Agilent Technologies AA-Duo
Agilent Technologies AA-Duo — Atomic spectroscopy.
Graphite furnace AAS method card for cadmium in water, using an UltrAA lamp at 4 mA, 228.8 nm, 0.5 nm slit, 10 µL sample and 5 µL modifier. The temperature program runs ash at 568 °C (sample) / 570 °C (standard) and atomization at 1577 °C (sample) / 1533 °C (standard), with 2100 °C as the atomization temperature setpoint. Three replicates yield 101.8 % QC recovery against a 2.50 Cd standard.
Full specifications
| Wavelength range | 228.8 nm |
| Elements | Cd |
| Fuel Flow | 2 L/min |
| Inert Gas Purity | 99.99 % |
| Modifier Volume | 5 μL |
| Qc Recovery | 101.8 % |
| Sample volume | 10 μL |
| Matrix mfr | Water Water |
| Chemical modifier mfr | 4 |
| Instrumentation mfr | 240 AAS |
| Hollow Cathode Lamp mfr | Cadmium Cadmium |
| Graphite tube platform mfr | Omega Omega |
| Mode mfr | Emission Emission |
| Chemical modifier volume mfr | 1 g/L |
| Bandpass mfr | 0.2 nm |
| D2 correction mfr | No No |
| Replicates mfr | 3 |
| Lamp type mfr | UltrAA lamp lamp |
| Slit Width | 0.5 nm |
| Lamp Current | 4 mA |
| Ash Temperature (sample) | 568 °C |
| Atomization Temperature | 2100 °C |
| Atomize Temperature (sample) | 1577 °C |
| Ash Temperature (standard) | 570 °C |
| Atomize Temperature (standard) | 1533 °C |
| Standard Solution | 2.50 Cd |
| Standard Reference Material mfr | 1 Cd |
Description
Method architecture
- Deterministic Cd-in-water program: 10 µL sample, 5 µL chemical modifier, 3 replicates.
- Optical chain: Cd UltrAA lamp at 4 mA, 228.8 nm, 0.5 nm slit, emission mode, D2 correction off.
- Calibration anchored to a 2.50 Cd standard; QC recovery 101.8 %.
Thermal behavior
- Ash stage: 568 °C for sample, 570 °C for standard.
- Atomization: 1577 °C sample, 1533 °C standard, against a 2100 °C atomization setpoint.
- The sample/standard split in both stages means the program is not a single shared temperature profile; each must be entered as its own stage value.
Operation
- Inert gas purity 99.99 %; fuel flow 2 L/min.
- Graphite tube platform: Omega; modifier volume 5 µL (1 g/L chemical modifier).
- No document digests were supplied, so installation, control and maintenance specifics cannot be stated beyond the parameters above.
Integration & Operating Notes
The method is specified at 99.99 % inert gas purity. Lower-purity gas introduces oxidant and trace-metal background into the graphite furnace, which is not corrected here because D2 correction is switched off.
The graphite tube with Omega platform is the atomizer consumable for this method. Tube condition directly affects the 1577/1533 °C atomization signals and the 101.8 % recovery; replacement tubes must be re-conditioned before the QC check is trusted.
Selection & Integration Q&A
Pick it for cadmium in water only: 228.8 nm, Cd UltrAA lamp at 4 mA, 0.5 nm slit, 10 µL sample plus 5 µL modifier, 2.50 Cd standard. For other elements or matrices the lamp, wavelength, modifier and temperature stages all change, so this program does not transfer.
Inert gas at 99.99 % purity and a 1 g/L chemical modifier dosed at 5 µL per 10 µL sample. D2 correction is off, so background control relies on the modifier and gas quality rather than on continuum correction.
The verified facts give no duty-cycle, ambient or installation limits, and no manual digest was supplied. Only the analytical parameters (temperatures, gas purity, volumes, lamp settings) are confirmed; environmental limits must be taken from the instrument manual, not this card.
A Cd UltrAA lamp, graphite tubes with Omega platform, 1 g/L chemical modifier, 2.50 Cd standard and 99.99 % inert gas. None of these are stated as part of base delivery; the tube in particular is a wear item that must be stocked and re-conditioned.