Agilent Technologies 8900-triple-quadrupole-icp-ms
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 8900-triple-quadrupole-icp-ms

Agilent Technologies 8900-triple-quadrupole-icp-ms — Atomic spectroscopy.

The Agilent 8900 ICP-QQQ is a second-generation triple-quadrupole ICP-MS with MS/MS capability, distinguished from single-quadrupole siblings (7850, 7900) by tandem mass filtering and from the 9500 by its standard/Advanced Applications configurations. It delivers 700 Mcps/ppm sensitivity and serves trace-element and speciation analysis in complex matrices.

Agilent Technologies ICP MS Instruments Series — Model Comparison & Selection

Model SensitivityMinimum dwell time, msAnalysis time, minData repeatability, %Instrument depth, mmInstrument height, mm
8900-triple-quadrupole-icp-ms Current 7000.1————
7900-icp-ms 109—————
9500-triple-quadrupole-icp-ms 100000.0535.54600530

Full specifications

Core Performance & Functional Specifications
Sensitivity 700 Mcps/ppm
Minimum Dwell Time 0.1 ms
UHMI matrix tolerance mfr 25 solids
Q1 resolution (MS/MS) mfr 0.7 u
Q2 resolution mfr 0.7 u
Dynamic range mfr 11 orders
SPS 4 capacity mfr 360 samples
SPS 6 capacity mfr 540 samples
I-AS rack capacity mfr 89 vials
I-AS small sample volume mfr 0.5 mL
LA-ICP-MS imaging acquisition time mfr 24 hours
Si detection limit (MS/MS, O2) mfr 10 ng/L
S detection limit (MS/MS, O2) mfr 10 ng/L
SiO2 nanoparticle size mfr 50 nm
matrix tolerance mfr 25 solids
S/Si/P detection limit specification mfr 50 ppt
sensitivity increase vs 8800 mfr factor of two two
dynamic range (Advanced/Semiconductor) mfr 11 orders
Y sensitivity (Advanced Applications) mfr 700 Mcps/ppm
Y sensitivity (Semiconductor) mfr 1.2 Gcps/ppm
detection limit (Th) mfr 2 ppq
detection limit (U) mfr 1 ppq
Si detection limit mfr 15 ppt
S detection limit mfr 5 ppt
background equivalent particle diameter mfr 25 nm
Electrical, Power & Drive Specifications
Plasma RF frequency mfr 27 MHz
Operating Conditions, Environment & Protection
background noise specification mfr 0.2 cps
Standards, Safety & General Data
dynamic range (Standard) mfr 10 orders
CeO+/Ce+ mfr 1

Description

Construction and series positioning

The 8900 ICP-QQQ is the second-generation triple-quadrupole instrument in Agilent's ICP-MS line. Unlike the single-quadrupole 7850 and 7900, it uses tandem mass spectrometry (MS/MS) with a reaction cell between two quadrupole mass filters, enabling interference removal by collision/reaction chemistry. Within the series it sits below the 9500 ICP-QQQ; the 8900 is offered in a Standard configuration and a #100 Advanced Applications configuration, the latter intended for demanding applications requiring the full MS/MS capability set.

Performance behavior

The verified sensitivity is 700 Mcps/ppm. No curve points, detection limits, or mass-range data were supplied, so no further performance claims are made here. The MS/MS mode is the defining operating behavior: precursor ion selection, reaction/collision in the cell, and product ion selection suppress polyatomic and isobaric interferences that single-quadrupole instruments cannot resolve.

Installation, control and operation

Document 5991 6900EN covers the instrument, its MS/MS capability, software, and options including SPS 4/SPS 6 and I-AS autosamplers, ADS, AVS MS, and a SEMI S2 compliance kit. Document 5991 6942EN distinguishes the Standard and #100 Advanced Applications configurations. No electrical, gas-supply, cooling, or dimensional specifications were verified; these must be taken from the full site-preparation guide before installation.

Selection & Integration Q&A

When should I choose the 8900 ICP-QQQ instead of the 7850 or 7900 single-quadrupole models?

Choose the 8900 when MS/MS interference removal is required for complex matrices or speciation; the 7850 and 7900 are single-quadrupole instruments without the tandem cell architecture. If the application needs the higher-tier 9500 platform, the 8900 is the intermediate triple-quadrupole choice.

What adjacent system components and accessories must be planned for beyond the base instrument?

Autosamplers (SPS 4, SPS 6, I-AS), ADS, AVS MS, and the SEMI S2 compliance kit are options, not base delivery. Site preparation must also cover cell gases and lab services, but exact electrical and gas requirements are not in the supplied digests and must be taken from the full site-prep guide.

What mounting and environmental limits apply to the 8900 installation?

The supplied document digests contain no verified dimensional, weight, cooling, or environmental specifications. Installation limits must be obtained from the complete site-preparation documentation before committing lab space; no values are asserted here to avoid unsupported claims.

What maintenance schedule and consumables apply, and what is not included in base delivery?

No maintenance intervals or consumable lists are present in the supplied digests. Autosamplers, ADS, AVS MS, and the SEMI S2 kit are explicitly options. Service intervals, cones, tubing, and tuning solutions must be confirmed from the full instrument manual, not assumed from this card.