TwisTorr 74 FS User Manual

Technical Reference Summary

Covered Models & Document Scope

Covers TwisTorr 74 FS turbomolecular pump models X3502-64170, X3502-64171, X3502-64172, X3502-64173. Compatible controllers listed: X3508-64301 (AG Rack RS232/485), X3508-64022 (AG Rack Profibus), X3509-64030 (on board 110/220V), X3509-64021 (on board 24V), X3510-64050 (PCB). Optional water cooling kits X3502-68002, X3502-68003; air cooling fan 9699290.

Section-by-Section Technical Breakdown

pp. 294
Technical Information – Description of the TwisTorr 74 FS

Describes the pump construction with self-positioning aluminium alloy stator disks having parallel spiral pumping channels. The motor is fed with 54 Vac three-phase at 1167 Hz, with a start-up ramp using a higher initial voltage/frequency ratio. Cooling is required before starting; water or air cooling options are available, and a thermistor near the upper bearing protects against overheating.

  • Stators are self-positioning machined aluminium alloy disks with spiral channels pumping centrifugally on one side and centripetally on the other.
  • Motor supply: 54 Vac three-phase at 1167 Hz; frequency ramps up during start-up with higher initial V/f ratio.
  • Pump must always be cooled before starting; running without cooling can damage internal components.
  • Water cooling kits X3502-68002 and X3502-68003; air cooling fan 9699290.
  • Thermistor near upper bearing trips at >60 °C (Error Code 2) and <0 °C (Error Code 1).
  • Pump can operate in any position and can be supported on the high vacuum flange.
  • Forevacuum connection is KF 16 NW on the side of the pump.
pp. 279-281
Installation

The pump must be installed before running; operating it unconnected exposes rotating parts. The controller power cord serves as the emergency disconnecting device and must be easily reachable. The pump must be used with a suitable Agilent controller and connected to a primary pump. Installation is indoors only, with specified environmental limits.

  • Install the pump before running it; never operate unconnected to the system.
  • Controller power cord is the disconnecting device; ensure easy access and space behind the socket.
  • Do not remove adhesive and protective cap before connecting the turbopump to the system.
  • Indoor use only; avoid rain, ice, snow, dust, aggressive gases, explosive or high fire risk environments.
  • Operating environment: max pressure 2 bar above atmospheric, temperature +5 °C to +35 °C, relative humidity 0–90 % non-condensing.
  • In magnetic fields, protect the pump with a ferromagnetic shield.
  • Compatible controllers: X3508-64301, X3508-64022, X3509-64030, X3509-64021, X3510-64050.
  • Installation category II per EN 61010-1; connect to a mains line satisfying this category.
pp. 282
Pump Fixing

The TwisTorr 74 FS can be installed in any position. The inlet flange must be mounted to the system counter-flange with a connection capable of withstanding 250 Nm around its axis. A table specifies clamp quantities and fixing torques for each flange type. The pump cannot be fixed by means of its base.

  • Install in any position; fix in a stable position.
  • Connection must withstand 250 Nm torque around the axis.
  • ISO 63 K: 4 × M10 clamps IC63250DCMZ, 22 Nm.
  • CFF 4 ½: 8 × mounting kit X3502-68007 / screws X3502-68008 or X3502-68009, 20 Nm.
  • CFF 2 ¾: 6 × American screw 1/4"-28 UNF 2B, 20 Nm.
  • KF 40: 1 × Clamp KF40, no torque specified.
  • The TwisTorr 74 FS cannot be fixed by means of its base.
pp. 283-284
Use of TwisTorr 74 FS

All operating instructions are in the controller manual. The pump may be heated to obtain better limit pressures, with inlet flange temperature limits of 120 °C for ConFlat and 80 °C for ISO flanges, and water cooling must always be used while heating. Avoid impacts and harsh movements during operation; use clean air or inert gas for venting at less than 2 bar above atmospheric.

