Agilent Technologies NovoCyte flow cytometer
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies NovoCyte flow cytometer

Agilent Technologies NovoCyte flow cytometer — Research flow cytometry.

The NovoCyte is a benchtop flow cytometer with up to five lasers and fully contained Class I laser optics, configured per application. It operates from 15–32 °C on a grounded three-pin supply and interfaces with a dedicated workstation, fluidics station, and optional NovoSampler/NovoSampler Pro. It is intended for multicolor fluorescence analysis of 12 × 75 mm, 1.5 mL, or 2.0 mL tube samples in research laboratories.

Full specifications

Core Performance & Functional Specifications
Carryover 0.1 percent
Laser Solid state laser with on-board thermal-electric cooling and guaranteed thermal stability and beam quality.
Laser power 20 laser
Sample Flow Rate 5 adjustable
Sheath Flow Rate 6.5 mL/min
Volumetric Absolute Count Precision 5 %
Laser Beam Profile mfr 10 beam.
Optical Alignment Procedure mfr Fixed, no operator alignment required required
Fluorescence Detection mfr Photomultiplier Tube Tube
FSC/SSC Sensitivity mfr 0.5 μm
Fluorescence Resolution mfr 3 CEN
Optical Filters mfr User Exchangeable Exchangeable
Cell Size mfr 0.2 μm
Flow Cell mfr 170 cell
Sample Acquisition Rate mfr 35000 events/second
Sample Delivery mfr Positive-displacement syringe pump enabling dire
Sample Aspiration Volume mfr 5 mL
Compatibility to Autosampler mfr No fluidic tubing disassembly or re-connection r
Dimensions (with reagents containers) (W × D × H) mfr 11.4 in
NovoClean solution container mfr 500 mL
NovoRinse solution container mfr 500 mL
Waste container mfr 3 L
Waste filter (1 pcs) mfr Waste filter filter
NovoRinse solution inlet filter (1 pcs) mfr NovoRinse solution inlet filter filter
NovoClean solution inlet filter (1 pcs) mfr NovoClean solution inlet filter filter
Clean the External Surface of the Instrument mfr Keep the external surface of the instrument clea
Decontamination mfr When suspecting contamination in the system. sys
Prime Fluidics System before Long-Term Storage mfr No plan to run the instrument for more than two
Prime Fluidics System after Long-Term Storage mfr Instrument has not been used for more than two w
Purge Fluidics System before Shipment mfr The instrument needs to be shipped. shipped.
Calibrate the NovoCyte Fluidics Station or NovoCyte Fluidics Cart mfr Abnormal warning about the liquid level in the r
Clean the Optical Filters mfr The optical filters get dirty. dirty.
Clean Fluidics System mfr • The CV of the data is larger than normal. • QC
Sample Injection Probe (SIP) Rinse mfr Automated flying collar wash of inner and outer
Data Acquisition Software mfr NovoExpress NovoExpress
Parameters mfr Height and Area for FSC, SSC and all Fluorescent
Dynamic Range mfr 24 scale
Compensation mfr Full inter-beam matrix, during or post acquisiti
File Export mfr 3.0 report
Data Report mfr Automatic report. Customizable report. Batch PDF
Usage Monitor mfr Comprehensive Transaction Log and System Log. Lo
Quality Control Instrument Daily QC mfr Automatic daily quality control with Agilent Flo
Production QC mfr Each instrument is verified for assembly integri
Sampling Manual Sample Loading mfr 12 holder
Operation Requirements mfr 15 kPa
Regulatory Compliance mfr Indoor and Outdoor Use Use
Equipment Class mfr Class I Laboratory Equipment Equipment
NovoFlow green threshold mfr 0.9 L
NovoRinse green threshold mfr 150 mL
NovoClean green threshold mfr 150 mL
Waste container green threshold mfr 2.8 L
NovoFlow red threshold mfr 0.7 L
NovoRinse red threshold mfr 100 mL
NovoClean red threshold mfr 100 mL
Waste container full threshold mfr 3 L
Automatic compensation stop condition mfr 10000 μL
Automatic compensation threshold mfr 100000
Compensation sample concentration mfr 1,000,000...2,000,000 mL
Dimensions, Weight & Mechanical Design
Weight 51 lb
Instrument Dimension (W × D × H) mfr 23.6 in
Instrument Weight mfr 86.0 lb
Materials, Media & Process Compatibility
NovoFlow sheath fluid container mfr 3 L
NovoFlow sheath fluid in-line filter (2 pcs) mfr NovoFlow sheath fluid in-line filter filter
NovoFlow sheath fluid inlet filter (1 pcs) mfr NovoFlow sheath fluid inlet filter filter
Fluid Level Sensing mfr Active sensing using weight sensors with automat
Fluid Container Capacity mfr 3 waste
Operating Conditions, Environment & Protection
Operating temperature 15 °C
Sound pressure 60 dB
Storage temperature 1 °C
Overvoltage Category mfr Class II II
Pollution degree of the intended environment mfr 2
Degree of Ingress protection mfr IPXO IPXO

