ACQUITY Arc MDLC System Customer Familiarization Guide
Technical Reference Summary
Covered Models & Document Scope
ACQUITY Arc MDLC System configurations covered by this Customer Familiarization Guide (715006691 VERSION 00). The document addresses systems with a CM-A (Column Manager-Active) and systems with a 30-cm CH (Column Heater) with external diverter valves. Detector options include two 2489 UV/Vis detectors, two 2998 PDA detectors, or one each of 2489 UV/Vis and 2998 PDA; an optional QDa detector is also supported. The guide is a performance-test/familiarization procedure, not a full operator manual.
Section-by-Section Technical Breakdown
Describes creating a chromatographic system in Empower 3, dragging modules into the system, and setting the primary flow source. Emphasizes that the first-dimension pump must be configured before the second-dimension pump and that configuration must be repeated when an instrument component is added or replaced.
- Empower 3 base is the minimum software; all instruments are released for Empower 3.
- Select File > New > Chromatographic System and drag modules from Available Instruments to New System Instruments.
- Configure the pump for the first dimension before configuring the pump for the second dimension.
- After adding or replacing an instrument component, repeat the system configuration procedure to achieve proper communication.
- Set the primary flow source via Console > Configure > Primary flow source, selecting the solvent manager for the first dimension.
Details Console configuration of the Column Manager-Active for MDLC operation. The CM-A must be operated in Advanced mode, configured for 4 columns, with specific column positions, ports, and preheater settings, followed by power cycling.
- To enable MDLC features, the CM-A must be operated in Advanced mode.
- Select 4 columns in the Instrument Configuration dialog.
- In Advanced mode, select 2 for left and right positions and 6 for left and right ports.
- Only one preheater can be installed in each trough, on either the left-hand or right-hand side.
- Select Active for column positions 1 and 2.
- Cycle power to the CM-A when the Configuration Changed dialog appears.
Outlines placing the QDa detector into Ready status via the Waters Console, including startup checks (source pressure test, critical readback system check) and resolution/calibration checks. Calibration and resolution checks are valid for one year.
- If the QDa is powered-on and in Ready status, no actions are needed.
- Resolution and calibration checks are valid for one year after Check Resolution or Check Calibration; if a check fails within that period, setup must be performed again.
- On startup from powered-down state, the source pressure test and critical readback system check are performed before Ready status.
- The source pressure test takes approximately three minutes; it also runs when the instrument door is closed.
- Resolution and calibration checks can be enabled or disabled in Configure > Startup Configuration; if disabled, manual checks must be performed annually or as required.
- To verify probe temperature has reached 600 °C, select MS Display.
Provides solvent preparation recipes for the 2489 UV/Vis and 2998 PDA detectors, including mobile phases A, B, C, D, seal wash, wash, and purge solvents. Also lists important recommendations on water quality, glassware cleaning, and solvent purity.
- Mobile phase A: 100% water with 0.5% formic acid (1 L water + 5 mL formic acid).
- Mobile phase B: 100% acetonitrile with 0.5% formic acid (1 L acetonitrile + 5 mL formic acid).
- Mobile phases C and D: 100% acetonitrile.
- Seal wash solvent: 90:10 water/acetonitrile (900 mL water + 100 mL acetonitrile).
- Use LCMS-grade solvents minimum; lower-quality solvents can cause signal suppression, high background, and loss of sensitivity.
- Do not wash glassware with detergents; rinse with high-purity solvents before use.
- Ensure the QSM-R flow path position parameter is set to Path 1.
Describes priming and flushing procedures for the QSM-R and SM FTN-R, including seal wash prime, purge solvent priming, solvent changeover, and equilibration settings. Specifies flow rates for D1 and D2 pumps and column temperature.
- To avoid fluid backup, ensure waste container is below the stack, waste lines have no crimps or bends, and degasser line ends are not immersed in waste fluid.
- Perform seal wash prime from dry.
- Prime SM FTN-R purge solvent with 30 cycles initially; later 60 cycles in solvent changeover.
- In MDLC systems, priming options appear twice, once for each QSM-R.
- Equilibration flow: 1.000 mL/min for D1 pump, 0.5 mL/min for D2 pump.
- Solvent A and B at 95/5; Position Path 1.
- Set column temperature to 50 °C; turn detector lamps on.
