Contents

  1. Overview p. 2
    Background and purpose of automated amino acid analysis
  2. Procedure pp. 3-4
    Preparation of amino acid standards and caffeine stock solutions
  3. Quality pp. 5-7
    Calibration linearity, sensitivity, repeatability, and retention time performance
  4. p. 8
    Summary of automation benefits and analytical performance
  5. p. 5 table, 7 rows
  6. p. 6 table, 22 rows
  7. p. 7 table, 22 rows
  8. p. 8 table, 2 rows

Table 1. Chromatographic conditions for analysis of derivatized amino acids.

ParameterValue
ColumnAgilent AdvanceBio AAA LC column, 3.0 × 100 mm, 2.7 μm with Agilent AdvanceBio AAA guard column, 3.0 × 5 mm, 2.7 μm
SolventA) 10 mM Na2HPO4 and 10 mM Na2B4O7, pH: 8.2 B) Acetonitrile/Methanol/Water (45:45:10; v:v:v)
GradientTime (min) %A %B 0 98 2 0.40 98 2 13.60 43 57 14 0 100 Stop time: 17 min Post time: 3 min
Flow Rate0.600 mL/min
Temperature40 °C
DetectionFLD: Excitation: 345 nm; emission: 455 nm After 13.00 min: change excitation: 265 nm; change emission: 315 nm PMT gain: 10 Peak width: > 0.025 min (18.52 Hz)
InjectionAgilent 1260 Infinity III Vialsampler with 100 µL analytical head and 100 µL loop – Injection volume: 1 µL – Needle wash: 3 s with acetonitrile: 0.1 M HCl (50:50; v:v) – Sampling speeds: draw speed: 200 µL/min; eject speed: 400 µL/min – Needle height position: offset 0.0 mm – Sample temperature: 8 °C Agilent 1260 Infinity III Multisampler with 100 µL analytical head and 100 µL loop – Injection volume: 1 µL – Needle wash: 3 s with acetonitrile: 0.1 M HCl (50:50; v:v) – Sampling speeds: draw speed: 200 µL/min; eject speed: 400 µL/min – Needle height position: use vial bottom sensing, offset 0.0 mm – Sample temperature: 8 °C

Table 3. Calibration linearity, sensitivity, and repeatability of amino acid analysis using injector workflows on an Agilent 1260 Infinity III Multisampler. Ten consecutive injections of the amino acid mixture with a final concentration of 22.5 pmol/µL for each amino acid and 50 pmol/µL ISTDs were used for relative standard deviation (RSD) calculations. The limit of detection (LOD) was determined by using a threshold of 3x the signal‑to‑noise (S/N) value.

Peak No.CompoundRT (min)RT RSD (%)Area RSD (%)LOD (pmol/µL)Calibration Range (pmol/µL)Calibration TypeR²
1L-Aspartic acid1.070.610.340.2270.45 to 90Linear0.99999
2L-Glutamic acid1.560.600.360.1170.45 to 90Linear0.99997
3L-Asparagine3.470.070.250.1120.45 to 90Linear0.99990
4L-Serine3.680.060.290.0680.45 to 90Linear0.99992
5L-Glutamine4.200.040.230.1210.45 to 90Linear0.99986
6L-Histidine4.380.030.980.1430.45 to 90Quadratic0.99999
7Glycine4.620.030.320.0900.45 to 90Linear0.99997
8L-Threonine4.770.030.200.1220.45 to 90Linear0.99997
9L-Arginine5.380.020.210.1070.45 to 90Linear0.99998
10L-Alanine5.720.020.210.1100.45 to 90Linear0.99998
11L-Tyrosine6.690.020.300.1100.45 to 90Linear0.99998
12L-Cysteine7.760.041.711.3992.25 to 90Quadratic0.99994
13L-Valine8.110.030.230.1090.45 to 90Linear0.99997
14L-Methionine8.280.030.910.1420.45 to 90Linear0.99998
15L-Norvaline*8.540.030.36NANANANA
16L-Tryptophan8.960.030.490.1330.45 to 90Linear0.99996
17L-Phenylalanine9.240.020.210.1260.45 to 90Linear0.99998
18L-Isoleucine9.390.020.330.1160.45 to 90Linear0.99999
19L-Leucine9.900.020.340.1210.45 to 90Linear0.99999
20L-Lysine10.300.022.320.390.45 to 90Quadratic0.99993
21Sarcosine*12.350.023.29NANANANA
22L-Proline12.840.023.181.6062.25 to 90Quadratic0.99943

* ISTD

Table 4. Calibration linearity, sensitivity, and repeatability of amino acid analysis using injector workflows on an Agilent 1260 Infinity III Vialsampler. Ten consecutive injections of the amino acid mixture with a final concentration of 22.5 pmol/µL for each amino acid and 50 pmol/µL ISTDs were used for RSD calculations. The LOD was determined by using a threshold of 3x the S/N value.

