Agilent Technologies NovoCyte Quanteon
Agilent Technologies NovoCyte Quanteon — Research flow cytometry.
The NovoCyte Quanteon is a research-only benchtop flow cytometer built around a fluidics station and workstation, with the instrument rear carrying one power port, four communication ports and five fluidic connection ports. It operates only within a 15–32 °C ambient window and accepts 12 × 75 mm, 1.5 mL or 2.0 mL sample tubes. It is intended for acquisition and analysis of labelled cell samples, not for diagnostic procedures.
Full specifications
| Carryover | 0.1 percent |
| Laser | Solid state laser with on-board thermal-electric cooling and guaranteed thermal stability and beam quality |
| Laser power | 100 mW |
| Sample Flow Rate | 5 adjustable |
| Volumetric Absolute Count Precision | 5 % |
| Laser Beam Profile mfr | 10 beam |
| Optical Alignment Procedure mfr | Fixed, no operator alignment required required |
| Fluorescence Detection mfr | Silicon Photomultiplier with high photon detecti |
| FSC/SSC Sensitivity mfr | 0.4 μm |
| Fluorescence Resolution mfr | 3 CEN |
| Optical Filters mfr | User exchangeable, “Smart” filter automatically |
| Cell Size mfr | 0.1 μm |
| Flow Cell mfr | 170 cell |
| Sample Acquisition Rate mfr | 100000 events/second |
| Sample Delivery mfr | Positive-displacement syringe pump enabling dire |
| NovoCyte Workstation mfr | 7 above. |
| Monitor mfr | 1920 above |
| Computer Operating System mfr | 10 installed |
| Clean the External Surface of the Instrument mfr | Keep the external surface of the instrument clea |
| Decontamination mfr | When suspecting contamination in the system. sys |
| Prime Fluidics System before Long-Term Storage mfr | No plan to run the instrument for more than two |
| Prime Fluidics System after Long-Term Storage mfr | Instrument has not been used for more than two w |
| Purge Fluidics System before Shipment mfr | The instrument needs to be shipped. shipped. |
| Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics Cart mfr | Abnormal warning about the liquid level in the r |
| Clean the Optical Filters mfr | The optical filters get dirty. dirty. |
| Clean Fluidics System mfr | • The CV of the data is larger than normal. • QC |
| Instrument Daily QC mfr | Automatic daily quality control with Agilent Flo |
| Production QC mfr | Each instrument is verified for assembly integri |
| Manual Sample Loading mfr | 12 holder. |
| Indoor and Outdoor Use mfr | Indoor use only only |
| Equipment Class mfr | Class I Laboratory Equipment Equipment |
| Average Sheath Flow Rate mfr | 6.5 mL/min |
| Sample Aspiration Volume mfr | 5 mL |
| Sample Recovery mfr | Available Available |
| Compatibility to Autosampler mfr | No fluidic tubing disassembly or re-connection r |
| Sample Injection Probe (SIP) Rinse mfr | Automated flying collar wash of inner and outer |
| Data Acquisition Software mfr | NovoExpress NovoExpress |
| Parameters mfr | Height and Area for FSC, SSC and all Fluorescent |
| Dynamic Range mfr | 24 scale |
| Compensation mfr | Full inter-beam matrix, during or post acquisiti |
| File Export mfr | 3.0 report |
| Data Report mfr | Automatic report. Customizable report. Batch PDF |
| Usage Monitor mfr | Comprehensive Transaction Log and System Log. Lo |
| NovoCyte Fluidics Cart NovoRinse container volume mfr | 500 mL |
| NovoCyte Fluidics Cart NovoClean container volume mfr | 500 mL |
| NovoCyte Fluidics Cart waste container volume mfr | 15 L |
| Minimum QC sample volume mfr | 400 μL |
| Auto-compensation stop condition (events) mfr | 10000 events |
| Auto-compensation stop condition (volume) mfr | 20 μL |
| Recommended compensation sample concentration mfr | 6...2 mL |
| Instrument Dimension (W × D × H) mfr | 24.4 Q |
| Instrument Weight mfr | 104.0 Q |
| Fluid Level Sensing mfr | Active sensing using weight sensors with automat |
| NovoCyte Fluidics Cart sheath fluid container volume mfr | 15 L |
| Operating temperature | 15 °C |
| Sound pressure | 60 dB |
| Storage temperature | 1 °C |
| Overvoltage Category mfr | Class II II |
| Pollution degree of the intended environment mfr | 2 |
| Degree of Ingress protection mfr | IPXO IPXO |
Description
Architecture and system composition
- The NovoCyte Quanteon is a multi-module system: the cytometer itself, the NovoCyte Workstation and Monitor, the NovoCyte Fluidics Station II, and optionally the NovoCyte Fluidics Cart and NovoSampler Q. The instrument rear concentrates all services: one power port, four communication ports and five fluidic connection ports, so cable and fluidic routing must be planned together.
