Instruction Manual - SMV-301/301RT
Technical Reference Summary
Covered Models & Document Scope
This manual covers the Shimadzu automatic Mooney viscometers SMV-301 (standard model) and SMV-301RT (functionality-enhanced model with stress relaxation measurement and variable rotor speed). EMC descriptions apply specifically to 344-04262-58 SMV-301 50Hz, 344-04263-58 SMV-301 60Hz, and 344-04264-58 SMV-301RT.
Section-by-Section Technical Breakdown
The SMV-301/301RT are automatic Mooney viscometers for determining Mooney viscosity, scorch time, minimum Mooney viscosity and related characteristics of unvulcanized rubber per JIS, ISO and ASTM methods. The instrument uses an LCD touch-panel for test condition setup and result display, and requires both hands on the down switches to start a test for safety.
- SMV-301 is the standard model; SMV-301RT adds stress relaxation measurement and variable rotor speed.
- Measurement conforms to JIS, ISO, ASTM and other standards.
- LCD touch-panel display for setting test conditions and displaying results.
- Two-hand operation required to start a test.
Lists the standard accessories supplied with one SMV-301 or SMV-301RT unit, including dies, rotor, seals, calibration weights, temperature sensors, power cord, piping hose, tool set and touch pen. Optional accessories, maintenance parts and consumables are not included.
- V-groove die: 1 set for upper and lower.
- L type rotor: 1 piece.
- O-ring set (10 pieces) and backup ring set (5 pieces).
- Standard weights for 20M (1 set) and for 40M (2 sets).
- Temperature sensor: 2 sets.
- Power cord 3 m, piping hose 3 m, tool set, touch pen.
Describes the test chamber, pressurizing unit, Mooney viscosity detector, drive unit, automatic rotor lifter, standard weight calibration unit, power switch, control panel, heating unit, rotor preheating chamber and windbreaker casing. Key operating pressures and rotor speeds are specified.
- Upper die pressure is 11500±500 N, proportional to pneumatic pressure; pneumatic pressure should be 0.48 to 0.50 MPa.
- Dies are attached with four M6 x 20 bolts; O-ring and backup ring at lower die center prevent specimen outflow.
- SMV-301 rotor speed is 2 rpm; SMV-301RT is variable from 0.1 to 20.0 rpm.
- Calibration with standard weights should be done approximately once per month.
- Power switch includes an earth leakage detector circuit.
- Windbreaker casing door interlock stops heating plate movement when opened.
Describes the standby switch and LCD panel of the measurement control unit. The standby switch lights when pressed, and closing the windbreaker door starts a test. The LCD panel is a touch screen for settings and data processing.
- Standby switch turns on and lights; closing the windbreaker door then starts a test.
- LCD panel displays test conditions and results; use the stylus or touch softly.
- Do not press the LCD with pointed objects or with excessive force.
- Always-lit or always-dark picture elements are an LCD specification, not a malfunction.
Specifies installation site conditions, power requirement and air source requirement. Installation, adjustment and re-installation must be performed by a Shimadzu representative.
- Operating temperature +10 °C to +40 °C; no rapid temperature fluctuations.
- Vibration acceleration 0.005 G or less.
- Supply voltage fluctuation within ±10%.
- Ground resistance 100 Ω max.
- Power: single-phase AC100 V ±10%, 1500 VA, 50/60 Hz.
- Air supply pressure 0.50 to 0.70 MPa; equipment operating pressure 0.48 to 0.50 MPa.
- Air source must be dry air; connection via R 1/4 female screw coupler.
Describes preparation before power-on: opening the stop valve, checking manometer reading of 0.48 to 0.50 MPa, and closing the stop valve before turning power off. Power-on displays the title screen and starts self-check.
- Turn stop valve counterclockwise to open; check manometer reads 0.48 to 0.50 MPa.
- Adjust air pressure per section 10.3 if reading is out of range.
- Before turning OFF the power switch, turn the stop valve clockwise to close it.
- Power switch is on the right side of the testing machine; self-check starts after title screen.
Covers general conditions (date/time, warm-up completion, buzzer, Mooney coefficients, rotor size, decay time and stress relaxation for SMV-301RT) and device setup (printer, print items, external control, hardware ID).
- Date/time backed up by rechargeable battery; may need resetting after installation or long idle period.
- Automatic warm-up can be scheduled up to one month in advance.
- Rotor size selection L or S; S type is optional and generally gives smaller Mooney values.
- Decay time setup for SMV-301RT: X 0 to 100%, t 0 to 600 s; zero disables that measurement.
- Stress relaxation test per ISO 289-4 and ASTM D1646 supported on SMV-301RT.
- Printer types: inkjet supporting ESC/P, or thermal CT-S4000.