Description

Construction and series positioning

Behavior and sample handling

Installation, control, and operation

Integration & Operating Notes

Electrical
Grounded three-pin socket and shipped cord only

The power cord must be inserted in a three-pin socket with a protective ground connection to minimize shock hazard. Use only the power cord shipped with the product; do not use it with any other product, and source replacement cords from Agilent to ensure electrical safety and regulatory compliance.

User Manual, Electrical Safety / p.20; Detachable Power Cord Declaration for Japan / p.25
Mounting
Two-person lift required: instrument exceeds 51 lb (23 kg)

The instrument weighs more than 51 lb (23 kg). Use sufficient assistance and proper lifting techniques when moving or lifting the product to avoid injury and equipment damage. Plan the bench location to keep the back-side power, communication, and fluidic ports accessible for routing.

User Manual, Ergonomics / p.19; NovoCyte Instrument (Front View) / p.46
Process Limits
No flammable gases, fumes, or oxygen-enriched atmosphere

Do not operate the instrument in the presence of flammable gases or fumes, oxygen-enriched atmosphere, or flammable liquid/explosive gas or reagents. Do not use solvents with an auto-ignition temperature below 200 °C; select reagents compatible with the instrument to avoid fire or explosion hazards.

User Manual, General Safety / p.17–18; Limitations / p.23
Control & I/O
Leak response: hold front power button 6–7 s

If leaks are detected, immediately turn off the power by pressing and holding the power button on the front of the instrument for about 6–7 seconds, then contact the local Agilent representative. This quick power-off mitigates electrical and biohazard risks.

User Manual, Electrical Safety / p.20
Maintenance
Replace sheath in-line filter when dirty

The sheath in-line filter must be replaced when dirty to avoid excessive debris in the display. This maintains data quality and prevents flow cell contamination; include filter inspection in the scheduled maintenance routine.

User Manual, Troubleshooting / p.136
Maintenance
Calibrate the fluidics station per System Maintenance

The NovoCyte Fluidics Station must be calibrated; the calibration procedure is described in System Maintenance. Regular calibration ensures accurate fluidic performance and should be scheduled alongside preventative maintenance and replacement of fluidic system consumables.

User Manual, Calibrate NovoCyte Fluidics Station / p.131
Maintenance
Back up workstation data regularly

Data is automatically stored on the workstation hard drive. Regularly back up data to a server or other storage devices to prevent data loss if the hard drive is corrupted or files are deleted.

User Manual, General Safety / p.19
Maintenance
Covers removed only by Agilent authorized personnel

Only Agilent authorized personnel are allowed to remove instrument covers; always disconnect power cables and external circuits before removing covers. This prevents electric shock and laser exposure, and unauthorized service voids warranty.

User Manual, General Safety / p.18

Selection & Integration Q&A

When should I choose the standard NovoCyte configuration versus a customized model or the NovoSampler/NovoSampler Pro option?

Choose the standard configuration when your panel fits its laser/detection layout (Appendix A). Customized models other than the standard configuration are available for different laser needs. Add NovoSampler or NovoSampler Pro only if automated tube sampling is required; the base instrument runs tubes manually via the SIP.

What adjacent equipment and connections must be planned for a complete NovoCyte installation?

The instrument requires a grounded three-pin power socket and the shipped power cord, plus a workstation and monitor and a fluidics station. The back side has four communication ports and five fluidic connection ports; route cables and fluidics accordingly. The fluidics cart and NovoSampler/NovoSampler Pro are optional.

What laboratory environment and duty-cycle limits apply to the NovoCyte?

Operate at 15–32 °C and store at 1–40 °C. Do not operate near flammable gases/fumes, oxygen-enriched atmospheres, or explosive reagents, and avoid solvents with auto-ignition below 200 °C. The instrument is ISM Group 1 Class A (CISPR 11), so EMC difficulties may arise outside non-domestic establishments.

What maintenance schedule and consumables/accessories are required beyond the base delivery?

Scheduled maintenance includes preventative maintenance, replacement of fluidic system consumables, and sheath in-line filter replacement when dirty; the fluidics station must be calibrated. Unscheduled tasks include priming, cleaning, decontamination, and purging before shipment. The fluidics cart and NovoSampler/NovoSampler Pro are optional, not included in the base instrument.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
NovoCyte Flow Cytometer User Guide
Manual ·152 pp.
View PDF