Near Infrared Photometric Accuracy of Agilent Cary 7000 UV-Vis-NIR Spectrophotometer

Technical Reference Summary

Covered Models & Document Scope

Technical overview evaluating photometric (absorbance) accuracy in the near infrared (NIR) region of the Agilent Cary 7000 universal measurement spectrophotometer (UMS) using three metal-on-quartz filter reference materials; no model/size variants or ordering codes are covered.

Technical Data

Parameter Value Condition
Absorbance accuracy requirement below 1 Abs ±0.01 % p.2 ↗ USP/Ph. Eur. guideline
Absorbance accuracy requirement above 1 Abs ±1 % p.2 ↗ USP/Ph. Eur. guideline
Certified absorbance value of filter 0.3688 Abs p.3 ↗ 1,100 nm
Certified absorbance tolerance of filter ±0.0025 Abs p.3 ↗ 1,100 nm
Measured absorbance (Cary 7000) 0.3694 Abs p.3 ↗ 1,100 nm, n = 3
Absorbance error 0.1894 % p.3 ↗ 1,100 nm

Operating Limits

  • Absorbance accuracy requirement below 1 Abs: ±0.01% p.2 ↗
  • Absorbance accuracy requirement above 1 Abs: ±1% p.2 ↗