Plate Heater-Shaker+ User Manual
Technical Reference Summary
Covered Models & Document Scope
This user manual covers the Plate Heater-Shaker+ device, a laboratory instrument controlled via OneLab software. The document does not define specific model numbers or ordering codes; the device is identified by the Wi-Fi network name pattern HeaterShaker-5xx.4xxx.xxxxx, where xxxx.xxxxx is the serial number. No optional variants or accessories are described beyond Plate Adapters, which are referenced but not listed with part numbers.
Section-by-Section Technical Breakdown
The device features a power socket, Ethernet port, built-in wireless adapter, built-in computer, LED indicator, labware holder/Plate Adapter mounting area, power button, and an open hatch. Plate Adapters can be removed and replaced by powering off, cooling the heating elements, removing labware, loosening a Torx T8 screw, and inserting a new adapter straight into the plate holder.
- Power off and ensure heating elements are cold before removing a Plate Adapter.
- Loosen the Torx screw counterclockwise using the supplied T8 screwdriver.
- Insert the new Plate Adapter straight into the plate holder and ensure it is well aligned.
- Tighten the Torx screw clockwise after insertion.
The Plate Heater-Shaker+ is connected to a power source and enters installation mode after the LED indicator turns blue and blinks. Users connect a computer, tablet, or smartphone to the device's Wi-Fi network named HeaterShaker-5xx.4xxx.xxxxx and access http://192.168.2.1 to complete installation. If installation mode does not start automatically, pressing the power button 5 times forces it.
- Only use the provided AC/DC power supply unit and cords.
- Wait for the LED indicator to turn blue and blink before proceeding.
- Connect to Wi-Fi network HeaterShaker-5xx.4xxx.xxxxx, where xxxx.xxxxx is the serial number found on top of the device.
- Open a web browser and navigate to http://192.168.2.1 to follow on-screen installation steps.
- The LED indicator stops blinking when the device is successfully connected.
- If the device does not automatically enter installation mode, press the power button 5 times.
Experiments are designed and executed through OneLab software using a drag-and-drop protocol editor. Shaking and Heating actions are created from the left-side menu, and a temperature can be set within a Shaking action. OneLab automatically selects the Plate Heater-Shaker+ and supports manual, semi-automated, or fully automated experiment setups.
- Create Shaking or Heating actions via the left-side menu in OneLab.
- A temperature can be set within a Shaking action.
- OneLab automatically selects the Plate Heater-Shaker+ for experiments.
- Each step of the experiment is recorded and available in OneLab.
- If after 5 minutes the inside sensor does not reach a thermal plateau, the current temperature is automatically set at 18°C.
The Plate Heater-Shaker+ has a maximum heating temperature of 99°C, temperature resolution of 0.1°C, and temperature accuracy of ±0.1°C without a Plate Adapter mounted. Mixing frequency ranges from 200 rpm to 2500 rpm with a constant 2 mm orbit, 1 rpm speed setting resolution, and ±25 rpm regulation accuracy. The device measures 170 x 162 x 100 mm, weighs 2.55 kg, and is powered by an external 24 VDC, 5 A, 120 W supply.
- Maximum heating temperature: 99°C.
- Temperature resolution: 0.1°C; accuracy: ±0.1°C without Plate Adapter.
- Mixing frequency: 200 rpm up to 2500 rpm; maximum speed may differ by labware type.
- Mixing orbit constant: 2 mm; speed setting resolution: 1 rpm; regulation accuracy: ±25 rpm.
- Relative humidity: 80% at 37°C; maximum operating altitude: 2000 m.
- Ingress protection: IP20; pollution degree 2.
- Dimensions (W x L x H): 170 x 162 x 100 mm; weight: 2.55 kg.
- External power supply: input 100-240 VAC, 1.4 A; output 24 VDC, 5 A, 120 W.
- Network connectivity: Gigabit Ethernet and Wi-Fi 2.4 GHz and 5 GHz.
- Cleaning only with a wet tissue or isopropanol.
When installing multiple connected devices, place them in front of Andrew+ as indicated by OneLab, open the hatches, plug power cables into the power sockets, pass cables into the dedicated path in front of the devices, and close the hatches.
- Place connected devices in front of Andrew+ as indicated by OneLab.
- Open the hatches before plugging in power cables.
- Pass cables into the dedicated path in front of the devices.
- Close the hatches after cabling.