Peak No.CompoundRT (min)RT RSD (%)Area RSD (%)LOD (pmol/µL)Calibration Range (pmol/µL)Calibration TypeR²
1L-Aspartic acid1.1940.3380.4690.1670.45 to 90Linear0.99994
2L-Glutamic acid1.8080.4560.6320.2670.45 to 90Linear0.99995
3L-Asparagine3.6570.1390.5120.1370.45 to 90Linear0.99980
4L-Serine3.8750.1100.5630.1370.45 to 90Linear0.99997
5L-Glutamine4.3890,0810.6230.1410.45 to 90Linear0.99967
6L-Histidine4.5730.0791.3130.2540.45 to 90Quadratic0.99972
7Glycine4.8290.0790.5590.1660.45 to 90Linear0.99996
8L-Threonine4.9660.0750.5410.2130.45 to 90Linear0.99984
9L-Arginine5.5690.0700.5880.1840.45 to 90Linear0.99998
10L-Alanine5.9330.0500.6390.1860.45 to 90Linear0.99997
11L-Tyrosine6.9120.0780.6430.1630.45 to 90Linear0.99998
12L-Cysteine7.9910.0481.1810.8860.90 to 90Quadratic0.99973
13L-Valine8.3260.0440.6760.1490.45 to 90Linear0.99998
14L-Methionine8.5170.0372.8920.5630.90 to 90Linear0.99962
15L-Norvaline*8.7640.0340.573NANANANA
16L-Tryptophan9.2090.0260.6530.1560.45 to 90Linear0.99981
17L-Phenylalanine9.4790.0250.6750.1870.45 to 90Linear0.99999
18L-Isoleucine9.6150.0250.5250.1690.45 to 90Linear0.99997
19L-Leucine10.1370.0170.5810.1890.45 to 90Linear0.99997
20L-Lysine10.5350.0121.2870.5640.45 to 90Quadratic0.99999
21Sarcosine*12.5780.0163.521NANANANA
22L-Proline13.0670.0132.9610.4770.90 to 90Quadratic0.99965

* ISTD

Table 5. Caffeine calibrations using calibration solutions prepared with the Agilent 1260 Infinity III Multisampler and Agilent 1260 Infinity III Vialsampler.

AutosamplerCalibration FunctionCorrelation R2Deviation Control Sample (%)
Agilent 1260 Infinity II Multisamplery = 28.3784x – 0.4930R² = 0.999943.58%
Agilent 1260 Infinity II Vialsamplery = 28.5657x – 0.7107R² = 0.999990.79%

Engineering Facts

13 facts · every claim linked to a page
Selection

The autosampler default configuration uses a 100 µL analytical head and 100 µL sample loop, so method volumes must fit this setup.

Determines whether the standard injector hardware can deliver the required sample and reagent volumes without modification.

Source
“100 µL analytical head and 100 µL sample loop (default setup)”
Selection

The Vialsampler requires the integrated column compartment while the Multisampler requires an additional Multicolumn Thermostat, affecting system configuration.

Impacts the module list and bench space needed when choosing between the two autosamplers.

Source
“Multisampler (G7167A) with 100 µL analytical head and 100 µL sample loop (default setup) and additional Multicolumn Thermostat (G7116A)”
Compliance

The injector workflow software requires OpenLab CDS version 2.8 or later.

Must be verified before purchase/installation to ensure the automation workflows are supported.

Source
“Agilent OpenLab CDS version 2.8 or later”
Operation

Reagents in the autosampler must be replaced daily and stored at 4 to 8 °C.

Defines the daily operating routine and sample compartment temperature control needed for reliable derivatization.

Source
“All reagents should be stored at 4 to 8 °C, and reagents in the autosampler should be exchanged daily.”
Maintenance

OPA/FMOC reagents must be aliquoted into amber vials and used within one week after opening.

Sets the reagent shelf-life limit that governs consumable planning and workflow scheduling.

Source
“stored for no longer than a week”
Operation

Stock solutions should be aliquoted before freezing to avoid repeated freeze-thaw cycles.

Prevents degradation of standards that would compromise calibration accuracy.

Source
“It is recommended to distribute the stock solutions before freezing them at –20 °C to avoid freeze-thaw cycles.”
Operation

Flat glass vial inserts are required to ensure proper mixing during the dilution series workflow.

Specifies the correct vial insert type needed for reliable automated serial dilutions.

Source
“the use of flat glass vial inserts (part number 5181-3377) is necessary to ensure proper mixing of the dilution.”
Operation

The Multisampler uses vial bottom sensing with a 0.0 mm offset for amino acid analysis, while the caffeine method uses a –3.0 mm offset.

Needle height settings must be configured per method to ensure correct sample aspiration.

Source
“Needle height position: use vial bottom sensing, offset 0.0 mm”
Performance

The amino acid method separates 20 amino acids and two ISTDs within 17 minutes.

Defines the achievable throughput and run time for the derivatization workflow.

Source
“The chromatogram shows a successful separation of 20 amino acids and two ISTDs within 17 minutes.”
Performance

Most compounds achieved RT RSD below 0.1% and peak area RSD below 1% over ten injections.

Demonstrates the precision achievable with injector workflows for method validation planning.

Source
“showing values below 0.1% for RT RSD and below 1% for peak area RSD”
Performance

LOD was mostly below 0.4 pmol/µL for all amino acids except L-cysteine and L-proline.

Sets realistic sensitivity expectations and flags analytes needing special attention.

Source
“excellent sensitivity with LOD mostly below 0.4 pmol/µL was obtained for all amino acids except L-cysteine and L-proline”
Performance

The caffeine dilution workflow covers a serial dilution range of 25 to 0.04 µg/mL.

Defines the usable calibration range for the automated dilution series.

Source
“a serial dilution of the caffeine stock solution within a range of 25 to 0.04 µg/mL”
Performance

The Vialsampler gave a 0.79% deviation on the control sample versus 3.58% for the Multisampler in caffeine calibration.

Helps select the autosampler based on dilution accuracy requirements.

Source
“Agilent 1260 Infinity II Vialsampler”