Agilent Technologies PCG 750
Agilent Technologies PCG 750 — Vacuum gauges.
The PCG 750 is a combination Pirani/capacitance diaphragm gauge with a DN 16 ISO-KF flange, covering 5×10⁻⁵ to 1500 mbar using both sensors continuously. It is the tungsten-filament base version of the PCG-75x series, distinguished from the PCG-752 by filament material and from setpoint/display/Profibus variants by its plain analog output. It serves as a compact, fast-response vacuum sensor for rough-to-fine vacuum monitoring and control.
Full specifications
| Measurement range | 5 mbar |
| Data Rate | 12 Mbaud |
| Internal Volume | 4.7 microliter |
| Operation | 10 °C |
| Response Time | 30 ms |
| Switching Time | 30 ms |
| Vacuum connection | 1.4435 millimeter |
| Measurement principle mfr | thermal conductance acc. to Pirani Pirani |
| Accuracy (N2) mfr | 2.5 reading |
| Repeatability (N2) mfr | 1 reading |
| Output impedance mfr | 2 proof |
| Load impedance mfr | 10 |
| Gauge identification mfr | 71.5 |
| HV adjustment mfr | 10 mbar |
| ATM adjustment mfr | 100 mbar |
| Switching functions setting range mfr | 5.0 mbar |
| Hysteresis mfr | 10 threshold |
| Switching characteristics mfr | Low Trip Point Point |
| Switching function type mfr | 1 function |
| Contact rating closed mfr | 30 resistive |
| Contact rating open mfr | LED lit solid solid |
| Ripple mfr | 1 Vpp |
| Sensor cable mfr | 0.14 conductor |
| Grounding concept vacuum connection to signal common mfr | 10 nF |
| RS232C data format mfr | 8 handshake |
| Materials exposed to vacuum - feedthrough mfr | glass glass |
| Materials exposed to vacuum - orifice mfr | stainless steel steel |
| Materials exposed to vacuum - diaphragm mfr | ceramic ceramic |
| Materials exposed to vacuum - further materials mfr | 1.4301 SnAg |
| Permissible pressure (absolute) mfr | 5 bar |
| Bursting pressure (absolute) mfr | 10 bar |
| Use mfr | 2000 NN |
| Grounding concept mfr | 10 nF |
| Materials exposed to vacuum - Feedthrough mfr | glass glass |
| Materials exposed to vacuum - Orifice mfr | stainless steel steel |
| Materials exposed to vacuum - Diaphragm mfr | ceramic ceramic |
| Materials exposed to vacuum - Further materials mfr | 1.4301 SnAg |
| AGPS45INT mfr | Power supply, PVG-PCG, international plug. plug. |
| PCG750KF16 mfr | -750 KF. |
| PCG750KF16RS mfr | -750 KF. |
| PCG750KF16SD1 mfr | -750 mbar. |
| PCG750KF16SD2 mfr | -750 Torr. |
| PCG750KF16SD3 mfr | -750 Pascal. |
| PCG750KF16SP mfr | -750 mbar. |
| PCG752KF16 mfr | -752 KF. |
| PCG752KF16SD1 mfr | -752 mbar. |
| PCG752KF16SD2 mfr | -752 Torr. |
| PCG752KF16SD3 mfr | -752 Pascal. |
| PCG752KF16SP mfr | -752 mbar. |
| Default setting mfr | 0.1 Chex |
| PCG-752 mfr | Ni Ni |
| Feedthrough mfr | glass glass |
| Orifice mfr | stainless steel steel |
| Diaphragm mfr | ceramic ceramic |
| Further materials mfr | 1.4301 SnAg |
| Year's mean mfr | 65 |
| During 60 days mfr | 85 |
| Grounding concept Vacuum connection to signal common mfr | 10 nF |
| Data format mfr | 8 handshake |
| About this manual mfr | 6 |
| Validity mfr | 6 |
| Definitions and terminology mfr | 7 |
| Definition of Caution, Warning and Note mfr | 7 |
| Warning Symbols mfr | 8 |
| Safety mfr | 9 |
| Proper use mfr | 9 |
| Improper use mfr | 10 |
