Turbo-V 301-AG Rack Controller User Manual
Technical Reference Summary
Covered Models & Document Scope
Covers the Turbo-V 301-AG Rack Controller, a microprocessor-controlled frequency converter for turbomolecular pumps, with front/remote/serial control, 24 Vdc cooling fan drive, vent valve control, pressure reading, optional Profibus interface, and automatic input voltage setting. The manual is multilingual (Italian, English, German, French, Spanish, Portuguese, Dutch, Danish, Swedish, Norwegian, Finnish, Hungarian, Polish, Czech, Slovak, Slovenian).
Section-by-Section Technical Breakdown
The Turbo-V 301 Rack series controllers are microprocessor-controlled frequency converters built from solid-state components with self-diagnosis and self-protection functions. They support front, remote, and serial control, 24 Vdc cooling fan drive, vent valve control, stop-speed reading, pressure reading, optional Profibus interface, and automatic input voltage setting.
- Front / remote / serial operation
- 24 Vdc control of the pump cooling fan
- Vent valve control
- Pump speed reading after stop command (stop speed reading)
- Pressure reading
- Profibus interface (option)
- Automatic input voltage setting
The controller can be installed on a bench or in a suitable rack, provided cooling air can circulate freely around the unit. It must not be installed in environments exposed to atmospheric agents (rain, snow, ice), steam, aggressive gases, or explosive/flammable atmospheres. Operating ambient temperature must be +5 °C to +45 °C with relative humidity 0–95% non-condensing.
- Use only the supplied three-conductor power cable with an approved plug and a properly earthed wall outlet.
- Always disconnect the power cable before installation or maintenance work.
- Operating temperature: +5 °C to +45 °C.
- Relative humidity: 0–95% (non-condensing).
- Ensure free air circulation around the controller.
The front panel includes buttons for displaying cycle number, cycle time, and pump life; selecting LOW SPEED; displaying pump current, temperature, power, and rotational speed; and sending START/STOP/RESET commands. A 2-line x 16-character alphanumeric LCD is provided. Pressing buttons 3 and 1 simultaneously for at least 2 seconds starts a programming routine for certain operating parameters.
- Button 1: display cycle number, cycle time, pump life.
- Button 2: LOW SPEED selection (active only in front panel mode); toggles stand-by speed.
- Button 3: display pump current, pump temperature, pump power, rotational speed (always active).
- Button 4: START/STOP/RESET (active only in front panel mode); after a fault stop, press once to reset and again to restart.
- Button 3 + Button 1 held ≥2 s: enter programming mode.
- LCD: dot matrix, 2 lines x 16 characters.
The rear panel provides an input connector for logic signals (with a special shorting bridge), an output connector for logic signals and pump frequency control plus programmable analog output, a 24 Vdc outlet for cooling fan/vent valve/gauge, a power input module with fuses and ENC filter, a Profibus address selector, an optional Profibus connector, a serial connector for remote control, and a pump connector.
- Input logic signal connector (supplied with shorting bridge).
- Output logic signals and pump frequency control plus programmable analog output.
- 24 Vdc outlet for cooling fan, vent valve, and gauge.
- Power input module including fuses, power outlet, and ENC filter.
- Profibus address selector.
- Profibus connector (option).
- Serial connector for remote control.
- Pump connector.
To switch on the controller, insert the power cable into the wall outlet. To start the pump, enable the interlock input and press the START button on the front panel. To stop the pump, press the STOP button on the front panel.
- Controller ON: connect power cable to mains outlet.
- Pump START: enable interlock input and press START.
- Pump STOP: press STOP on front panel.
Turbo-V 301 Rack series controllers are maintenance-free. Any work must be performed by authorized personnel. In case of failure, Agilent repair service or the Agilent advanced exchange service can provide a replacement controller.
- Controllers are maintenance-free.
- Service only by authorized personnel.
- Disconnect power cable before any work on the controller.
- Agilent advanced exchange service available for replacement.
The controller self-diagnosis circuit displays error messages on the LCD in case of certain faults. The messages include CHECK CONNECTION TO PUMP, WAITING INTERLOCK, PUMP OVERTEMP., and CONTROLLER OVERTEMP., each with a description and corrective action.
- CHECK CONNECTION TO PUMP: faulty pump-controller connection or pump temperature below 0 °C; check cable and press START twice.
- WAITING INTERLOCK: interlock signal active due to open short between pin 3 and pin 8 of connector J1/P1 or external interlock open; restore short or close external interlock.
- PUMP OVERTEMP.: upper bearing or pump temperature exceeded 60 °C; wait for cooldown and press START twice.
- CONTROLLER OVERTEMP.: controller temperature exceeded 70 °C or heatsink temperature exceeded 60 °C; wait for cooldown and press START twice.