  • Read the controller manual before use.
  • Heating limits: inlet flange ≤120 °C for ConFlat, ≤80 °C for ISO; always use water cooling while heating.
  • Never use the turbopump when the inlet flange is not connected to the vacuum chamber.
  • Do not touch the pump or accessories during heating; high temperatures may cause burns.
  • Avoid impacts, oscillations or harsh movements during operation; bearings may be damaged.
  • Use air or inert gas free from dust or particles for venting; vent port pressure must be less than 2 bar above atmospheric.
  • For particulate or aggressive pollutants, use the integrated purge/vent port with inert gas (He, N2, Ar).
  • Never use with corrosive gases or vapors.
  • Designed for high throughput of N2, Ar and lighter gases; consult Agilent for gases heavier than Ar.
pp. 285
Maintenance

The TwisTorr 74 FS requires no maintenance; all operations must be performed by authorized personnel. Before any work, disconnect power, vent the pump, wait for the rotor to stop completely and for surface temperature to fall below 50 °C. Agilent repair or advanced exchange service is available.

  • Pump is maintenance-free; interventions only by authorized personnel.
  • Before work: disconnect power, vent via the dedicated valve, wait for complete rotor stop and surface temperature below 50 °C.
  • Agilent advanced exchange service provides a regenerated replacement pump.
  • A completed 'Request for Return' form must be sent to the local sales office and a copy placed in the packaging before shipping.
  • Disposal must follow specific national regulations.
pp. 277
Preparation for Installation

The pump is supplied in protective packaging; check for transport damage and avoid dropping or shock. Do not handle vacuum-exposed components with bare hands. Keep the pump closed until installation and perform a complete power on/off cycle once a month if unused for many months.

  • Contact local sales office if packaging shows transport damage.
  • Do not drop or subject the pump to sudden impact or shock vibration during unpacking.
  • Packaging is 100% recyclable and complies with EEC Directive 94/62.
  • Do not use bare hands on vacuum-exposed components; use gloves to prevent outgassing.
  • Keep pump closed until installation to avoid contamination.
  • If unused for many months, perform a complete power on/off cycle once a month.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Motor is fed with 54 Vac three-phase at 1167 Hz during normal operation, with a start-up ramp using a higher initial voltage/frequency ratio. p.294 ↗ The pump requires a compatible Agilent controller; do not connect directly to mains. TwisTorr 74 FS
Maximum rotational speed is 70000 rpm. p.273 ↗ High kinetic energy; strict adherence to installation instructions is required to prevent rotor crash consequences. TwisTorr 74 FS
Thermistor near upper bearing trips at >60 °C (Error Code 2) and <0 °C (Error Code 1). p.294 ↗ Ensure cooling is active before start and ambient temperature is within +5 to +35 °C to avoid trips. TwisTorr 74 FS
Pump must always be cooled before starting; water cooling kits X3502-68002 and X3502-68003 or air cooling fan 9699290 are available. p.294 ↗ Select and install cooling before commissioning; running without cooling can damage internal components. TwisTorr 74 FS
Inlet flange connection must withstand 250 Nm torque around its axis. p.282 ↗ Use a robust counter-flange and correct clamps/screws to prevent pump detachment in case of rotor failure. TwisTorr 74 FS
ISO 63 K flange requires 4 × M10 clamps IC63250DCMZ tightened to 22 Nm. p.282 ↗ Use specified clamp type and torque for safe installation. TwisTorr 74 FS with ISO 63 K inlet
CFF 4 ½ flange requires 8 × mounting kit X3502-68007 / screws X3502-68008 or X3502-68009 tightened to 20 Nm. p.282 ↗ Use Agilent hardware for ConFlat installation. TwisTorr 74 FS with CFF 4 ½ inlet
CFF 2 ¾ flange requires 6 × American screw 1/4"-28 UNF 2B tightened to 20 Nm. p.282 ↗ Use specified screws and torque for CFF 2 ¾ installation. TwisTorr 74 FS with CFF 2 ¾ inlet
Maximum inlet flange temperature during bakeout: 120 °C for ConFlat, 80 °C for ISO; always use water cooling while heating. p.283 ↗ Do not exceed bakeout limits; water cooling is mandatory during bakeout. TwisTorr 74 FS
Vent port pressure must be less than 2 bar above atmospheric pressure; use air or inert gas free from dust or particles. p.283 ↗ Install a suitable vent valve and filtered gas supply to avoid bearing damage. TwisTorr 74 FS
Operating environment: temperature +5 °C to +35 °C, relative humidity 0–90 % non-condensing, maximum pressure 2 bar above atmospheric. p.281 ↗ Ensure indoor installation within these limits; otherwise pump performance and safety may be impaired. TwisTorr 74 FS
Storage temperature range -40 °C to 70 °C; relative humidity 0–90 % non-condensing; shelf life 12 months from shipping date. p.276 ↗ Respect storage limits; if shelf life exceeded, return pump to factory. TwisTorr 74 FS
Pump can be installed in any position and can be supported on the high vacuum flange; it cannot be fixed by means of its base. p.294 ↗ Plan mounting using the inlet flange only; do not rely on the pump base for fixing. TwisTorr 74 FS
Forevacuum connection is KF 16 NW on the side of the pump. p.294 ↗ Provide a compatible KF 16 NW foreline connection. TwisTorr 74 FS
Compatible controllers: X3508-64301 (AG Rack RS232/485), X3508-64022 (AG Rack Profibus), X3509-64030 (on board 110/220V), X3509-64021 (on board 24V), X3510-64050 (PCB). p.281 ↗ Select one of these controllers; pump cannot operate without a controller. TwisTorr 74 FS
Before maintenance, disconnect power, vent the pump, wait for complete rotor stop and surface temperature below 50 °C. p.285 ↗ Follow lockout/vent/cool-down procedure to avoid injury and equipment damage. TwisTorr 74 FS
In magnetic fields, the pump must be protected using a ferromagnetic shield. p.281 ↗ Assess stray magnetic fields at installation location and provide shielding if needed. TwisTorr 74 FS
For pumping gases containing particulate or aggressive pollutants, use the integrated purge/vent port with a steady flow of inert gas (He, N2, Ar). p.283 ↗ Connect purge gas supply to protect bearings when pumping dirty or corrosive media. TwisTorr 74 FS
Pump is designed for high throughput of N2, Ar and lighter gases; consult Agilent for gases heavier than Ar. p.284 ↗ Verify gas compatibility before use; heavier gases may require special configuration. TwisTorr 74 FS
Installation category II per EN 61010-1; connect to a mains line satisfying this category. p.281 ↗ Ensure site electrical installation meets overvoltage category II requirements. TwisTorr 74 FS

