Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 1260 Infinity II Prime LC System

Agilent Technologies 1260 Infinity II Prime LC System — Liquid chromatography with Flow rate 0.005 liter/minute, Pumping speed 3×10⁻⁴ m³/h.

The 1260 Infinity II Prime LC System is a modular HPLC/UHPLC platform built around the Flexible Pump G7104C, delivering up to 5 mL/min at pressures up to 800 bar. It distinguishes itself within the InfinityLab series by covering both HPLC and UHPLC methods on one system, while the 1290 Bio variant extends pressure to 1300 bar for biocompatible UHPLC. It serves as the front-end separation engine in analytical workflows, feeding detectors such as the 1260 Infinity II DAD WR (190–950 nm, 120 Hz).

Flow rate
0.005 liter/minute
Pumping speed
3×10⁻⁴ m³/h

Full specifications

Core Performance & Functional Specifications
Flow rate 5 mL/min
Pumping speed 5 mL/min
Wavelength range 190...950 nm
Maximum Data Acquisition Rate 120 Hz
Maximum Individual Bottle Volume 2 L
Maximum pressure 800 bar
Pressure limit 800 bar
Solvent Cabinet Maximum Volume 8 L
Stroke Automatic
Nominal inline filter pore size mfr 0.3 μL
Compressibility mfr Use solvent types types
Composition mfr 95 %A
Stoptime mfr as pump pump
Timetable mfr 2.5 %B
Posttime mfr 1 min
Autobalance mfr Prerun Prerun
G7120-60007 (Bottle Head Assembly, long (1730 mm)) mfr Solvent Bottle Bottle
G5664-68712 (Analytical tubing kit 0.25 mm i.d. PTFE-ESD) mfr Detector Detector

Description

Construction and series positioning

Behavior under load and media

Installation, connection, and control

Integration & Operating Notes

Process Limits
Do not exceed 800 bar maximum pressure

The 1260 Infinity II Prime LC System is rated for a maximum pressure of 800 bar. Operating above this limit risks damage to the pump, column, and flow path components. For methods requiring pressures up to 1300 bar, the 1290 Infinity III Bio Online LC System must be used instead.

Features, page 8
Cooling & Media
Avoid DMF with PVDF leak sensor

Dimethylformamide (DMF) is incompatible with the PVDF leak sensor and causes corrosion, which can lead to false leak detection or sensor failure. Mobile phases containing DMF must be avoided; if used, the leak sensor must be inspected regularly for corrosion.

Leak Sensor, page 45
Electrical
Set DIP switches before LAN configuration

The rear DIP switches determine the network address mode: all switches DOWN = default IP 192.168.254.11; switch 4 UP = DHCP; switch 5 UP = STORED address. These must be set before attempting LAN configuration, which should be executed from the module with direct LAN connection to the controller software, usually the detector with the highest data rate.

Instrument Configuration, page 48; General Information on LAN Configuration, page 47
Maintenance
Back-flush inline filter weekly

The optional inline filter (part number 5067-5407, nominal pore size 0.3 µm) protects the system from blockages caused by buffers or solvent precipitation. It must be back-flushed weekly to maintain performance and prevent pressure buildup.

Optional Inline Filter, page 77
Maintenance
Exchange water-based solvents daily

Water-based solvents must be exchanged daily and selected to be used up within 1–2 days to prevent algae growth and contamination. Use only HPLC-grade solvents and water filtered through 0.2 µm filters.

How to Deal With Solvents, page 77
Process Limits
Use lower channels for buffers and flush regularly

Buffer solutions should be placed on lower channels (A and/or D) of the 4CGV gradient valve. All 4CGV channels must be flushed regularly with water to remove salt deposits and prevent precipitation and blockage.

Select Channels for Gradient Valve (4CGV), page 77
Control & I/O
Set ERI timeout to match sample delivery response

The ERI interface timeout (in minutes) defines how long the software waits for an answer from the sample delivery device. If no answer is received, the sample is marked as missed and the software continues. Set the timeout according to the hardware response time to avoid missed samples.

Set Up the ERI Interface, page 67
Mounting
Use recommended waste container and routing

The recommended waste container is 5043-1221 (6 L waste can with 1 Stay Safe cap GL45 with 4 ports). Waste tubes must not immerse in the liquid to prevent back-siphonage. Ensure leak plane outlets are unobstructed and waste tubes are correctly routed to allow leak detection.

Waste Concept, page 43; Handling Leak and Waste, page 40

Selection & Integration Q&A

When should I choose the 1260 Infinity II Prime LC System over the 1290 Infinity III Bio Online LC System?

Choose the 1260 Infinity II Prime LC System for HPLC and UHPLC methods requiring pressures up to 800 bar and flow rates up to 5 mL/min. Select the 1290 Infinity III Bio Online LC System when pressures up to 1300 bar or biocompatible flow paths are needed, such as for bio-molecule analysis or high-pressure UHPLC.

What are the capillary and tubing requirements for connecting the 1260 Infinity II Prime LC System modules?

For 1260 Infinity III systems, use capillary 5500-1246 (0.17 mm x 500 mm SI/SI) from pump to sampler and 5500-1191 (Quick Turn 0.12 mm x 280 mm) from column/MCT valve to detector. These ensure correct flow path connections and minimal delay volume. For 1290 Bio systems, MP35N capillaries are required.

What are the mounting and environment considerations for the leak sensor and waste management?

The leak plane with PVDF leak sensor stops the system when liquid is detected. Ensure leak plane outlets are unobstructed and waste tubes are correctly routed. Do not use DMF as it corrodes the sensor. Use waste container 5043-1221 (6 L) and do not immerse waste tubes in liquid to prevent back-siphonage.

What maintenance schedule and accessories are required for the inline filter and solvent handling?

The optional inline filter (5067-5407, 0.3 µm) must be back-flushed weekly. Water-based solvents must be exchanged daily and used within 1–2 days; use HPLC-grade solvents and water filtered through 0.2 µm filters. The inline filter is not included in base delivery and must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
1260 Infinity III Prime Online LC and 1290 Infinity III Bio Online LC System Manual
Manual ·146 pp.
View PDF