This section provides detailed technical data including controller specifications, outline, connections, R1–R2 output diagrams, programmable analog output diagrams, J2 serial connector, letter protocol description, front panel use, programming, remote I/O use, serial mode use, Profibus option, error messages, and orderable parts.
- Controller specifications and outline.
- Controller connections and output diagrams.
- Serial connector J2 and letter protocol.
- Front panel programming and use.
- Remote I/O and serial mode operation.
- Profibus option details.
- Orderable parts.
The Profibus option provides an input buffer with defined offsets for reading voltage, pump bearing temperature, controller power supply temperature, controller CPU section temperature, status, error code bit field, and set point/valve statuses. Temperature thresholds trigger fail status: pump bearing >60 °C, controller power section >75 °C, controller CPU section >80 °C.
- Input buffer offset 0: voltage provided to motor (0-100 V).
- Offset 1: pump bearing temperature (0-100 °C); >60 °C causes fail and pump stop.
- Offset 2: controller power supply temperature (0-100 °C); >75 °C causes 'controller over-temperature' fail.
- Offset 3: controller CPU section temperature (0-100 °C); >80 °C causes 'controller overtemperature' fail.
- Offset 4: status B0-3 (0=Stop, 1=Waiting for interlock, 2=Ramp, 3=Auto tuning, 4=Braking, 5=Normal operation, 6=Fail).
- Offset 5: error code bit field (B7 Too High Load, B6 Short Circuit, B5 Over Voltage, B4 Verify Line Select, B3 Power Fail, B2 Controller Over-Temperature, B1 Pump Over-Temperature, B0 Check Connection To Pump).
- Offset 6: set point R1/R2 status, vent valve status, purge valve status.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Operating ambient temperature range is +5 °C to +45 °C. p.91 ↗ | Ensure installation location maintains this temperature range; otherwise controller may malfunction or derate. | Turbo-V 301-AG Rack Controller |
| Relative humidity must be 0–95% non-condensing. p.91 ↗ | Avoid condensing environments; condensation can cause electrical shorts. | Turbo-V 301-AG Rack Controller |
| Pump bearing overtemperature trip at 60 °C (upper bearing or pump body). p.73 ↗ | If PUMP OVERTEMP. occurs, wait for cooldown below threshold and press START twice to restart. | Turbo-V 301-AG Rack Controller |
| Controller overtemperature trip at 70 °C internal or 60 °C heatsink. p.73 ↗ | Ensure adequate cooling airflow; if CONTROLLER OVERTEMP. occurs, wait for cooldown and press START twice. | Turbo-V 301-AG Rack Controller |
| Profibus input buffer: pump bearing temperature >60 °C causes fail status and pump stop. p.278 ↗ | Monitor WIN 201 via Profibus to detect bearing overheating early. | Turbo-V 301-AG Rack Controller with Profibus option |
| Profibus input buffer: controller power supply temperature >75 °C causes 'controller over-temperature' fail. p.278 ↗ | Monitor WIN 216; ensure cooling to prevent power section overheating. | Turbo-V 301-AG Rack Controller with Profibus option |
| Profibus input buffer: controller CPU section temperature >80 °C causes 'controller overtemperature' fail. p.278 ↗ | Monitor WIN 211; ensure adequate ventilation around controller. | Turbo-V 301-AG Rack Controller with Profibus option |
| Interlock signal is on connector P1; a short between pin 3 and pin 8 of connector J1 (or P1) must be present or external interlock closed. p.73 ↗ | If WAITING INTERLOCK appears, check the short between pin 3 and pin 8 or close external interlock. | Turbo-V 301-AG Rack Controller |
| After a fault stop, press START once to reset the controller and a second time to restart the pump. p.81 ↗ | Two presses are required to recover from fault conditions. | Turbo-V 301-AG Rack Controller |
| Pressing buttons 3 and 1 simultaneously for at least 2 seconds enters programming mode for certain operating parameters. p.81 ↗ | Use this key combination to access parameter programming via front panel. | Turbo-V 301-AG Rack Controller |
| LOW SPEED button toggles pump to stand-by speed; active only when front panel control mode is selected. p.81 ↗ | Use LOW SPEED for standby operation; not available in remote or serial mode. | Turbo-V 301-AG Rack Controller |
| Start Output Mode default is 0 = Starting (output ON only when pump status = Starting); option 1 = running (output ON when pump is running). p.243 ↗ | Configure WIN 177 according to whether the start output should indicate starting phase or running state. | Turbo-V 301-AG Rack Controller |