| Safety guideline for PCG-750, PCG-752 mfr | 11 |
| Disposal mfr | 12 |
| Service mfr | 13 |
| Cable length | 100 m |
| RS232C operation cable length mfr | 30 m |
| RS232C transmission rate mfr | 57600 baud |
| Mounting orientation mfr | any any |
| Cable length, system wiring mfr | according to Profibus specifications specificati |
| Cable length FCC 68 mfr | 30 m |
| Transmission rate mfr | 57600 baud |
| RS232C cable length mfr | 30 m |
| Voltage vs. pressure mfr | 1.286 logarithmic |
| Supply voltage at the gauge mfr | 15 VDC |
| Fuse to be connected mfr | 1 AT |
| Fuse to be connected 3) mfr | 1 AT |
| Profibus Data Rate | 12 Mbaud |
| Output signal measurement range mfr | 0.61 V |
| Diagnostic port mfr | 2.5 pin |
| Power consumption without fieldbus mfr | 2.5 W |
| Power consumption with fieldbus mfr | 3 W |
| Profibus interface physical mfr | 485 |
| Profibus node address default setting mfr | 0.1 Chex |
| Profibus connection cable mfr | 9 cable |
| Profibus cable length, system wiring mfr | according to Profibus specifications specificati |
| Weight w/o fieldbus interface mfr | 115 g |
| Weight with fieldbus interface mfr | 230 g |
| Profibus connection mfr | 9 |
| w/o fieldbus interface mfr | 115 g |
| with fieldbus interface mfr | 230 g |
| Specification, data format, communication protocol mfr | 485 |
| Weight without fieldbus interface mfr | 115 g |
| Materials exposed to vacuum - filament PCG-752 mfr | Ni Ni |
| Materials exposed to vacuum - Filament PCG-752 mfr | Ni Ni |
| Filament mfr | 160 °C |
| Operating temperature | 10 °C |
| Protection rating | 40 |
| Storage temperature | 20 °C |
| Relative humidity year's mean mfr | 65 |
| Relative humidity during 60 days mfr | 85 |
| Relative humidity - Year's mean mfr | 65 |
| Relative humidity - During 60 days mfr | 85 |
Description
Construction and series positioning
- The PCG 750 combines a Pirani thermal-conductance sensor and a diaphragm capacitive sensor in one body with a DN 16 ISO-KF vacuum connection. Both sensors are constantly active: Pirani signal dominates below ~1 mbar, capacitive signal above ~10–100 mbar (gas-dependent crossover), and a proportional blend in between.
- Within the PCG-75x family, this is the base gauge with tungsten filament (PCG-752 uses a nickel-plated filament). It lacks the setpoint relays, LCD display, and Profibus interface of the SD/SP variants; those functions are available only in sibling models and are not part of this unit.
- Internal volume is 4.7 cm³, adding minimal dead volume to small chambers. Weight is 115–130 g without fieldbus.
Measurement behavior and media
- The output is logarithmic: U = c + 1.286 log₁₀ p, valid from 5×10⁻⁵ to 1500 mbar, with a measurement-range voltage of +0.61 to +10.23 V. Controllers must interpret this logarithmic signal; the error signal is 0 V by default.
- Accuracy (N₂) is ±50% of reading at 5×10⁻⁴…1×10⁻³ mbar, ±15% at 1×10⁻³…100 mbar, ±5% at 100…950 mbar, and ±2.5% at 950…1050 mbar; repeatability is ±2% over 1×10⁻³…1100 mbar.
- In the Pirani range below 1 mbar, readings depend on gas type; apply calibration factors (e.g., He 0.8, Ar 1.7, H₂ 0.5, CO₂ 0.9) to obtain effective pressure. Above ~1 mbar the capacitive sensor is gas-independent.
