Agilent Technologies NovoCyte Quanteon
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies NovoCyte Quanteon

Agilent Technologies NovoCyte Quanteon — Research flow cytometry.

The NovoCyte Quanteon is a research-only benchtop flow cytometer built around a fluidics station and workstation, with the instrument rear carrying one power port, four communication ports and five fluidic connection ports. It operates only within a 15–32 °C ambient window and accepts 12 × 75 mm, 1.5 mL or 2.0 mL sample tubes. It is intended for acquisition and analysis of labelled cell samples, not for diagnostic procedures.

Full specifications

Core Performance & Functional Specifications
Carryover 0.1 percent
Laser Solid state laser with on-board thermal-electric cooling and guaranteed thermal stability and beam quality
Laser power 100 mW
Sample Flow Rate 5 adjustable
Volumetric Absolute Count Precision 5 %
Laser Beam Profile mfr 10 beam
Optical Alignment Procedure mfr Fixed, no operator alignment required required
Fluorescence Detection mfr Silicon Photomultiplier with high photon detecti
FSC/SSC Sensitivity mfr 0.4 μm
Fluorescence Resolution mfr 3 CEN
Optical Filters mfr User exchangeable, “Smart” filter automatically
Cell Size mfr 0.1 μm
Flow Cell mfr 170 cell
Sample Acquisition Rate mfr 100000 events/second
Sample Delivery mfr Positive-displacement syringe pump enabling dire
NovoCyte Workstation mfr 7 above.
Monitor mfr 1920 above
Computer Operating System mfr 10 installed
Clean the External Surface of the Instrument mfr Keep the external surface of the instrument clea
Decontamination mfr When suspecting contamination in the system. sys
Prime Fluidics System before Long-Term Storage mfr No plan to run the instrument for more than two
Prime Fluidics System after Long-Term Storage mfr Instrument has not been used for more than two w
Purge Fluidics System before Shipment mfr The instrument needs to be shipped. shipped.
Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics Cart mfr Abnormal warning about the liquid level in the r
Clean the Optical Filters mfr The optical filters get dirty. dirty.
Clean Fluidics System mfr • The CV of the data is larger than normal. • QC
Instrument Daily QC mfr Automatic daily quality control with Agilent Flo
Production QC mfr Each instrument is verified for assembly integri
Manual Sample Loading mfr 12 holder.
Indoor and Outdoor Use mfr Indoor use only only
Equipment Class mfr Class I Laboratory Equipment Equipment
Average Sheath Flow Rate mfr 6.5 mL/min
Sample Aspiration Volume mfr 5 mL
Sample Recovery mfr Available Available
Compatibility to Autosampler mfr No fluidic tubing disassembly or re-connection r
Sample Injection Probe (SIP) Rinse mfr Automated flying collar wash of inner and outer
Data Acquisition Software mfr NovoExpress NovoExpress
Parameters mfr Height and Area for FSC, SSC and all Fluorescent
Dynamic Range mfr 24 scale
Compensation mfr Full inter-beam matrix, during or post acquisiti
File Export mfr 3.0 report
Data Report mfr Automatic report. Customizable report. Batch PDF
Usage Monitor mfr Comprehensive Transaction Log and System Log. Lo
NovoCyte Fluidics Cart NovoRinse container volume mfr 500 mL
NovoCyte Fluidics Cart NovoClean container volume mfr 500 mL
NovoCyte Fluidics Cart waste container volume mfr 15 L
Minimum QC sample volume mfr 400 μL
Auto-compensation stop condition (events) mfr 10000 events
Auto-compensation stop condition (volume) mfr 20 μL
Recommended compensation sample concentration mfr 6...2 mL
Dimensions, Weight & Mechanical Design
Instrument Dimension (W × D × H) mfr 24.4 Q
Instrument Weight mfr 104.0 Q
Materials, Media & Process Compatibility
Fluid Level Sensing mfr Active sensing using weight sensors with automat
NovoCyte Fluidics Cart sheath fluid container volume mfr 15 L
Operating Conditions, Environment & Protection
Operating temperature 15 °C
Sound pressure 60 dB
Storage temperature 1 °C
Overvoltage Category mfr Class II II
Pollution degree of the intended environment mfr 2
Degree of Ingress protection mfr IPXO IPXO