| Pressure unit of measure selectable via WIN 163: 0 = mBar, 1 = Pa, 2 = Torr. p.243 ↗ | Set correct unit for pressure readings to avoid misinterpretation. | Turbo-V 301-AG Rack Controller |
| Stop speed reading can be enabled/disabled via WIN 167 (0 = disable, 1 = enable). p.243 ↗ | Enable to read pump speed after stop command. | Turbo-V 301-AG Rack Controller |
| R2 Set Point Type selectable via WIN 171: 0 = Freq, 1 = Power, 2 = Time, 3 = Normal, 4 = Pressure. p.243 ↗ | Configure R2 set point according to the parameter used for triggering. | Turbo-V 301-AG Rack Controller |
| R2 Set Point Signal Activation Type via WIN 174: 0 = high level active, 1 = low level active. p.243 ↗ | Match activation polarity to external control system. | Turbo-V 301-AG Rack Controller |
| Pump status codes: Stop=0, Waiting intlk=1, Starting=2, Auto-tuning=3, Braking=4, Normal=5, Fail=6. p.243 ↗ | Use these codes for remote monitoring and diagnostics. | Turbo-V 301-AG Rack Controller |
| Error code bit field: B7 Too High Load, B6 Short Circuit, B5 Over Voltage, B4 Verify Line Select, B3 Power Fail, B2 Controller Over-Temperature, B1 Pump Over-Temperature, B0 Check Connection To Pump. p.279 ↗ | Decode error code byte (offset 5) to identify specific fault. | Turbo-V 301-AG Rack Controller with Profibus option |
| Use only the supplied three-conductor power cable with approved plug and properly earthed outlet. p.91 ↗ | Prevents electric shock and ensures CE compliance. | Turbo-V 301-AG Rack Controller |
| Always disconnect power cable before any installation or maintenance work. p.91 ↗ | High voltages inside controller can cause severe injury or death. | Turbo-V 301-AG Rack Controller |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| CHECK CONNECTION TO PUMP p.73 ↗ | Faulty connection between pump and controller, or pump temperature below 0 °C. | Check that the connection cable between pump and controller is properly fastened at both ends and not interrupted. Press START twice to restart the pump. | Tab. 1 |
| WAITING INTERLOCK p.73 ↗ | Interlock signal on connector P1 is active due to interruption of the short circuit between pin 3 and pin 8 of connector J1, or due to opening of the external interlock signal. | Restore the short circuit between pin 3 and pin 8 of connector P1, or close the external interlock signal. | Tab. 1 |
| PUMP OVERTEMP. p.73 ↗ | Temperature of the upper bearing or pump exceeded 60 °C. | Wait until temperature falls below threshold. Press START twice to restart the pump. | Tab. 1 |
| CONTROLLER OVERTEMP. p.73 ↗ | Controller temperature exceeded 70 °C, or controller heatsink temperature exceeded 60 °C. | Wait until temperature falls below threshold. Press START twice to restart the pump. | Tab. 1 |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Operating ambient temperature | +5 to +45 °C p.91 ↗ | |
| Operating relative humidity | 0 – 95 % (non-condensing) p.91 ↗ | |
| Pump bearing overtemperature trip | 60 °C p.73 ↗ | upper bearing or pump body |
| Controller overtemperature trip | 70 °C p.73 ↗ | controller internal temperature |
| Controller heatsink overtemperature trip | 60 °C p.73 ↗ | controller radiator temperature |
| Pump bearing temperature fail threshold | 60 °C p.278 ↗ | Profibus input buffer, WIN 201 |
| Controller power supply temperature fail threshold | 75 °C p.278 ↗ | Profibus input buffer, WIN 216 |
| Controller CPU section temperature fail threshold | 80 °C p.278 ↗ | Profibus input buffer, WIN 211 |
| Pump temperature reading range | 0 to 70 °C p.243 ↗ | WIN 204 |
| Pump current reading | 0-100 mA dc p.243 ↗ | WIN 200 |
| Pump voltage reading | 0-100 Vdc p.243 ↗ | WIN 201 |
| Pump power reading | 0-100 W p.243 ↗ | WIN 202 (pump current x pump voltage duty cycle) |
| Driving frequency reading | 0-100 Hz p.243 ↗ | WIN 203 |
| Controller heatsink temperature reading | 0-100 °C p.243 ↗ | WIN 211 |
| Pressure unit of measure | 0 = mBar, 1 = Pa, 2 = Torr p.243 ↗ | WIN 163 |
| Stop speed reading | 0 = disable, 1 = enable p.243 ↗ | WIN 167 |
| R2 Set Point Type | 0 = Freq, 1 = Power, 2 = Time, 3 = Normal, 4 = Pressure p.243 ↗ | WIN 171 |
| R2 Set Point Signal Activation Type | 0 = high level active, 1 = low level active p.243 ↗ | WIN 174 |
| Start Output Mode | 0 = Starting (Output ON only with pump Status = Starting) DEFAULT; 1 = running (Output ON when the pump is running) p.243 ↗ | WIN 177 |
| Pump status codes | Stop = 0, Waiting intlk = 1, Starting = 2, Auto-tuning = 3, Braking = 4, Normal = 5, Fail = 6 p.243 ↗ | WIN 205 |