a table of troubles, causes, and measures for common problems such as spectrometer inner pressure out of green range, unstable temperature, poor analysis accuracy, and temperature buzzer. Also covers error messages during analysis with causes and corrective actions.

  • Spectrometer inner pressure not in green range: check condenser lens O-ring, vacuum pump oil level and condition.
  • Temperature not stable: confirm analysis chamber temperature 15–30 °C, shut down and restart.
  • Poor analysis accuracy: check argon leaks and electrode contamination; clean emission table and sample plate.
  • Buzzer sounds when spectrometer temperature exceeds 55 °C: turn OFF MAIN, check room temperature, restart after cooling.
  • Error messages during analysis include Photo Detector HV OFF, communication loss, control CPU stopped, and incorrect analysis condition settings.
pp. 124-168
Maintenance and Inspection

Lists consumables, maintenance parts, and spare parts, and describes maintenance and inspection items, flow, daily maintenance, periodic maintenance, and preparation for analysis after maintenance. Includes procedures for cleaning the emission table, condenser lens, replacing pump oil, and checking argon bubbler water level.

  • Daily maintenance includes cleaning emission table and checking argon bubbler water level.
  • Periodic maintenance includes cleaning condenser lens and replacing pump oil.
  • Replacement cycles for periodic replacement parts are rough estimates; may require earlier replacement depending on usage.
  • After maintenance, perform waste discharge and preparation for analysis.
pp. 180-186
Technical Data

Contains specifications and installation requirements for the PDA-8000. Includes power requirements, environmental conditions, and installation category.

  • Power supply: AC 200/220/230/240 V, 1.5 kVA maximum, 50/60 Hz.
  • Installation category II, pollution degree 2, altitude up to 2000 m, indoor use only.
  • Operating temperature 15 °C to 30 °C, humidity 15% to 70% RH.
  • Grounding resistance 30 Ω or below (10 Ω or below advisable, Class A grounding).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Power supply: AC 200/220/230/240 V, 1.5 kVA maximum, 50/60 Hz. p.6 ↗ Ensure power distribution and breaker capacity can handle 1.5 kVA; provide a leakage breaker for the power distribution box. PDA-8000
Grounding resistance must be 30 Ω or below (10 Ω or below advisable, Class A grounding); individual grounding required. p.6 ↗ Shared grounding may cause interference; install a dedicated ground with low resistance. PDA-8000
Operating environment: indoor, temperature 15 °C to 30 °C, humidity 15% to 70% RH, no direct wind from air-conditioners, no large vibration. p.7 ↗ Install in a temperature-controlled room free from drafts and vibration to ensure stable analysis. PDA-8000
Installation category II, pollution degree 2, altitude up to 2000 m, indoor use only. p.7 ↗ Verify site electrical installation meets these overvoltage and pollution requirements. PDA-8000
Argon gas supply pressure must be 140 kPa to 200 kPa and stable; fluctuation affects analysis values. p.7 ↗ Use a two-stage regulator and monitor pressure daily; replace cylinder when residual pressure is 2 MPa or less. PDA-8000
Argon gas flow rates: Sleep 0.1 L/min, Standby 0.5 L/min, Analysis 5 L/min (differs depending on analyzed sample). p.7 ↗ Set flow meter knobs accordingly and verify at the center of the ball. PDA-8000
Argon gas bubbler water level must be 8 cm to 10 cm; install bubbler at height of 60 cm or less from floor. p.17 ↗ Incorrect water level causes fluctuation in analysis values; check daily. PDA-8000
After long-term shutdown, 5 hours are required from startup before the system is stable enough for use. p.14 ↗ Plan startup 5 hours before analysis; wait for spectrometer temperature to stabilize. PDA-8000
When spectrometer temperature exceeds 55 °C, the temperature alarm buzzer sounds. p.10 ↗ Turn OFF MAIN, check room temperature (must be 15–30 °C), and restart after cooling. PDA-8000
Do not place anything heavier than 5 kg on the sample plate. p.8 ↗ Heavier objects may deform the sample plate and affect analysis accuracy. PDA-8000
Do not attempt to analyze strongly magnetized samples. p.8 ↗ Strong magnetism prevents correct analysis results. PDA-8000
Optional stand cooling kit (211-78899-92) is required for continuous analysis at high-energy discharge. p.9 ↗ For continuous high-energy analysis, order and install the cooling kit to prevent emission table overheating. PDA-8000
Emission table cover interlock disables spark discharge when cover is open; cover is locked during emission. p.6 ↗ Do not disable the cover switch; forced opening breaks the switch and poses electric shock risk. PDA-8000
Power failure procedure: turn OFF SOURCE, then MAIN; after recovery, turn ON MAIN, then SOURCE, then vacuum pump on [Instrument Status] window. p.10 ↗ Follow this sequence to avoid equipment damage and ensure proper restart. PDA-8000
During power interruption, door lock is released to allow sample removal; analysis data is not saved. p.10 ↗ Expect data loss on power failure; reanalyze sample after recovery. PDA-8000
Replacement parts are available for seven years after product discontinuation. p.16 ↗ Plan long-term maintenance and spare parts procurement accordingly. PDA-8000

