Agilent Technologies 7250 GC Q TOF
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 7250 GC Q TOF

Agilent Technologies 7250 GC Q TOF — Gas chromatography with Noise level 70 dB.

The Agilent 7250 GC Q-TOF is a quadrupole time-of-flight GC/MS mainframe with three turbomolecular pumps, an LE-EI ion source (optional Axial CI), and either a DS202 rotary vane or IDP dry scroll foreline pump. It is distinguished within the 7250 series by its accurate-mass TOF analyzer and collision cell, supporting EI and CI operation with total gas flow into the MS limited to 2.0 mL/min recommended (2.4 mL/min maximum). It serves as the mass-selective detector in a GC/MS system for high-resolution accurate-mass analysis.

Noise level
70 dB

Full specifications

Core Performance & Functional Specifications
Noise level 70 dB
Gas Flow 2.4 mL/min
Maximum Column Id 0.32 mm
Maximum Recommended Gas Flow 2.0 mL/min
Quad Vacuum 3.50 Torr
Rough Vacuum 1.25 mTorr
Tof Vacuum 2.61 Torr
Flight tube extension over top of instrument mfr 136 cm
Optimal carrier gas flow mfr 1.0...1.5 mL/min
Collision cell nitrogen gas flow mfr 1 mL/min
Collision cell helium gas flow mfr 4 mL/min
High-vacuum pump 1 mfr Split-flow turbo turbo
High-vacuum pump 2 mfr Turbo Turbo
High-vacuum pump 3 mfr Turbo Turbo
Optimal carrier gas flow, mL/min mfr 1.0...1.5
Collision cell Nitrogen gas flow, mL/min mfr 1
Collision cell Helium gas flow, mL/min mfr 4
MS Source mfr 230 dependent.
Complete assembly mfr Axial CI Ion Source Assy Assy
Flight tube extension above instrument mfr 136 cm
Ammonia supply maximum pressure mfr 5 psig
Turbo pump speed at 10–15 min mfr 80
Turbo pump final speed mfr 95
Thermal equilibration wait mfr 2 hours
Dimensions, Weight & Mechanical Design
Depth 92 cm
Foreline Pump Weight 25 kg
Height 53.5 cm
Instrument Depth 92 cm
Instrument Height 53.5 cm
Instrument Width 69 cm
Width 69 cm
Instrument mainframe weight mfr 159 kg
Mainframe weight mfr 159 kg
Interfaces, Connections & Control
GC/MS interface operating range mfr 250...350 °C
Operating Conditions, Environment & Protection
Sound pressure 70 dB
PCI mfr 300 °C
NCI mfr 150 °C
Quadrupole maximum setpoint mfr 200 °C
Ion source maximum setpoint mfr 350 °C

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection, and operation

Integration & Operating Notes

Process Limits
Total gas flow into the MS must not exceed 2.4 mL/min

Column flow plus collision cell gas flow should stay at or below 2.0 mL/min recommended; 2.4 mL/min is the maximum before spectral performance and sensitivity degrade. Select column length and id so that combined flows remain within limits during tuning and acquisition.

User Manual, Chapter 2, Table 2, page 24
Process Limits
Hydrogen carrier gas is prohibited; use ≥99.9995% pure carrier

Carrier gas must be at least 99.9995% pure. Hydrogen gas use is specifically prohibited; do not plumb hydrogen to the system. Use high-purity helium or another approved carrier.

User Manual, Chapter 2, page 25
Process Limits
Ammonia supply pressure must stay below 5 psig

In CI mode with ammonia, supply pressure must be less than 5 psig; higher pressure can cause ammonia to condense from gas to liquid. Install a pressure regulator limiting ammonia delivery pressure below 5 psig.

User Manual, Chapter 4, page 66
Control & I/O
Do not heat GC zones before carrier gas flow is on

Do not turn on any GC heated zones until carrier gas flow is on; heating a column without carrier gas flow damages the column. This interlock protects both column and MS.

User Manual, Chapter 3, page 56
Control & I/O
Turbo speed must reach 80% within 10–15 minutes during pump down

During pump down, turbo pump speed should reach 80% within 10 to 15 minutes and eventually 95%; otherwise MS electronics shut off the foreline pump and a power cycle is required. Monitor pump speed; failure to reach speed indicates a vacuum problem requiring troubleshooting.

User Manual, Chapter 3, page 56
Control & I/O
Wait 2 hours after 'Okay to run' for thermal equilibrium

Turn on carrier gas flow after vacuum pumps have been running for 15 minutes; after 'Okay to run' appears, wait 2 hours for thermal equilibrium before acquiring data. Data acquired before thermal equilibrium may not be reproducible.

User Manual, Chapter 3, page 56
Maintenance
Ballast foreline pump daily when running ammonia ≥5 h/day

When running ammonia five or more hours per day, the foreline pump must be ballasted (flushed with air) for at least one hour per day, and the MS must be purged with methane after flowing ammonia. This minimizes damage to pump seals and removes residual ammonia.

User Manual, Chapter 4, page 66
Maintenance
Check calibration vial every 6 months

Check the calibration vial every 6 months to ensure tune and mass calibration reference is available and valid.

User Manual, Chapter 5, page 82
Mounting
Analyzer side plate thumbscrew must be finger-tight only

The top thumbscrew on the analyzer side plate must be fastened finger-tight; do not over-tighten as this can cause air leaks. Collision cell chamber top plate shipping brackets must remain fastened during normal operation.

User Manual, Chapter 1, page 17
Mounting
Interface tip seal must be in place for LE-EI ion source

The interface tip seal must be in place for the LE-EI ion source; the analyzer should close all the way with no resistance except spring tension from the tip seal. Misalignment or missing seal causes leaks or damage; align by wiggling the side plate on its hinge.

User Manual, Chapter 2, pages 33-34

Selection & Integration Q&A

When should I choose the 7250 Q-TOF with the standard LE-EI source versus the optional Axial CI source?

Choose the LE-EI source for routine EI operation (standard or low-energy). Specify the optional Axial CI ion source only if chemical ionization with isobutane, ammonia, or carbon dioxide is required; CI mode adds reagent gas handling and maintenance procedures.

What are the carrier gas purity and column compatibility requirements for the 7250 Q-TOF?

Carrier gas must be at least 99.9995% pure; hydrogen is prohibited. Maximum column id is 0.32 mm at 30 m length. Optimal carrier gas flow is 1.0–1.5 mL/min, and total gas flow into the MS (column plus collision cell) must not exceed 2.0 mL/min recommended (2.4 mL/min maximum).

What maintenance schedule and consumables are required for the 7250 Q-TOF, and what is not included in base delivery?

Check foreline pump oil weekly; replace rotary vane pump oil and oil mist filter every 6 months (or IDP tip seals yearly). Check calibration vial every 6 months. Columns, carrier gas, collision cell gases, and reagent gases are not included and must be supplied separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
7250 Accurate-Mass Q-TOF GC/MS Operating Manual
Manual ·128 pp.
View PDF