Details column flushing and equilibration steps: flush D1, D2, and trap columns to waste with 100% acetonitrile, then equilibrate at 95%A:5%B at 1.0 mL/min until baseline is stable (approximately 30 minutes).
- Flush D1 column with 100% acetonitrile for 20 minutes at 1.0 mL/min.
- Flush D2 column for 20 minutes and trap column for 3 minutes at 1.0 mL/min.
- Set column heater temperature to 50 °C and verify temperature is reached.
- Equilibrate at 95%A:5%B at 1.0 mL/min until baseline is stable (approximately 30 minutes).
- Flow may be reduced during equilibration to conserve mobile phase, but must be increased to 1.0 mL/min before running performance tests.
Provides instructions for preparing stock solution and test samples from ACQUITY UPLC MS Startup Solution (700002741-2). The single heart-cut test sample contains five analytes at 50 ppm each.
- Stock solution 700002741-2 contains acetaminophen, caffeine, sulfadimethoxine, reserpine, and terfenadine each at 1.0000 mg/mL.
- Single Heart-Cut test sample: pipette 100 µL stock into 2-mL vial, add 1,900 µL deionized water, vortex; final concentration 50 ppm each.
- Blank: 1 mL deionized water in a sample vial.
- On-column injection volume is 20 µL.
Describes acquiring and processing data for dimension 1 to determine the retention time of sulfadimethoxine. Specifies run time, injection volume, replicate injections, and acceptance criteria for %RSD of retention time and peak area.
- D1 analysis: injection volume 20 µL, run time 10 minutes.
- Injections: blank (2 replicates), sample (6 replicates).
- Approximate retention time of sulfadimethoxine is 3.9 minutes.
- Specification: sulfadimethoxine retention time %RSD ≤ 1%.
- Specification: sulfadimethoxine peak area %RSD ≤ 3%.
- For a system with one detector only, ensure the 'in' port of the detector is connected to the proper column outlet when set in D1 or D2 mode.
Describes performing the single heart-cut by adjusting event tables based on the D1 retention time, then running and processing the D2 analysis. Specifies run time, injection volume, and acceptance criteria.
- Heart-cut window: mean RT ± 0.1 minutes, assuming the 250-µL loop is installed on the left-side diverter valve.
- Enter RT minus 0.1 min into first line of event tables; RT plus 0.1 min into second line.
- D2 analysis: injection volume 20 µL, run time 20 minutes (all configurations).
- Injections: blank (2 replicates), sample (6 replicates).
- Specification: sulfadimethoxine retention time %RSD ≤ 1%.
- Specification: sulfadimethoxine peak area %RSD ≤ 5%.
Provides the procedure to power down the system via the Console, including solvent changeover with custom conditions, priming, and equilibration settings with zero flow.
- Use Control > Start up system > Solvent Changeover > Use Custom Conditions.
- Priming options: Solvent A, B, C, D, and Seal Wash; priming duration 5 minutes.
- SM FTN-R priming wash solvent 120 seconds; purge solvent 60 cycles.
- Equilibration flow: 0.00 mL/min for both D1 and D2 pumps.
- Solvent A and B at 95/5; Position Path 1.
- Set sample temperature to off if installed; set column temperature to off.
- After applying final conditions, select Do not wait.
Provides instrument method parameter settings for configurations using a CM-A and for configurations using a 30-cm CH with external diverter valves. Includes gradient tables, detector parameters, and event table settings.
- For CM-A configurations, set Advanced settings stroke volume to 66 µL for D2 in the QSM-R.
- For 30-cm CH configurations, set Advanced settings stroke volume to 66 µL for D2 in the QSM-R.
- For D1 retention time determination with CM-A, clear the Run Events selection in the CM-A events table and save method with 'D1' suffix.
- If using optional QDa detector, change run time in Functions tab to 0–10 minutes for D1 method.
- For 30-cm CH with external diverter valves, the detector event tables control valve switching via signal cables.
- The 2998 PDA event table can control both external diverter valves; 2489 and QDa event tables can control only one valve each.
- For 2489 TUV detector, always set threshold event absorbance units to 4.000 AU.
- Control D1 left-side external diverter valve with D1 detector's Switch 1; control right-side valve with D2 detector's Switch 1.
