Instruction Manual - USF-2000/USF-2000A - Software

Technical Reference Summary

Covered Models & Document Scope

This manual (Part No. 349-04408) covers the USF-2000/USF-2000A ultrasonic fatigue testing machine software, supporting both longitudinal (vertical) vibration and torsional vibration test machines. It describes software operation, test condition setup, execution modes, data file layout, hardware DIP switch settings for ultrasonic power supplies (2000BCD and DCX20: 2.5), interface board settings, software installation, and break detection function setup.

Section-by-Section Technical Breakdown

pp. 13-14
Software Overview

The software performs 20 kHz resonance fatigue tests for both longitudinal and torsional vibration machines. It supports normal (constant stress) and stress fluctuation tests, as well as continuous and intermittent operation. Five execution modes are provided: test condition setting, test execution, system settings, check frequency, and frequency adjustment.

  • Fatigue test at 20 kHz using resonance.
  • Selectable longitudinal vibration or torsion test depending on machine.
  • Normal test applies constant stress; stress fluctuation test changes stress per cycle.
  • Continuous operation applies constant stress; intermittent operation repeats oscillation/stop to suppress specimen heating.
  • Five modes: test condition setting, test execution, system settings, check frequency, frequency adjustment.
pp. 15-18
Menu Processing

The File menu provides reading, saving, and printing of test conditions, saving of collected data, and exit. The System menu provides system settings, frequency check, frequency adjustment, and options. The Help menu shows version information.

  • File menu: read test conditions, save test conditions, save as, save data, print test conditions, exit.
  • System menu: System Settings, frequency check, frequency adjustment, option.
  • Options screen has Language tab (Japanese/English) and Test type tab (longitudinal/torsional); restart required after change.
  • Help menu contains only version information.
pp. 20-28
Screen Details and Functionality (Vertical Vibration Test) - Test Conditions

This section describes setting test piece material, stress conversion factor, amplifier output correction factor, test piece shape, and test conditions for normal and stress fluctuation tests. It includes input ranges for material properties, specimen geometry, stress, oscillation/stop times, and test end criteria.

  • Material: Young's modulus 1~10^6 MPa, density 0.01~1000000 g/cm3.
  • Stress conversion factor can be input or calculated from specimen shape and material.
  • Amplifier output correction factor range 0.800 to 1.200.
  • Specimen shapes: circular tapered (chord or arc), notch, straight.
  • Normal test: nominal stress, intermittent operation, oscillation time 110~500 ms, stop time 100~5000 ms.
  • Stress fluctuation test: up to 10 steps, loop count 1~100, stress, oscillation 110~500 ms, stop 0 or 100~5000 ms, total cycle 20000~1×10^12.
  • Test end: counter preset 20000~1×10^12 times, frequency fluctuation width 10~1000 Hz.
pp. 29-30
Correction Magnification Adjustment

This section describes adjusting the amplifier output correction magnification by measuring the actual edge amplitude and calculating a new correction factor. It includes oscillation settings and a graph of amplifier output vs. stress amplitude.

  • Target edge amplitude displayed from nominal stress; measured edge amplitude input range 0.80 to 600.00 μm.
  • New correction magnification range 0.800 to 1.200.
  • Oscillation time 110~500 ms, stop time 100~5000 ms.
  • Graph shows amplifier output (%) vs. stress amplitude and edge amplitude.
pp. 31-36
Run Test

This section describes the test execution screen, starting and stopping tests, data collection settings, and test end conditions. It includes counter, frequency, stress amplitude displays, and data collection cycle settings.

  • Test execution screen shows step, counter, frequency, stress amplitude, edge amplitude.
  • Data display maximum 32000 points; data collection up to 1.0×10^10 times.
  • Data collection cycle: logarithmic or linear; initial cycle 20000~1×10^11, end cycle up to 1×10^12, interval 20000~1×10^11.
  • Test ends on cycle over, frequency fluctuation range exceeded, alarm, emergency stop, break detection, or manual stop.
  • External trigger signal is HIGH when test starts.
pp. 36-37
System Setting

This section describes calibration of the horn end face amplitude by measuring the horn end face and setting values. It includes selection of ultrasonic power supply type, horn tip diameter, intermittent operation settings, and amplifier amplitude calibration.

  • Calibration value shipped with straight test piece calibration.
  • Ultrasonic power supply type selection changes input/output voltage range.
  • Horn tip diameter setting.
  • Amplifier output amplitude 10-120%, horn side amplitude 1~120 μm.
  • Rise time setting for intermittent operation.
pp. 29-46
Screen Details and Functionality (Torsion Test) - Test Conditions

This section mirrors the vertical vibration test conditions but for torsional vibration tests. It covers material, stress conversion factor, amplifier output correction factor, specimen shape, normal and stress fluctuation test conditions, and test end settings.

  • Similar input ranges as vertical vibration test.
  • Adjust correction factor mode available.
  • Run test, system setting, check frequency, frequency adjustment for torsion.
pp. 48-51
File Layout

This section describes the layout of collected data files for longitudinal vibration and torsion tests, for both normal and stress fluctuation tests.