- External control via USB requires optional PC software; hardware ID is a 4-digit number.
Describes setting test conditions such as temperature, Mooney viscosity measurement time, preheat time, after time (pressure holding time), scorch points, sample number, remark and graph scale.
- Test conditions include temperature, preheat time, after time, scorch points and Mooney viscosity measurement time.
- Condition files can be saved and loaded.
- Statistical data processing is available.
Covers daily maintenance, greasing, and replacement of consumable items. Periodic inspection and calibration should be requested from a Shimadzu representative.
- Daily maintenance per section 9.1.
- Greasing per section 9.2.
- Replacing consumable items per section 9.3.
- Periodic inspection/calibration by Shimadzu representative.
- Replacement cycles are rough estimates and may be shorter depending on environment and usage.
Provides symptom-based troubleshooting, error messages, corrective actions, air pressure adjustment and diagnosis by maintenance programs. If the problem is not solved, contact a Shimadzu representative.
- Symptoms include blank or abnormal LCD, upper platen not lifting/lowering, temperature not rising or abnormally high, abnormal Mooney values, smoke or abnormal smell.
- Error messages include Printer Error, RAM X Error and A/D Data Err.
- Corrective actions in section 10.2.
- Air pressure adjustment in section 10.3.
- Diagnosis by maintenance programs in section 10.4.
Lists consumables and maintenance parts. Replacement parts are available for seven years after product discontinuation, subject to manufacturer conditions.
- Consumables listed in section 11.1.
- Maintenance parts listed in section 11.2.
- Replacement parts available for seven years after discontinuation.
Covers principles of measurement, examples of tests and standard specifications.
- Principles of measurement in section 12.1.
- Examples of tests in section 12.2.
- Standard specifications in section 12.3.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Upper die pressure is 11500±500 N, proportional to pneumatic pressure; pneumatic pressure should be 0.48 to 0.50 MPa. p.33 ↗ | Maintain air pressure within 0.48–0.50 MPa to achieve the specified die pressure; adjust via pressure regulation valve while monitoring the pressure gage. | SMV-301 and SMV-301RT |
| SMV-301 rotor speed is 2 rpm; SMV-301RT rotor speed is variable from 0.1 to 20.0 rpm. p.34 ↗ | Select model based on required rotor speed range; RT model needed for variable-speed or stress relaxation work. | SMV-301 / SMV-301RT |
| Calibration using standard weights should be done at an interval of approximately one month. p.34 ↗ | Schedule monthly sensitivity calibration to maintain Mooney viscosity measurement accuracy. | SMV-301 and SMV-301RT |
| Power requirement: single-phase AC100 V ±10%, 1500 VA, 50/60 Hz; ground resistance 100 Ω max. p.42 ↗ | Verify supply voltage stability and install a protective ground with ≤100 Ω resistance before operation. | SMV-301 and SMV-301RT |
| Air supply pressure 0.50 to 0.70 MPa; equipment operating pressure 0.48 to 0.50 MPa; air source must be dry air. p.42 ↗ | Provide dry compressed air within supply range and regulate to 0.48–0.50 MPa at the instrument; moisture may cause malfunction. | SMV-301 and SMV-301RT |
| Installation site: operating temperature +10 °C to +40 °C, vibration acceleration 0.005 G or less, supply voltage fluctuation within ±10%, no condensation, no corrosive gas. p.41 ↗ | Site the instrument away from vibration, direct sunlight, temperature fluctuations and corrosive atmospheres to ensure stable operation. | SMV-301 and SMV-301RT |
| When turning power off and on again, wait at least 5 seconds before turning it on. p.8 ↗ | Avoid rapid power cycling to prevent equipment failure. | SMV-301 and SMV-301RT |
| Decay time setup for SMV-301RT: X (%) 0 to 100%, t (s) 0 to 600 s; measurements based on a parameter set to zero are not performed. p.54 ↗ | Configure stress relaxation parameters within these ranges; zero disables the corresponding measurement. | SMV-301RT |
| Stress relaxation test supported per ISO 289-4 and ASTM D1646 on SMV-301RT; ISO 289-4 calculates MSR and standard error, ASTM D1646 calculates intercept k, slope a, correlation r, area A, or decay time. p.55 ↗ | Use SMV-301RT for standards-compliant stress relaxation analysis; select the appropriate standard in the software. | SMV-301RT |
| Rotor size selection L or S; S type is optional and Mooney viscosity values measured with S type are generally smaller than with L type. p.53 ↗ | Do not directly compare results between L and S rotors; the configured size is only for output labeling, not conversion. | SMV-301 and SMV-301RT |