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Error Code = 2 p.294 ↗ Thermistor reading near upper bearing exceeds 60 °C (overheating). Check cooling (water or air) and ensure it is operating; allow pump to cool before restart. See Controller Manual
Error Code = 1 p.294 ↗ Thermistor reading near upper bearing is below 0 °C. Raise ambient temperature within +5 °C to +35 °C operating range before restart. See Controller Manual

Technical Data

Parameter Value Condition
Motor supply voltage 54 Vac p.294 ↗ three-phase, 1167 Hz
Motor supply frequency 1167 Hz p.294 ↗ normal operation
Maximum rotational speed 70000 rpm p.273 ↗
Operating temperature range +5 to +35 °C p.281 ↗ during operation
Storage temperature range -40 to 70 °C p.276 ↗ shipping, moving, storing
Relative humidity range 0 to 90 % p.276 ↗ non-condensing
Maximum pressure at vent port 2 bar p.283 ↗ above atmospheric pressure
Maximum inlet flange temperature during bakeout 120 °C p.283 ↗ ConFlat flange
Maximum inlet flange temperature during bakeout 80 °C p.283 ↗ ISO flange
Bearing overtemperature trip 60 °C p.294 ↗ thermistor near upper bearing; Error Code = 2
Bearing undertemperature trip 0 °C p.294 ↗ thermistor near upper bearing; Error Code = 1
Shelf life 12 months p.276 ↗ from shipping date
Forevacuum connection KF 16 NW p.294 ↗ side of pump
Inlet flange fixing torque capability 250 Nm p.282 ↗ around axis
ISO 63 K flange clamp quantity 4 p.282 ↗ M10 clamps IC63250DCMZ
ISO 63 K flange fixing torque 22 Nm p.282 ↗ M10 clamps IC63250DCMZ
CFF 4 ½ flange clamp quantity 8 p.282 ↗ Mounting Kit X3502-68007 / screws X3502-68008 / X3502-68009
CFF 4 ½ flange fixing torque 20 Nm p.282 ↗ Mounting Kit X3502-68007 / screws X3502-68008 / X3502-68009
CFF 2 ¾ flange clamp quantity 6 p.282 ↗ American screw 1/4"-28 UNF 2B
CFF 2 ¾ flange fixing torque 20 Nm p.282 ↗ American screw 1/4"-28 UNF 2B
KF 40 flange clamp quantity 1 p.282 ↗ Clamp KF40