- Response time and switching time are <30 ms, making the gauge suitable for fast pressure control and interlock applications.
Installation, connection, and operation
- Mount in any orientation but free of vibrations; indoors only. The vacuum connection must stay ≤80 °C; ambient operating range is +10…+50 °C. Maximum absolute pressure is 5 bar; bursting pressure is 10 bar.
- Supply is +15…+30 VDC SELV with ≤1 Vpp ripple, fused at 1 AT. Power consumption is ≤2.5 W without fieldbus. Use shielded sensor cable (0.14 mm²/conductor, ≤100 m) and observe single-point grounding to avoid ground loops.
- The 8-pin FCC-68 connector provides supply, analog output, gauge identification, and relay contacts. A 2.5 mm 3-pin RS232C diagnostic port (57600 baud, 8 data bits, no parity, no handshake) allows reading and parameter setting; cable length ≤30 m.
- Allow at least 10 minutes stabilization after power-up. Setpoints SP1/SP2 are factory set to the lower range limit and do not switch by default; configure them before relying on relay outputs. Do not press the <SP> button if the analog signal controls a process, as it suppresses the measurement signal.
Integration & Operating Notes
The gauge must be powered from a grounded protective extra-low voltage (SELV) source and the circuit must include a 1 AT fuse. Exceeding the 30 VDC upper limit or injecting excessive ripple can damage the electronics or corrupt the analog output; Agilent controllers already meet this requirement.
The permissible absolute pressure is ≤5 bar and bursting pressure is 10 bar. Use overpressure-suitable clamps and O-rings with an outer centering ring for the KF connection. The ceramic diaphragm sensor is fragile; avoid sudden pressure surges or mechanical shock that could crack it.
Any orientation is allowed, but the gauge must not be subjected to vibration (e.g., directly on a roughing pump line). The vacuum connection temperature must stay ≤80 °C; if baking the system, ensure the flange does not exceed this limit. Indoor use only, with relative humidity ≤65% annual mean and no condensation.
Both setpoints are factory set to the lower measurement range limit (5×10⁻⁵ mbar) and therefore do not switch by default. If the relay contacts are wired into a valve or interlock circuit, set the trip points and hysteresis (10% of threshold) via the RS232C interface or buttons first; otherwise the interlock will never trigger.
Pushing the <SP> button suppresses the measurement signal and outputs the threshold value instead. If a controller or PLC uses the analog output for pressure control, this action will cause a sudden false reading and possible malfunction. Use the RS232C diagnostic port for parameter changes instead.
Zero drift can occur after long operation or contamination. ATM adjustment (default 1000 mbar) must be done at atmospheric pressure after at least 10 minutes stabilization; HV adjustment (default 5×10⁻⁵ mbar) requires evacuation to p << 10⁻⁵ mbar and a 2-minute wait. Successful HV adjustment is indicated by a 4.99×10⁻⁵ mbar output. Replace the sensor if contamination is severe.
Selection & Integration Q&A
Choose the PCG 750 for standard vacuum monitoring with a tungsten filament and plain analog output. The PCG 752 offers a nickel-plated filament for slightly different chemical compatibility. If you need local readout, relay setpoints, or Profibus, select the SD or SP variants; those functions are not present on the base PCG 750.
Use a SELV supply of +15…+30 VDC with ≤1 Vpp ripple, fused at 1 AT. The 8-pin FCC-68 connector carries supply, analog output, and relay contacts. Use shielded cable with 0.14 mm² conductors, max 100 m, and ground the shield on one side. The analog output is logarithmic (0.61–10.23 V) and must be interpreted accordingly.
The gauge is factory calibrated; perform ATM and HV adjustments if drift occurs, and replace the sensor if severely contaminated. The base delivery includes the gauge only—the power supply (AGPS45INT), mating FCC-68 connector, RS232C cable, and Profibus GSD file are not included and must be ordered separately. Setpoints are factory set to the lower limit and need configuration.