Agilent Technologies Cary 660 FTIR

Agilent Technologies Cary 660 FTIR — Molecular spectroscopy with Flow rate 10 liter/minute.

The Cary 660 is a mid-series FTIR spectrometer in the Cary 600 Series, sharing the penthouse architecture with the 670 and 680 while the 610 microscope is a separate accessory. It uses a 632.8 nm continuous laser (<1 mW embedded, <600 μW accessible) and requires dry purge gas at 10 L/min. It serves as the optical engine for routine molecular spectroscopy, with MCT/InSb detectors needing liquid-nitrogen cooling.

Flow rate
10 liter/minute

Full specifications

Core Performance & Functional Specifications
Flow rate 10 L/min
Wavelength range 632.8 nm
Dewar Equilibrium Time 20 minutes
Maximum Accessible Laser Power 600 μW
Maximum Beam Divergence (embedded Laser) 1.3 milliradians
Minimum Side Clearance 50 mm
Warm-up Time To Operating Temperature 30 minutes
Wavelength (embedded Laser) 632.8 nm
Regulator required above source pressure mfr 60 psi
Laser pulse duration mfr Continuous Continuous
Pulse duration (embedded laser) mfr Continuous Continuous
Initial Dewar fill volume mfr 500 milliliters
Dewar top-off volume mfr 200 milliliters
Minimum rear clearance mfr 150 mm
Purge gas source pressure regulator threshold mfr 60 psi
Pulse duration mfr Continuous Continuous
Agilent Cary 600 Series FTIR spectrometer mfr 4 equivalent
Agilent Cary 610 FTIR microscope (typical) mfr 1 equivalent
High-sensitivity MCT mfr Liquid nitrogen nitrogen
Linearized high-sensitivity MCT mfr Liquid nitrogen nitrogen
Linearized broadband MCT mfr Liquid nitrogen nitrogen
Cooled DLaTGS mfr Peltier Peltier
Top — Power/Scan mfr Flashing green green
Bottom — A/D over-range mfr Flashing blue blue
Recommended purge flow rate for spectrometer and microscope mfr 10 L/min
Regulator threshold pressure mfr 60 psi
Dewar fill volume mfr 475 milliliters
Dewar top-up volume mfr 200 milliliters
Rear clearance mfr 150 mm
Electrical, Power & Drive Specifications
Maximum Accessible Power (embedded Laser) 1 mW
Fuse type (spectrometer) mfr 4 equivalent
Fuse type (microscope) mfr 1 equivalent
Fuse type mfr 1 equivalent
Materials, Media & Process Compatibility
Liquid Nitrogen Temperature -196 °C
Operating Conditions, Environment & Protection
Purge Gas Dew Point -70 °C

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Purge gas must be dried to -70 °C dew point or use dry nitrogen

Moisture in the purge supply can compromise optical performance and detector cooling. Use water-free air dried to a dew point of -70 °C or dry nitrogen, and monitor flow with a flow meter; a second flow meter is recommended for independent microscope control.

User's Guide, Section 2, Purge gas, page 22
Process Limits
Install a pressure regulator if purge source exceeds 60 psi

Source pressures above 60 psi (420 kPa) require a regulator to protect the instrument from overpressure. Maximum purge flow is 10 L/min; 10 L/min is recommended when purging both spectrometer and microscope.

User's Guide, Section 2, Purge gas, page 22
Mounting
Maintain 50 mm side and 150 mm rear clearance for air circulation

Insufficient clearance can cause overheating and impair instrument performance. Ensure at least 50 mm (2 in.) on the sides and 150 mm (6 in.) at the rear of the system.

User's Guide, Section 3, Unpacking, page 28
Electrical
Replace fuses only with specified T4 AH 250 V type

The spectrometer uses a T4 AH 250 V fuse (IEC601272-2 Sheet 5, 5×20 mm, Littelfuse 0215004 or equivalent). Do not substitute the lower-rated microscope fuse (1FS1 T2.0 AH 250 V) in the spectrometer.

User's Guide, Section 2, Table 1, page 22
Control & I/O
Use LED states to confirm readiness and diagnose over-range

Solid green power/scan LED indicates initialized and ready; flashing green indicates scanning; solid yellow indicates initialization failure. A flashing blue A/D over-range LED means the A/D exceeds its maximum value; reduce sensitivity or increase beam attenuation.

User's Guide, Section 4, Table 3, page 38
Maintenance
Clean exterior with water or mild detergent only

Use a soft cloth dampened with water or mild detergent; do not use organic solvents or abrasive cleaning agents, which can damage instrument surfaces and markings. Routine maintenance procedures are in the Resolutions Pro Help.

User's Guide, Section 5, Cleaning, page 48

Selection & Integration Q&A

When should I choose the Cary 660 instead of a sibling model like the 670 or 680?

The Cary 660 is the base spectrometer in the 600 Series; the 670 and 680 are higher-tier siblings. Choose the 660 for routine FTIR without the additional capabilities of the 670/680. The 610 microscope is a separate accessory that connects via the penthouse EXT DET ports.

What are the requirements for connecting a Cary 610 microscope to the Cary 660?

The microscope detector output connects to the spectrometer EXT DET 1 or EXT DET 2 penthouse connector. A second purge flow meter is recommended for independent microscope control. The microscope uses its own 1FS1 T2.0 AH 250 V fuse, not the spectrometer fuse.

What maintenance and consumables are required, and what is not included in base delivery?

Liquid nitrogen (500 mL initial plus 200 mL top-off) is required for MCT/InSb detectors but not supplied. Purge gas (dry air or nitrogen) and a pressure regulator (if source >60 psi) are user-supplied. Clean exterior with water or mild detergent only; routine maintenance is in the Resolutions Pro Help.