Agilent Technologies blood alcohol analysis, BAC analysis
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies blood alcohol analysis, BAC analysis

Agilent Technologies blood alcohol analysis, BAC analysis — Gas chromatography with Motor power 2.35 kW.

The Agilent blood alcohol analysis (BAC) configuration pairs an 8890 GC with a split/splitless inlet and dual flame ionization detectors with an 8697 headspace sampler, using DB-BAC1/DB-BAC2 Ultra Inert columns. It is a dedicated forensic BAC system, distinguished within the 8890 series by headspace sampling and dual-FID confirmation, intended for ethanol and volatile congener quantification in blood samples.

Motor power
2.35 kW

Full specifications

Core Performance & Functional Specifications
Motor power 2350 W
Retention Time, Ethanol 1.348 min
Retention Time, T-butanol 1.953 min
Vial Volume 20 mL
GC system mfr 8890
Headspace sampler mfr 8697
Detector mfr dual flame ionization detector detector
Sample volume in headspace vial mfr 0.5 mL
Number of sequential injections mfr 12
Resolution, methanol mfr 2.7
Resolution, acetaldehyde mfr 7.9
Resolution, ethanol mfr 2.0
Resolution, isopropanol mfr 7.2
Resolution, t-butanol mfr 5.2
Resolution, propanal mfr 1.3
Resolution, n-propanol mfr 2.0
Resolution, acetone mfr 3.8
Resolution, acetonitrile mfr 3.9
Resolution, 2-butanol mfr 3.0
Resolution, ethyl acetate mfr 12.4
Resolution, 2-butanone mfr 4.1
Dimensions, Weight & Mechanical Design
Depth 54 cm
Height 50 cm
Weight 50 kg
Width 54 cm
GC liner inner diameter mfr 2 mm
Electrical, Power & Drive Specifications
Frequency 50 % Hz
Power Outlet Current 20 Amp
Line voltage mfr 120 VAC
Current rating mfr 19.6 amps
Operating Conditions, Environment & Protection
Maximum Altitude 4615 m
Operating humidity range mfr 5...95
Storage humidity range mfr 5...95
Standards, Safety & General Data
Internal standard concentration mfr 0.1

Description

Construction and positioning

Analytical behavior

Installation and operation

Integration & Operating Notes

Electrical
Dedicated 20 A branch circuit required for 120 V configuration

The recorded current rating of 19.6 A at 120 VAC leaves almost no margin on a 20 A outlet; the site preparation checklist should be followed to confirm breaker, wiring and receptacle sizing before installation, and the 2350 W figure should be cross-checked against the actual configuration ordered.

8890 GC site preparation checklist D0007040 Rev. 3.0; Factory Specification Sheet
Environment
Laboratory ambient must stay within 15–35 °C and 5–95 % humidity

Operation outside the stated temperature and humidity ranges can affect retention-time stability and detector response; the site should be assessed for HVAC capability and altitude (maximum 4615 m) before the system is installed.

Factory Specification Sheet
Process Limits
Headspace sampling fixes vial and sample volumes

The method uses 20 mL headspace vials with 0.5 mL sample volume; deviating from these volumes changes the phase ratio and can shift the reported retention times and resolutions, so sample preparation must follow the application note exactly.

Application Note 5994-3126EN
Maintenance
Consumables and standards are not part of the base system

Columns (DB-BAC1/DB-BAC2 Ultra Inert), liners, septa, headspace vials and calibration standards carry their own part numbers in the application note and must be ordered separately; the 2 mm liner is specified for this method.

Application Note 5994-3126EN

Selection & Integration Q&A

When should this BAC configuration be chosen instead of a standard 8890 GC package?

Choose this configuration when the workflow is forensic blood alcohol testing with headspace introduction and dual-FID confirmation; a standard 8890 GC without the 8697 headspace sampler and BAC columns is suited to general-purpose analysis instead.

What power and site services must be in place before installing the system?

The 8890 GC site preparation checklist covers 120 VAC Americas or 220/230/240 VAC configurations; the recorded 120 VAC/19.6 A rating requires a 20 A outlet, and the checklist also addresses the data system PC and power cord options.

Which items must be ordered separately from the base system?

The DB-BAC1/DB-BAC2 Ultra Inert columns, 2 mm GC liner, 20 mL headspace vials, septa and calibration standards are listed with their own part numbers in the application note and are not included in the base delivery.