Agilent Technologies Accurate and Reliable GC System with Faster Cycle Times
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Accurate and Reliable GC System with Faster Cycle Times

Agilent Technologies Accurate and Reliable GC System with Faster Cycle Times — Gas chromatography.

The Agilent 8890B GC System is a dual-column, dual-FID gas chromatograph configured for residual solvent analysis per USP <467> and ChP 0861, with an 8697 headspace sampler. It runs a 19.3-minute high-resolution method on 50 m columns and supports helium or nitrogen carrier gas. This modification targets routine compendial residual-solvent testing with fast cycle times.

Agilent Technologies GC Systems Series — Model Comparison & Selection

Model Operating temperature, °CSample volume, milliliterCarrier gasCollision gas nitrogen, mL/minColumn temperature, °CData rate, Hz
Accurate and Reliable GC System with Faster Cycle Times Current —0.00221.58510
Reliable routine analysis gas chromatography system 8...425—————

Full specifications

Core Performance & Functional Specifications
Carrier gas 2
Collision Gas (nitrogen) 1.5 mL/min
Column temperature 85 °C
Data Rate 10 Hz
Electron Energy 70 eV
Oven Ramp 1 Rate 3 °C/min
Oven Ramp Rate 2 50 °C/min
Purge Flow To Split Vent 50 mL/min
Quench Gas (helium) 2.25 mL/min
Run Time 19.3 minutes
Sample volume 1.5 µL
Septum Purge 6 mL/min
Switching Time 4.9 min
Carrier gas and flow rate mfr 1 flow
Tune mfr Etune: atunes.eiex.tune.xml s.eiex.tune.xml
Cleaning interval mfr 350...400 injections
Headspace vial pressurization gas mfr 2
Headspace inject time mfr 0.5 min
Detector mfr FID FID
S/N acceptance criterion, Class 1 (1,1,1-trichloroethane) mfr 5
S/N acceptance criterion, Class 1 (other solvents) mfr 3
S/N acceptance criterion, Class 1 (benzene) mfr 5
S/N acceptance criterion, Class 2A mfr 1
S/N acceptance criterion, Class 2B (1,2-dimethoxyethane and ...) mfr 1.5
RT %RSD, 1,1-Dichloroethene mfr 0.021
Area %RSD, 1,1-Dichloroethene mfr 1.275
RT %RSD, Benzene mfr 0.004
Area %RSD, Benzene mfr 1.833
RT %RSD, Toluene mfr 0.003
Area %RSD, Toluene mfr 0.641
RT %RSD, Methanol mfr 0.047
Area %RSD, Methanol mfr 2.521
Column dimensions mfr 50 μm
Oven ramp 2 hold mfr 1 minute
Vial pressurization gas mfr 2
Inject time mfr 0.5 min
Vial Pressurization Gas mfr 2
Loop Size mfr 1 mL
Vial Size mfr 20 mL
Vial Shaking mfr 8
Vial Equilibration Time mfr 40 min
Inject Time mfr 0.5 min
Vial Fill Pressure mfr 15 psi
Loop Fill Mode mfr Custom Custom
Loop Ramp Rate mfr 40 psi/min
Loop Final Pressure mfr 4 psi
Liner mfr 1 id
CFT Device mfr Unpurged two-way splitter splitter
Tubing mfr 0.32 tubing
FID (Both Channels) mfr 250
A mfr 1
B mfr 1
Split ratio mfr 10
FID hydrogen flow mfr 30 mL/min
FID air flow mfr 300 mL/min
Operating Conditions, Environment & Protection
Fid Temperature 250 °C
Initial Oven Temperature 40 °C
Inlet Temperature 250 °C
Ion Source Temperature 280 °C
Transfer Line Temperature 300 °C
Signal-to-noise, Class 1 (1,1,1-trichloroethane) mfr 5
Signal-to-noise, Class 1 (benzene) mfr 5
Signal-to-noise, Class 1 (other solvents) mfr 3
Signal-to-noise, Class 2A mfr 1
Signal-to-noise, Class 2B (1,2-dimethoxyethane and others) mfr 1.5

Description

Construction and positioning

Behavior under method conditions

Installation, control and operation

Integration & Operating Notes

Process Limits
Carrier gas choice affects method validation

The system is evaluated with both helium and nitrogen carrier gases; switching carrier gas requires revalidation of retention times and acceptance criteria because selectivity and efficiency differ between the two.

Application Note, Residual Solvent Analysis 2025 USP <467> / ChP 0861
Maintenance
Plan inlet maintenance every 350–400 injections

The specified cleaning interval is 350–400 injections; deferring septum/liner maintenance beyond this window risks carryover and degraded S/N, particularly for Class 1 solvents with acceptance criteria of 3–5.

Factory Specification Sheet
Control & I/O
Tune file must match the installed EI source

The method references the tune file atunes.eiex.tune.xml; using a different tune file changes ion-source response and can shift the S/N margins used for Class 1 and Class 2 acceptance.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose this 8890B residual-solvent configuration instead of a sibling GC System model?

Choose this dual-column, dual-FID 8890B with 8697 headspace sampler when the workflow is compendial residual-solvent testing per USP <467>/ChP 0861 with helium or nitrogen carrier gas. Sibling models target flexible/expandable intelligent GC, routine GC analysis, or Micro GC fast chromatography instead.

What adjacent system parts and consumables are required to run the residual-solvent method?

The method requires an Agilent 8697 headspace sampler, DB-Select 624 UI and DB-WAX UI columns (50 m × 0.200 mm × 0.4 µm for the 19.3-min high-resolution method), FID detectors, and either helium or nitrogen carrier gas. The 20 m × 0.100 mm, 0.2 µm column is needed only for the 13.83-min high-speed variant.

Which accessories and consumables are not included in the base delivery?

The base system does not include the application-specific columns (DB-Select 624 UI, DB-WAX UI), the 8697 headspace sampler, carrier/collision/quench gases, or the tune file atunes.eiex.tune.xml; these must be specified separately. The Deans switch for MSTFA diversion is a separate configuration for 7000/7010 GC/TQ, not part of this model.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Residual Solvents Analysis Based on Agilent DB-Select 624 UI Columns
Brochure ·1,5 МБ ·14 pp.
View PDF