- Sample introduction is by sample injection probe (SIP) into the sample tube holder; accepted tube formats are 12 × 75 mm, 1.5 mL and 2.0 mL tubes. Plate and tube-rack handling is delegated to the NovoSampler family when fitted.
- Status is reported locally by the power switch light (steady green = powered on, flashing green = shutdown cleaning in progress) and by an LED status indicator (green normal, orange warning, red error), with detail surfaced in the NovoExpress status bar.
Behaviour under load and in operation
- Event rate is governed by three interacting settings: threshold, sample flow rate and sample concentration. A threshold set too high yields zero or low event counts; too low a threshold or an over-concentrated sample produces excessive event rate. Corrective action is to adjust threshold, reduce flow rate, or dilute/re-prepare the sample.
- Signal stability depends on fluidic integrity: air bubbles in the flow cell cause unstable signal, gaps in plots, high CV and QC failure and are cleared with the Debubble function; a clogged flow cell produces QC failure, high CV or low counts and is cleared with the Unclog function.
- Instrument QC must be run with fresh, correctly stored QC particles; expired or improperly stored particles invalidate the QC result and require a new tube prepared and mixed before re-running.
Installation, control and operation
- Installation covers the instrument, workstation, fluidics station, optional cart and optional NovoSampler Q, plus power supply connection; the manual provides space, environment, lifting and carrying requirements.
- Routine operation follows a fixed sequence: startup and pre-inspection, power on, workstation on, add instrument reagents, empty waste, verify/modify configuration, run instrument QC, then acquire samples. Acquisition parameters include stop condition, flow rate and threshold; fluorescence compensation and data analysis are performed in NovoExpress.
- Communication with the workstation is over USB; only one instance of NovoExpress may run, and workstation hibernation must be avoided because it breaks the instrument–workstation link.
Integration & Operating Notes
Deformed waste tubing at the rear of the instrument causes leakage from the sample injection probe. Position the instrument with enough space between its back panel and the wall, and move it forward if the waste line is pinched; the rear also carries the power, communication and fluidic ports that need service access.
A steady green power switch light means the instrument is on; a flashing green light means the shutdown cleaning procedure is still running, so power must not be cut at that point. Plan the power outlet so the switch and its indicator remain visible after installation.
The sample tube holder accepts 12 × 75 mm tubes, 1.5 mL tubes and 2.0 mL tubes. Other formats can cause poor tube fit and sampling faults; when using plates or tube racks with a NovoSampler, load them in the correct orientation and fully seat them on the sample tray to keep the injection probe aligned.
A clogged flow cell is cleared with the Unclog function in NovoExpress, and air in the flow cell with the Debubble function; both are the first corrective actions for QC failure, high CV, low event count or unstable signal. Escalate to fluidic cleaning or consumable replacement only if these do not resolve the fault.
Leakage from the sample injection probe with intact tubing points to a clogged waste filter, which must be replaced. The fluidic system also uses a 0.05 µm sheath fluid ultrafiltration filter and other consumables replaced on a scheduled basis; priming is required before and after long-term storage, and purging before shipment.
Selection & Integration Q&A
The Quanteon is the cytometer module of a modular system; the manual documents it together with the Fluidics Station II, Workstation and optional NovoSampler Q. Choose it when you need the full fluidics-station-plus-workstation architecture with optional automated plate/tube sampling; the sibling variants are not documented in the supplied facts, so compare laser and detection channel configurations in Appendix A before deciding.
The rear panel carries one power port, four communication ports and five fluidic connection ports. The communication ports serve the workstation, the Fluidics Station II and accessories such as the NovoSampler Q, and the workstation links over USB. Reserve rear clearance for these lines and for the waste tubing, and provide the fluidics station and optional cart within reach of the fluidic ports.
Operate only between 15 and 32 °C ambient; store the instrument and accessories between 1 and 40 °C. The workstation must not hibernate during use, since hibernation drops the NovoExpress connection. Before and after long-term storage the fluidic system must be primed, and the instrument purged before shipment.
Scheduled maintenance includes replacing fluidic system consumables, the 0.05 µm sheath fluid ultrafiltration filter and the waste filter, plus debubbling, surface cleaning, decontamination, optical filter cleaning and Fluidics Station II calibration. Instrument reagents, QC particles, sheath fluid and waste containers are user-supplied; the NovoSampler Q and Fluidics Cart are options, not base delivery.