Description

Architecture and system composition

Behaviour under load and in operation

Installation, control and operation

Integration & Operating Notes

Mounting
Leave rear clearance so waste tubing cannot kink

Deformed waste tubing at the rear of the instrument causes leakage from the sample injection probe. Position the instrument with enough space between its back panel and the wall, and move it forward if the waste line is pinched; the rear also carries the power, communication and fluidic ports that need service access.

User Manual, Troubleshooting, p.135
Electrical
Power state is read from the power switch light

A steady green power switch light means the instrument is on; a flashing green light means the shutdown cleaning procedure is still running, so power must not be cut at that point. Plan the power outlet so the switch and its indicator remain visible after installation.

User Manual, Introduction, p.46
Process Limits
Use only the specified sample tube formats in the holder

The sample tube holder accepts 12 × 75 mm tubes, 1.5 mL tubes and 2.0 mL tubes. Other formats can cause poor tube fit and sampling faults; when using plates or tube racks with a NovoSampler, load them in the correct orientation and fully seat them on the sample tray to keep the injection probe aligned.

User Manual, Troubleshooting, p.135
Maintenance
Clear clogs and bubbles from software before other remedies

A clogged flow cell is cleared with the Unclog function in NovoExpress, and air in the flow cell with the Debubble function; both are the first corrective actions for QC failure, high CV, low event count or unstable signal. Escalate to fluidic cleaning or consumable replacement only if these do not resolve the fault.

User Manual, Troubleshooting, p.133
Maintenance
Replace the waste filter to stop SIP leakage

Leakage from the sample injection probe with intact tubing points to a clogged waste filter, which must be replaced. The fluidic system also uses a 0.05 µm sheath fluid ultrafiltration filter and other consumables replaced on a scheduled basis; priming is required before and after long-term storage, and purging before shipment.

User Manual, System Maintenance

Selection & Integration Q&A

When should I specify the NovoCyte Quanteon rather than a simpler configuration in this family?

The Quanteon is the cytometer module of a modular system; the manual documents it together with the Fluidics Station II, Workstation and optional NovoSampler Q. Choose it when you need the full fluidics-station-plus-workstation architecture with optional automated plate/tube sampling; the sibling variants are not documented in the supplied facts, so compare laser and detection channel configurations in Appendix A before deciding.

What must be provided on the bench for the instrument's rear connections and adjacent modules?

The rear panel carries one power port, four communication ports and five fluidic connection ports. The communication ports serve the workstation, the Fluidics Station II and accessories such as the NovoSampler Q, and the workstation links over USB. Reserve rear clearance for these lines and for the waste tubing, and provide the fluidics station and optional cart within reach of the fluidic ports.

What ambient and duty conditions apply to installation and storage?

Operate only between 15 and 32 °C ambient; store the instrument and accessories between 1 and 40 °C. The workstation must not hibernate during use, since hibernation drops the NovoExpress connection. Before and after long-term storage the fluidic system must be primed, and the instrument purged before shipment.

What maintenance items and consumables are needed beyond the base instrument?

Scheduled maintenance includes replacing fluidic system consumables, the 0.05 µm sheath fluid ultrafiltration filter and the waste filter, plus debubbling, surface cleaning, decontamination, optical filter cleaning and Fluidics Station II calibration. Instrument reagents, QC particles, sheath fluid and waste containers are user-supplied; the NovoSampler Q and Fluidics Cart are options, not base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
NovoCyte Quanteon Flow Cytometer User Guide
Manual ·148 pp.
View PDF