The device automatically starts up when plugged into a power source. When in standby mode (plugged in but turned off), press the power button to turn on. When in idle state (no experiment running), press the power button for 5 seconds to turn off.
- Device automatically starts up when plugged into a power source.
- From standby mode, press the power button to turn on.
- From idle state, press the power button for 5 seconds to turn off.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Maximum heating temperature is 99°C. p.6 ↗ | Do not set heating above 99°C; ensure labware and samples are compatible with this limit. | Plate Heater-Shaker+ |
| Temperature accuracy is ±0.1°C without a Plate Adapter mounted. p.6 ↗ | For precise temperature control, account for potential offset introduced by a Plate Adapter. | Plate Heater-Shaker+ |
| Mixing frequency range is 200 rpm up to 2500 rpm; maximum speed may differ according to labware type. p.6 ↗ | Verify maximum safe speed for the specific labware before operation. | Plate Heater-Shaker+ |
| Mixing orbit is constant at 2 mm. p.6 ↗ | Shaking amplitude is fixed; protocol design should rely on this orbit. | Plate Heater-Shaker+ |
| Speed setting resolution is 1 rpm and mixing regulation accuracy is ±25 rpm. p.6 ↗ | Actual speed may deviate by up to ±25 rpm from the setpoint; account for this in sensitive protocols. | Plate Heater-Shaker+ |
| External power supply input is 100-240 VAC, 1.4 A; output is 24 VDC, 5 A, 120 W. p.6 ↗ | Use only the provided AC/DC power supply unit and cords; ensure mains supply matches input range. | Plate Heater-Shaker+ |
| Operating voltage is 24 VDC and overvoltage category of DC port is Cat I. p.6 ↗ | Connect only to the supplied power supply; do not connect to higher-voltage circuits. | Plate Heater-Shaker+ |
| Maximum operating altitude is 2000 m. p.6 ↗ | Do not operate the device above 2000 m altitude. | Plate Heater-Shaker+ |
| Relative humidity limit is 80% at 37°C. p.6 ↗ | Operate only within specified humidity conditions to avoid condensation or damage. | Plate Heater-Shaker+ |
| Ingress protection is IP20; avoid spilling liquid on the device. p.6 ↗ | The device is not protected against liquid ingress; keep liquids away. | Plate Heater-Shaker+ |
| Dimensions are 170 x 162 x 100 mm (W x L x H) and weight is 2.55 kg; dimensions may change according to labware. p.6 ↗ | Allow sufficient bench space and account for labware overhang. | Plate Heater-Shaker+ |
| Network connectivity includes Gigabit Ethernet and Wi-Fi 2.4 GHz and 5 GHz. p.6 ↗ | Ensure network infrastructure supports these standards for reliable OneLab communication. | Plate Heater-Shaker+ |
| Cleaning must be done only with a wet tissue or isopropanol. p.6 ↗ | Do not use other solvents or cleaning agents that may damage surfaces. | Plate Heater-Shaker+ |
| If after 5 minutes the inside sensor does not reach a thermal plateau, the current temperature is automatically set at 18°C. p.5 ↗ | This allows selection of the device for lower-temperature experiments without waiting for room temperature equilibration. | Plate Heater-Shaker+ |
| To turn off the device from idle state, press the power button for 5 seconds. p.6 ↗ | A short press will not power off the device; hold for 5 seconds. | Plate Heater-Shaker+ |
| If the device does not automatically enter installation mode, press the power button 5 times. p.4 ↗ | This forces installation mode for network setup. | Plate Heater-Shaker+ |
| Plate Adapter replacement requires a Torx T8 screwdriver (supplied). p.3 ↗ | Ensure the T8 screwdriver is available for adapter changes. | Plate Heater-Shaker+ |
| The heating plate can reach temperatures in excess of 99°C; pay attention to residual heat after switching off and transport only when cold. p.2 ↗ | Allow sufficient cooling time before handling or moving the device. | Plate Heater-Shaker+ |
| All input/output ports are Safety Extra Low Voltage (SELV) circuits and should only be connected to other SELV circuits. p.2 ↗ | Do not connect to non-SELV circuits to avoid damaging the unit. | Plate Heater-Shaker+ |
| The user must maintain a distance of at least 20 cm from the product (Wi-Fi antenna) during installation and operation. p.2 ↗ | Ensure personnel maintain this distance to limit RF exposure. | Plate Heater-Shaker+ |