  • Data file layout for normal test (longitudinal vibration).
  • Data file layout for stress fluctuation test (longitudinal vibration).
  • Data file layout for normal test (torsional).
  • Data file layout for stress fluctuation test (torsional).
pp. 52-54
Set of Hardware

This section describes hardware settings: internal DIP switch of ultrasonic power supply (2000BCD), ultrasonic power supply (DCX20: 2.5), and interface board settings (old type).

  • Ultrasonic power supply (2000BCD) internal DIP switch settings.
  • Ultrasonic power supply (DCX20: 2.5) settings.
  • Interface board settings for old type boards.
pp. 56-61
Installation of the Software

This section describes turning off the power-saving mode of the personal computer, recognition of various boards, installation of InterfaceAD board driver (old type), new installation of the ultrasonic fatigue testing machine software, and version upgrade.

  • Turn off power-saving mode of PC.
  • Recognize various boards.
  • Install InterfaceAD board driver if old type board.
  • New installation and version up of software.
pp. 64-69
Break Detection Function

This section describes setting up the break detection function using WAVE LOGGER PRO and the ultrasonic fatigue test software.

  • Set of WAVE LOGGER PRO.
  • Set of ultrasonic fatigue test software for break detection.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Fatigue test frequency is 20 kHz using resonance. p.13 ↗ The system is designed for very high cycle fatigue testing at ultrasonic frequency. USF-2000/USF-2000A
Young's modulus input range: 1 ~ 10^6 MPa; density input range: 0.01 ~ 1000000 g/cm3. p.21 ↗ Material properties must be entered within these ranges for stress conversion calculations. USF-2000/USF-2000A
Amplifier output correction factor range: 0.800 to 1.200. p.22 ↗ Correction factor can be adjusted within this range to calibrate amplifier output. USF-2000/USF-2000A
Specimen maximum diameter D1: 1 ~ 20 mm; minimum diameter D2: 1 mm to D1. p.22 ↗ Specimen geometry must be within these limits for proper stress calculation. USF-2000/USF-2000A
Notch bottom radius R: 0.1 ~ 10 mm; notch angle θ: 10 ~ 150 deg. p.23 ↗ Notch geometry must be within these ranges; stress concentration factor is calculated automatically. USF-2000/USF-2000A
Intermittent operation oscillation time: 110 ~ 500 ms; stop time: 100 ~ 5000 ms. p.25 ↗ These times control the ON/OFF cycle to suppress specimen heating. USF-2000/USF-2000A
Counter preset (number of cycles) range: 20000 ~ 1×10^12 times. p.26 ↗ Test can be programmed to stop after a specified number of cycles within this range. USF-2000/USF-2000A
Frequency fluctuation width range: 10 ~ 1000 Hz. p.26 ↗ If resonance frequency variation exceeds this width, the test stops and is regarded as specimen fatigue. USF-2000/USF-2000A
Stress fluctuation test: up to 10 steps, loop count 1 ~ 100, total cycle 20000 ~ 1×10^12. p.26 ↗ Complex stress sequences can be programmed with multiple steps and loops. USF-2000/USF-2000A
Measured edge amplitude input range: 0.80 to 600.00 μm. p.30 ↗ Used in correction magnification adjustment to calculate new correction factor. USF-2000/USF-2000A
Data display maximum 32000 points; data collection up to 1.0×10^10 times. p.32 ↗ Data collection is limited to these maximums; plan test duration accordingly. USF-2000/USF-2000A
Data collection initial cycle: 20000 ~ 1×10^11; end cycle: initial ~ 1×10^12; interval: 20000 ~ 1×10^11. p.33 ↗ Data collection cycles can be set logarithmically or linearly within these ranges. USF-2000/USF-2000A
Amplifier output amplitude: 10-120%; horn side amplitude: 1 ~ 120 μm. p.37 ↗ Calibration settings must be within these ranges. USF-2000/USF-2000A
External trigger signal is HIGH when test is started. p.35 ↗ Can be used to synchronize external devices with test start. USF-2000/USF-2000A
If test stress is below minimum stress, test is aborted with warning; if below recommended stress, warning allows continuation. p.34 ↗ Operator must confirm to proceed with stress below recommended minimum. USF-2000/USF-2000A
Configuration file SuperSonic.INI located in C:\Program Files\Shimadzu Corporation\USF-2000 (default). p.28 ↗ Maximum oscillation and stop times can be extended by editing this file; backup recommended. USF-2000/USF-2000A

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Cycle over p.35 ↗ Test completed the set number of cycles. Test ends; message displayed. To continue, press Test Start (counter not reset) or reset counter. Section 3.3.4
Frequency fluctuation range exceeded p.35 ↗ Resonance frequency variation exceeded the set fluctuation width, regarded as specimen fatigue. Test stops; message displayed. Inspect specimen. Section 3.3.4
Alarm p.35 ↗ Test machine has become an alarm state. Test stops; message displayed. Check machine alarm. Section 3.3.4
emergency stop p.36 ↗ Emergency stop button pressed. Test stops; message displayed. Release emergency stop and restart if safe. Section 3.3.4
Break p.36 ↗ Break detection function detected specimen break. Test ends; message displayed. Replace specimen. Section 3.3.4
(No message display) p.36 ↗ Manual stop via Test Stop button. Test stops; no message. Counter does not become zero. Section 3.3.4