Agilent Technologies 4200 TapeStation
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 4200 TapeStation

Agilent Technologies 4200 TapeStation — Automated electrophoresis.

The Agilent 4200 TapeStation is an automated electrophoresis instrument for DNA and RNA sample quality control, using ScreenTape consumables to process up to 16 samples per device with a 1 µL sample requirement. It provides sizing from 35 to 1,000 bp and quantitative range of 25–500 ng/µL, with 16-minute analysis time. This model is distinguished within the Tapestation series by its 96-well plate compatibility and robotics for higher-throughput workflows.

Agilent Technologies Tapestation Instruments Series — Model Comparison & Selection

Model SensitivityDimensions w d hWeight, kgStorage temperature, °CAnalysis time, minApplication
4200 TapeStation Current 543525.5-40...7016—
4150 TapeStation 525417.8-40...7016—
4150 TapeStationDx —————automated electrophoresi
4200 TapeStationDx System —————automated electrophoresi

Full specifications

Core Performance & Functional Specifications
Sensitivity 5 ng/µL
Analysis time 16 min
Dimensions (w × D × H) 435 mm
Kit Stability 6 months
Quantitative precision mfr 10 % CV
Quantitative accuracy mfr 20
Quantitative range mfr 25...500 ng/μL
RINe functional range mfr 25...500 ng/μL
Maximum buffer concentration in sample mfr 200 NaCl
Sample volume required mfr 1 μL
Kit size mfr 112 samples
Sizing range mfr 35...1,000 bp
Safety standards mfr 2 only.
ISM Classification mfr 1
Computer Specifications mfr Checks the computer specifications and configura
Insufficient Mixing mfr Buffer and sample was not sufficiently vortexed.
Sample concentration outside recommended range for application mfr Linear concentration estimation cannot be achiev
Agilent D1000 ScreenTape Assay mfr 2991
Agilent High Sensitivity D1000 ScreenTape Assay mfr 2991
Agilent RNA ScreenTape Assay mfr 2991
Agilent High Sensitivity RNA ScreenTape Assay mfr 2991
Agilent Genomic DNA ScreenTape Assay mfr 2991
Agilent D5000 ScreenTape Assay mfr 2991
Agilent High Sensitivity D5000 ScreenTape Assay mfr 2991
Agilent Cell-free DNA ScreenTape Assay mfr 2991
Needle To Pump mfr Yes Yes
Tape Handling mfr Yes Yes
Sample concentration outside recommended range for application. mfr Linear concentration estimation cannot be achiev
Sample prepared too long before analysis mfr Evaporation or settling may alter the sample to
Incorrect Marker peaks picked up mfr Depending on the sample pattern, wrong peaks can
Clear space around instrument mfr 75 mm
Additional space above instrument for lid mfr 25 cm
Minimum wait before unpacking after shipment mfr 5 h
Vortex mixing speed mfr 2000 rpm
Vortex mixing time mfr 1 minute
Minimum distance from magnets for pacemaker wearers mfr 55 mm
Minimum distance from magnets for sensitive devices mfr 25 mm
Dimensions, Weight & Mechanical Design
Weight 25.5 kg
Electrical, Power & Drive Specifications
Power consumption 50 W
Line voltage mfr 100...240
Line frequency mfr 50
Interfaces, Connections & Control
Interface mfr USB USB
Incorrect protocol used mfr Differences exist between Standard and High Sens
Materials, Media & Process Compatibility
Samples per consumable mfr 16
Operating Conditions, Environment & Protection
Operating Altitude 3000 m
Sound pressure 70 decibel
Storage temperature -40...70 °C
Operating humidity mfr 15...80 condensing
Software Environment mfr Checks the compatibility of TapeStation Controll

Description

Construction and positioning within the series

The 4200 TapeStation (models G2991A and G2991B) is a benchtop automated electrophoresis system that integrates robotics, fluidics, and optical detection to analyze nucleic acid samples. It sits above the 4150 TapeStation in the Tapestation series, adding a 96-well plate format and extended sample capacity. The system uses ScreenTape devices—single-use consumables with 16 lanes each—and processes up to 16 samples per consumable. The instrument includes a piercing needle, electrode cartridge, and a camera-based detection system. It is controlled by the TapeStation Controller software running on a connected laptop.

Behavior under load and assay performance

Installation, connection, and control specifics

The 4200 TapeStation connects to a laptop via USB. The TapeStation Controller software manages instrument communication and includes System Diagnostics with Software Installation Verification, System Verification, and Hardware Troubleshooting test suites. These tests check mechanical, electrical, and optical components using the TapeStation Test Tape (PN 5067-5601). Hardware Troubleshooting takes approximately 30 minutes. The optional liquid handling test requires a foil-sealed 96-well plate with specific water volumes in designated wells and two tube strips. The recommended foil seal for sample plates is PN 5067-5154. If no valid TapeStation system is connected, only Software Installation Verification tests can be run. Tape Handling, Teach Pump-ML, and Teach Pump-ML-Z tests are specific to G2991A and G2991B instruments.

Integration & Operating Notes

Process Limits
Lane insufficiency requires a second ScreenTape within 15 minutes

If the system detects insufficient lanes during a run, a second ScreenTape device with sufficient unused lanes must be added within 15 minutes to finalize the analysis. Keep spare ScreenTape devices on hand and respond promptly to lane-insufficiency messages to avoid losing the run.

User Manual, Software Messages, page 201

Selection & Integration Q&A

When should I choose the 4200 TapeStation over the 4150 TapeStation?

Choose the 4200 when you need 96-well plate compatibility and higher throughput; it processes up to 16 samples per ScreenTape and supports robotics for automated plate handling. The 4150 is limited to 16 samples per run without plate automation. For lower sample numbers, the 4150 may suffice.

What are the requirements for connecting the 4200 TapeStation to a laptop and ensuring reliable communication?

The instrument connects via USB to a laptop running TapeStation Controller software. Ensure the laptop meets the software specifications and that only one instance of the software is running. Use the Windows Device Manager to verify USB communication. If connection fails, check USB cable and port, and restart the software.

What maintenance schedule and consumables are required for the 4200 TapeStation, and what is not included in the base delivery?

Replace the piercing needle before 4480 samples (warning at 3840) and the electrode cartridge annually or when messages appear. ScreenTape devices, Test Tape (5067-5601), foil seals (5067-5154), and sample plates are consumables not included. The Test Tape is required for diagnostics. Schedule maintenance every 6 months for kit stability.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
4200 TapeStation System Manual
Manual ·256 pp.
View PDF