Agilent Technologies Real-time LC Monitoring, Infinity III Prime Online LC System
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Real-time LC Monitoring, Infinity III Prime Online LC System

Agilent Technologies Real-time LC Monitoring, Infinity III Prime Online LC System — Liquid chromatography.

The 1260 Infinity II Prime Online LC System (G3167AA) is a high-performance GPC/SEC system configured for online reaction monitoring, integrating a 1260 Infinity II Prime LC with a reactor sampling device. It operates at 1.0 mL/min analytical flow, 45 °C column temperature, and 30 psi inlet pressure, with a 5 µL sample volume and 10 mm detector flow cell. This modification is distinguished within the 1260 series by its online sampling interface, enabling automated withdrawal and injection of reaction aliquots for real-time molecular-weight analysis.

Full specifications

Core Performance & Functional Specifications
Analytical Flow Rate 1.0 mL/min
Column temperature 45 °C
Inlet Pressure 30 psi
Sample volume 5 µL
Solvent composition mfr 15 % B
Feed speed mfr 80 rate
Needle wash mfr 3 % FA
Diode array detector wavelength mfr 273 nm
Draw speed mfr 100 μL/min
Wait time mfr 3.6 s
Dilution mfr 1
Mixing time mfr 10 s
Cleaning volume mfr 1.6 mL
Sampling interval mfr 10 min
Regulator pressure mfr 15 psi
Flow meter mfr 0.1 liter/min
Column dimensions mfr 3.0 mm
Dimensions, Weight & Mechanical Design
Detector Flow Cell Path Length 10 mm

Description

Construction and Series Positioning

Behavior Under Load and in Different Media

Installation, Connection, and Control

Integration & Operating Notes

Process Limits
Inlet pressure must not exceed 30 psi

The specified inlet pressure is 30 psi; exceeding this may damage the sampling device or the LC column. Ensure the reactor pressure is regulated to 15 psi at the regulator and that the flow meter indicates 0.1 L/min to maintain proper sampling conditions.

Factory Specification Sheet; Reactor Sampling Device Application Note
Control & I/O
Sampling interval must be synchronized with LC run time

The sampling interval is 10 min, which must be coordinated with the LC analytical run time to avoid overlapping injections. The SRS software must be configured to trigger sampling at the correct interval and to account for the draw speed (100 µL/min), wait time (3.6 s), and mixing time (10 s).

Reactor Sampling Device Application Note
Maintenance
Regular cleaning of sampling device is required

The needle wash uses 3% FA and a cleaning volume of 1.6 mL per cycle. Ensure the wash solution is replenished and the sampling device is cleaned according to the manufacturer's schedule to prevent carryover and clogging.

Reactor Sampling Device Application Note
Electrical
Verify power requirements for the 1260 Infinity II Prime LC

The LC system requires a stable power supply; refer to the 1260 Infinity II Prime LC manual for voltage and frequency specifications. The sampling device may have additional power requirements; ensure all components are connected to a suitable power source.

1260 Infinity II Prime LC System Manual

Selection & Integration Q&A

When should I choose the 1260 Infinity II Prime Online LC System over the 1220 Infinity II LC System or the 1260 Analytical Scale LC Purification System?

Choose the 1260 Infinity II Prime Online LC System when you need automated online sampling from a reactor for real-time GPC/SEC analysis, with a 10 min sampling interval and 5 µL sample volume. The 1220 Infinity II LC System is a general analytical LC without online sampling; the 1260 Analytical Scale LC Purification System is for purification, not online monitoring.

What are the requirements for connecting the reactor sampling device to the 1260 Infinity II Prime Online LC System?

The system requires a reactor sampling device (e.g., Snapdragon Chemistry SRS or Mettler Toledo EasySampler 1210) with a draw speed of 100 µL/min, a regulator pressure of 15 psi, and a flow meter reading of 0.1 L/min. The sampling device must be compatible with the SRS software for control and synchronization with the LC.

What maintenance is required and what accessories are not included in the base delivery?

Regular maintenance includes replenishing the needle wash solution (3% FA) and cleaning the sampling device to prevent carryover. The base system does not include the reactor sampling device (e.g., SRS or EasySampler), which must be purchased separately. Additionally, columns, solvents, and software licenses may be required but are not part of the standard delivery.