- For 2489 UV/Vis always use Switch 1; for 2998 PDA use Switch 1 for D1 (left) and Switch 2 for D2 (right).
Recommends regular annual QDa maintenance and refers to the ACQUITY QDa Detector Overview and Maintenance Guide for additional information.
- Waters recommends regular QDa maintenance annually.
- Consult the ACQUITY QDa Detector Overview and Maintenance Guide for details.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Empower 3 base is the minimum software version; all instruments are released for Empower 3. p.6 ↗ | System control and data acquisition require Empower 3 or later. | Whole series |
| The CM-A must be operated in Advanced mode to enable MDLC features. p.11 ↗ | Failure to set Advanced mode will prevent MDLC operation. | Configurations with CM-A |
| Only one preheater can be installed in each trough, on either the left-hand or right-hand side. p.12 ↗ | Preheater configuration must be selected accordingly in the Console. | Configurations with CM-A |
| Resolution and calibration checks are valid for one year after Check Resolution or Check Calibration; if a check fails within that period, setup must be performed again. p.13 ↗ | Annual verification is required; failed checks necessitate recalibration. | QDa detector |
| The source pressure test takes approximately three minutes and also runs when the instrument door is closed. p.15 ↗ | Allow sufficient time for startup; door closure triggers a pressure test. | QDa detector |
| To verify probe temperature has reached 600 °C, select MS Display. p.14 ↗ | Probe temperature must reach 600 °C for proper operation. | QDa detector |
| Mobile phase A: 100% water with 0.5% formic acid (1 L water + 5 mL formic acid). p.17 ↗ | Use LCMS-grade water and analytical-grade formic acid. | 2489 UV/Vis and 2998 PDA configurations |
| Mobile phase B: 100% acetonitrile with 0.5% formic acid (1 L acetonitrile + 5 mL formic acid). p.17 ↗ | Use LCMS-grade acetonitrile and analytical-grade formic acid. | 2489 UV/Vis and 2998 PDA configurations |
| Seal wash solvent: 90:10 water/acetonitrile (900 mL water + 100 mL acetonitrile). p.17 ↗ | Prepare seal wash solvent as specified for proper seal lubrication. | Whole series |
| Equilibration flow: 1.000 mL/min for D1 pump and 0.5 mL/min for D2 pump; solvent A and B at 95/5; Position Path 1. p.18 ↗ | These settings are used during system equilibration before tests. | Whole series |
| Set column temperature to 50 °C and verify temperature is reached before equilibration. p.19 ↗ | Column heater must be at 50 °C for performance tests. | Whole series |
| Flush D1 column with 100% acetonitrile for 20 minutes at 1.0 mL/min; flush D2 column for 20 minutes and trap column for 3 minutes at 1.0 mL/min. p.19 ↗ | Follow these flush durations to prepare columns for use. | Whole series |
| Equilibrate at 95%A:5%B at 1.0 mL/min until baseline is stable (approximately 30 minutes). p.19 ↗ | Allow sufficient equilibration time before running tests. | Whole series |
| On-column injection volume is 20 µL for both D1 and D2 analyses. p.21 ↗ | Use 20 µL injection volume for all performance tests. | Whole series |
| D1 analysis run time is 10 minutes; D2 analysis run time is 20 minutes (all configurations). p.21 ↗ | Set run times accordingly in instrument methods. | Whole series |
| Approximate retention time of sulfadimethoxine in D1 is 3.9 minutes. p.22 ↗ | Use this as a guide for setting heart-cut window. | Whole series |
| Heart-cut window: mean RT ± 0.1 minutes, assuming the 250-µL loop is installed on the left-side diverter valve. p.23 ↗ | Adjust event tables to capture sulfadimethoxine peak. | Whole series |
| D1 acceptance criteria: sulfadimethoxine retention time %RSD ≤ 1% and peak area %RSD ≤ 3%. p.22 ↗ | Test results must meet these specifications. | Whole series |
| D2 acceptance criteria: sulfadimethoxine retention time %RSD ≤ 1% and peak area %RSD ≤ 5%. p.26 ↗ | Test results must meet these specifications. | Whole series |
| For 2489 TUV detector, always set threshold event absorbance units to 4.000 AU. p.37 ↗ | Prevents sample loss to waste when using 2489 detector. | Configurations with 2489 UV/Vis and external diverter valves |