| Printer types supported: inkjet printer supporting ESC/P commands, or designated thermal printer CT-S4000. p.57 ↗ | Select a compatible printer when configuring automatic printout. | SMV-301 and SMV-301RT |
| External control via USB requires optional PC software; hardware ID is a 4-digit number set in the PC software. p.59 ↗ | For PC-controlled operation, install the optional SMV-301 PC Control Software and configure matching 4-digit hardware ID. | SMV-301 and SMV-301RT |
| Windbreaker casing door interlock: when door is opened while heating plate is moving, the heating plate stops immediately; during testing the heating plate position is kept and test continues; at end of test the heating plate does not move up. p.36 ↗ | Understand interlock behavior for safe operation and to avoid unexpected test interruption. | SMV-301 and SMV-301RT |
| If the rotating part protection cover is removed, the cylinder does not move and the motor does not rotate. p.13 ↗ | Confirm the rotating part protection cover is correctly fixed before turning on power. | SMV-301 and SMV-301RT |
| Before maintenance inside the viscometer: press down switch to close test space, open front door, turn stop valve knob clockwise to tighten, release residual air from drain valve at filter bottom, then turn power OFF. p.13 ↗ | Follow this sequence to prevent upper platen drop and ensure safe maintenance. | SMV-301 and SMV-301RT |
| Replacement parts are available for seven years after product discontinuation, subject to manufacturer conditions. p.20 ↗ | Plan long-term spare parts procurement within this availability window. | SMV-301 and SMV-301RT |
| EMC compliance: EN61326:1997 + A1:1998 + A2:2001 + A3:2003, class A emissions and minimum immunity requirement, for models 344-04262-58, 344-04263-58, 344-04264-58. p.24 ↗ | Class A product not intended for residential use; maintain distance from sensitive devices and use a separate power source from strong EMI noise sources. | 344-04262-58 SMV-301 50Hz, 344-04263-58 SMV-301 60Hz, 344-04264-58 SMV-301RT |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| After turning the power ON, nothing appears on the LCD panel. p.112 ↗ | Not stated in the manual. | Contact Shimadzu representative. | Refer to 10.2.1, 10.2.2, 10.2.3 |
| the display is abnormal. p.112 ↗ | Not stated in the manual. | Contact Shimadzu representative. | Refer to 10.2.1, 10.2.2, 10.2.3 |
| Abnormality in the display of screen pages or displayed characters: the display is stuck at a test condition setup screen. p.112 ↗ | Not stated in the manual. | Contact Shimadzu representative. | Refer to 10.2.1, 10.2.2, 10.2.3 |
| incomplete or missing display p.112 ↗ | Not stated in the manual. | Contact Shimadzu representative. | Refer to 10.2.1, 10.2.2, 10.2.3 |
| the display is stuck at a testing screen p.112 ↗ | Not stated in the manual. | Contact Shimadzu representative. | Refer to 10.2.1, 10.2.2, 10.2.3 |
| The upper platen... does not lift. p.112 ↗ | Not stated in the manual. | Refer to 10.2.4. | 10.2.4 |
| does not lower. p.112 ↗ | Not stated in the manual. | Refer to 10.2.5. | 10.2.5 |
| The temperature of the upper or lower platen... does not rise. p.112 ↗ | Not stated in the manual. | Refer to 10.2.6. | 10.2.6 |
| rises up abnormally high. p.112 ↗ | Not stated in the manual. | Refer to 10.2.7. | 10.2.7 |
| A measured Mooney viscosity value is... abnormally high. p.112 ↗ | Not stated in the manual. | Refer to 11.2.8. | 11.2.8 |
| abnormally low. p.112 ↗ | Not stated in the manual. | Refer to 11.2.9. | 11.2.9 |
| fluctuating periodically. p.112 ↗ | Not stated in the manual. | Refer to 11.2.10. | 11.2.10 |
| Smoke comes out of the viscometer. p.112 ↗ | Not stated in the manual. | Turn off the power immediately and contact Shimadzu representative. | Refer to 10.2.11, 10.2.12 |
| Abnormal smell comes out of the viscometer. p.112 ↗ | Not stated in the manual. | Turn off the power immediately and contact Shimadzu representative. | Refer to 10.2.11, 10.2.12 |
| The clock displays erroneous time. p.112 ↗ | Not stated in the manual. | Refer to 11.2.13. | 11.2.13 |
| Printer Error p.112 ↗ | Printer power OFF; Printer cable disconnected; Printer out of paper; Paper feed error. | Not stated in the manual. | Error messages table |
| RAM X Error p.112 ↗ | RAM IC failure (X=0 M6, X=1 M6, X=2 M6, X=3 M6). | Ask for repair. | Error messages table |
| A/D Data Err. p.112 ↗ | A/D converter board failure. | Ask for repair. | Error messages table |