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
The spectrometer inner pressure is not within the green range. p.115 ↗ The O-ring of the condenser lens is not correctly installed; the oil level in the vacuum pump is not the specified one; the pump oil in the vacuum pump is soiled, lowering the exhaust ability. Perform maintenance and inspection of the condenser lens and vacuum pump. See Chapter 7, '4.3 Cleaning the Condenser Lens' (page 146) and '4.5 Replacing the Pump Oil' (page 151). Chapter 7, 4.3 and 4.5
The spectrometer temperature is not stable or does not reach the setting temperature. p.115 ↗ Abnormal temperature of the spectrometer; the fan supplying hot air has a problem. Confirm that the temperature of the analysis chamber is 15 °C to 30 °C. After that, shut down the system, and then restart it. See Chapter 2, '2. Shutdown' (page 19) and '1. Startup' (page 14). Chapter 2, 1 and 2
Poor analysis accuracy p.115 ↗ The argon gas leaks during analysis; the electrode is contaminated. Set the sample correctly and perform reanalysis. If the analysis accuracy cannot be restored, clean the emission table and sample plate. See Chapter 7, '4.1 Cleaning the Emission Table' (page 129). Chapter 7, 4.1
The buzzer of the temperature display sounds. When the spectrometer temperature exceeds 55 °C. p.115 ↗ Abnormal temperature of the spectrometer; a temperature control error message is displayed on the [Instrument Status] window. Confirm that the temperature of the analysis chamber is 15 °C to 30 °C. After that, shut down the system, and after confirming that the displayed temperature of the spectrometer has lowered to room temperature, restart the system. See Chapter 2, '2. Shutdown' (page 19) and '1. Startup' (page 14). Chapter 2, 1 and 2
An Error Occurs before Starting Analysis Condition Setup. p.121 ↗ An error has occurred before analysis processing. Check the conditions of the entire instrument on the [Instrument Status] window. 3.2 Errors During Analysis
Photo Detector HV is OFF. p.121 ↗ This is displayed when the analysis started with the photomultiplier HV [OFF]. Click ON [Photo Detector HV] in the [Instrument Control] area of the [Instrument Status]. 3.2 Errors During Analysis
An Error Occurs after Starting Analysis Condition Setup. p.121 ↗ There is a possibility that communication with the instrument is lost, or the control CPU of the instrument is stopped. Check if [Control CPU] is normal in the [Instrument Status] area of the [Instrument Status] window. 3.2 Errors During Analysis
An Error Occurs during Analysis Mode Setup. p.121 ↗ There is a possibility that communication with the instrument is lost, or the control CPU of the instrument is stopped. Check if [Control CPU] is normal in the [Instrument Status] area of the [Instrument Status] window. 3.2 Errors During Analysis
An Error Occurs during Integration Mode Setup. p.121 ↗ Setting of the integration mode is incorrect. Check [Analytical Condition] - [Measurement] on the [Analytical Group Information] window. 3.2 Errors During Analysis
An Error Occurs during Sampling Condition Setup. p.121 ↗ Setting of the analysis sampling is incorrect. Check [Analytical Condition] - [Discharge Condition] on the [Analytical Group Information] window. 3.2 Errors During Analysis
An Error Occurs during Sequence Condition Check. p.121 ↗ The SEQ setting is incorrect. Check [Analytical Condition] - [Discharge Condition] on the [Analytical Group Information] window. 3.2 Errors During Analysis
An Error Occurs during Integration Condition Setup. p.121 ↗ Setting of the measurement conditions is incorrect. Check [Analytical Condition] - [Measurement] on the [Analytical Group Information] window. 3.2 Errors During Analysis
An Error Occurs during Analysis Cycle Setup. p.121 ↗ Setting of the analysis cycle is incorrect. Check [Analytical Condition] - [Discharge Condition] on the [Analytical Group Information] window. 3.2 Errors During Analysis
An Error Occurs during ATT Setup. p.121 ↗ Setting of the photomultiplier sensitivity is incorrect. Check [Analytical Condition] - [Attenuator] on the [Analytical Group Information] window. 3.2 Errors During Analysis