No photo available yet

Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Cary 670 FTIR

Agilent Technologies Cary 670 FTIR — Molecular spectroscopy.

The Cary 670 FTIR is a laboratory Fourier-transform infrared spectrometer with an embedded 632.8 nm continuous helium-neon laser (maximum accessible power 600 µW, 1 mW embedded) used for interferometer referencing. It requires a dry purge gas supply (dew point -70 °C) at up to 10 L/min and 30 minutes warm-up to operating temperature. Its 20-minute dewar equilibrium time and -196 °C liquid-nitrogen dewar support cooled detector operation.

Full specifications

Core Performance & Functional Specifications
Dewar Equilibrium Time 20 minutes
Maximum Accessible Laser Power 600 μW
Maximum Beam Divergence (embedded Laser) 1.3 milliradians
Warm-up Time To Operating Temperature 30 minutes
Wavelength (embedded Laser) 632.8 nm
Purge flow rate (maximum) mfr 10 L/min
Purge flow rate (recommended) mfr 10 L/min
Regulator required above source pressure mfr 60 psi
Laser pulse duration mfr Continuous Continuous
Pulse duration (embedded laser) mfr Continuous Continuous
Initial Dewar fill volume mfr 500 milliliters
Dewar top-off volume mfr 200 milliliters
Minimum side clearance mfr 50 mm
Minimum rear clearance mfr 150 mm
Electrical, Power & Drive Specifications
Maximum Accessible Power (embedded Laser) 1 mW
Fuse type (spectrometer) mfr 4 equivalent
Fuse type (microscope) mfr 1 equivalent
Operating Conditions, Environment & Protection
Purge Gas Dew Point -70 °C

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Process Limits
Regulator required when source pressure exceeds 60 psi

The purge inlet must not see source pressure above 60 psi without a downstream regulator; direct connection to a high-pressure cylinder or house line can overpressure the instrument's internal purge path. Install the regulator at the source and set delivery to the spectrometer's inlet requirement.

Factory Specification Sheet
Cooling & Media
Fill the dewar with 500 mL initially and top off with 200 mL

Liquid nitrogen at -196 °C is required for cooled detector operation. The initial fill volume is 500 mL; subsequent top-offs use 200 mL. Allow 20 minutes for dewar equilibrium before acquiring data, and handle liquid nitrogen with appropriate cryogenic PPE and venting.

Factory Specification Sheet
Electrical
Fuse complement differs between spectrometer and microscope

The spectrometer uses 4 equivalent fuses; the microscope uses 1 equivalent fuse. Verify the correct fuse type and rating before replacing any fuse, and isolate power before opening the fuse holder. No voltage or current rating is stated in the verified facts.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the Cary 670 FTIR rather than another model in the Cary FTIR family?

The verified facts define no sibling models, so a direct comparison cannot be made. The Cary 670 FTIR is specified with a liquid-nitrogen dewar (-196 °C, 500 mL initial fill) and a 632.8 nm He-Ne reference laser; select it when these features match the application. Confirm alternatives against their own datasheets.

What purge gas supply and pressure regulation does the Cary 670 FTIR require from adjacent equipment?

The purge gas must be dried to a -70 °C dew point and supplied at 10 L/min (recommended and maximum). If source pressure exceeds 60 psi, a regulator must be installed upstream. The instrument does not include a gas dryer or regulator in the base delivery.

What consumables and service items are needed for the Cary 670 FTIR that are not included in the base delivery?

Liquid nitrogen for the dewar (500 mL initial fill, 200 mL top-off) and a dry purge gas source at -70 °C dew point are required but not supplied. The spectrometer uses 4 equivalent fuses and the microscope 1 equivalent fuse; spare fuses are not listed as included. No maintenance schedule is given in the verified facts.

Dimensional Drawings & Technical Documentation