Agilent Technologies 5977B GC MSD
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 5977B GC MSD

Agilent Technologies 5977B GC MSD — Gas chromatography.

The 5977B GC MSD is a single-quadrupole mass selective detector for 8890, 9000, 8860 and 7890 GCs, with a 41 × 30 × 54 cm housing and 39 kg mainframe. This turbo-pump configuration supports 1–2 mL/min helium column flow, up to 0.53 mm columns and field CI upgrade. It serves as the mass-selective detector in capillary GC/MS systems.

Full specifications

Core Performance & Functional Specifications
Foreline Reading 34.7 mTorr
Gas Flow 2 mL/min
Ion Vacuum Gauge Reading 3.18 Torr
Maximum Column Id 0.53 mm
Maximum Recommended Gas Flow 1.5 mL/min
Optimal He Column Flow 1 mL/min
Max column id mfr 0.25 mm
CI capability mfr No No
Inert ion sources available mfr Yes Yes
GC compatibility mfr 9000 Series
Foreline pumps available mfr Pfeiffer Duo Duo
DIP capability (3rd party) mfr Yes Yes
High vacuum pump mfr Diffusion Diffusion
CI capability‡ mfr No No
DIP** capability (3rd party) mfr Yes Yes
Methane/PCI mfr 17
Methane/NCI mfr 17
Isobutane/PCI mfr 39
Isobutane/NCI mfr 17
Ammonia/PCI mfr 18
Ammonia/NCI mfr 17
Minimum column velocity during conditioning mfr 30 cm/s
Carrier gas purity for conditioning mfr 99.9995 better
Column conditioning hold time mfr 30 minutes
Column extension past ferrule mfr 4...6 mm
Column extension for XTR/SS/Inert/CI mfr 1 mm
Final nut tightening mfr 4...1 turn
Minimum pumpdown time before verification mfr 60 minutes
Injection volume for sensitivity check mfr 1 μL
Thermal equilibrium wait after pumpdown mfr 2 hours
Bakeout time before CI tuning mfr 2 hours
Dimensions, Weight & Mechanical Design
Foreline Pump Weight 11 kg
Weight 39 kg
Housing height mfr 41 cm
Housing width mfr 30 cm
Housing depth mfr 54 cm
MSD housing height mfr 41 cm
MSD housing width mfr 30 cm
MSD housing depth mfr 54 cm
Diffusion pump mainframe weight mfr 39 kg
EI/CI turbo pump mainframe weight mfr 46 kg
Electrical, Power & Drive Specifications
Primary fuse rating mfr 12.5 V
EMV threshold for CI autotune concern mfr 2600 V
Operating Conditions, Environment & Protection
Sound Pressure Lp 70 dB
Standards, Safety & General Data
Standard EI turbo pump mainframe weight mfr 44 kg

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Keep total gas flow at or below 4 mL/min for turbo models

Total flow includes column flow plus reagent gas flow. The maximum recommended total gas flow into the MSD is 4 mL/min for turbo-pump models; operation above the 6.5 mL/min absolute maximum causes expected degradation of spectral performance and sensitivity. Size column and reagent gas flows together so the sum stays within the recommended limit.

User Manual, Table 3, page 18
Process Limits
Ammonia reagent gas supply must stay below 5 psig

Higher pressures can cause ammonia to condense to liquid. Keep the tank upright below the flow module and coil tubing into vertical loops. A 5% ammonia/95% helium or 5% ammonia/95% methane mixture is frequently used to minimize negative effects of ammonia while achieving good CI.

User Manual, To Use Other Reagent Gases, page 158
Electrical
Use exact T12.5A, 250 V primary fuses, two required

Primary fuses are T12.5A, 250 V (part 2110-1398), two required. Use exact replacement fuses; the most likely cause of primary fuse failure is a foreline pump problem, so investigate the pump before replacing fuses.

User Manual, To Replace the Primary Fuses, page 32

Selection & Integration Q&A

When should I choose this turbo-pump 5977B modification instead of a diffusion-pump variant in the series?

Choose the turbo configuration when you need 1–2 mL/min optimal helium flow, up to 4 mL/min recommended total flow, columns up to 0.53 mm ID, or field CI upgradeability. Diffusion-pump models are limited to 1 mL/min optimal, 1.5 mL/min recommended, 0.25 mm ID and no CI capability.

What column and gas-flow conditions must adjacent GC parts satisfy for this MSD?

The GC must deliver 1–2 mL/min optimal helium column flow and keep total gas flow (column plus reagent gas) at or below 4 mL/min recommended, 6.5 mL/min absolute maximum. Columns up to 0.53 mm ID (30 m) are supported; install via self-tightening or standard column nut with the GC/MSD interface tip seal.

What maintenance schedule and consumables are required beyond base delivery?

Check foreline pump fluid weekly and replace every 6 months (every 3 months with ammonia CI); diffusion pump fluid and dry foreline pump seals annually; IDP-3 tip seal annually; leaks monthly; calibration vials every 6 months. Primary fuses are T12.5A, 250 V (part 2110-1398), two required; Apiezon L high vacuum grease is needed for the analyzer side plate O-ring.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
GC/MSD System Operating Manual
Manual ·254 pp.
View PDF