Agilent Technologies PVG 500 Pirani Gauge
Agilent Technologies PVG 500 Pirani Gauge — Vacuum gauges.
The PVG-500 is a Pirani thermal-conductance vacuum gauge with a tungsten filament, covering 5×10⁻⁴ to 1000 mbar for air, O₂, CO and N₂. It is the base DN 16 ISO-KF variant without switching functions, distinguished from the PVG-502 by filament material and from S-suffixed siblings by the absence of setpoint relays. It serves as a compact, low-power pressure sensor in rough-to-medium vacuum control loops.
Full specifications
| Measurement range | 5 mbar |
| Electrical connection | 68 male |
| Internal Volume | 1.5 microliter |
| Operation | 5 |
| Repeatability | 2 reading |
| Resolution | 1 reading |
| Response Time | 80 ms |
| Vacuum connection | 80 millimeter |
| Measurement principle mfr | thermal conductance according to Pirani Pirani |
| Accuracy (N2) mfr | 50 reading |
| Output impedance mfr | 2 |
| Minimum loaded impedance mfr | 10 |
| Gauge identification mfr | 27.0 common |
| Adjustment mfr | one tactile switch for ATM and HV adjustment adj |
| Switching function setting range mfr | 2 mbar |
| Switching function hysteresis mfr | 10 threshold |
| Relay contact rating mfr | 30 floating |
| Ripple mfr | 1 |
| Sensor cable mfr | 68 male |
| Grounding concept mfr | 1 conduct |
| Materials exposed to vacuum mfr | 1.4301 NiFe |
| Admissible pressure mfr | 10 gases |
| Use mfr | 2000 NN |
| Measurement signal equation constant c mfr | 6.143 |
| Calibration factor C mfr | 0.7 |
| Switching functions setting range mfr | 2 mbar |
| Switching functions hysteresis mfr | 10 threshold |
| Relay contact mfr | 30 floating |
| At gauge mfr | 14 |
| Vacuum connection to signal common mfr | 1 |
| Supply common to signal common mfr | conducted separately, for differential measureme |
| Storage mfr | 20 |
| Product Description mfr | 15 |
| Technical Specifications mfr | 16 |
| PVG-500 and PVG-502 Outline mfr | 18 |
| Measurement Signal vs. Pressure mfr | 19 |
| Installation mfr | 21 |
| Power Connection mfr | 23 |
| Adjusting the Gauge mfr | 25 |
| Switching functions mfr | 26 |
| Adjusting the Setpoints mfr | 27 |
| Deinstallation mfr | 28 |
| Maintenance, Repair mfr | 29 |
| Accessories and Spare Parts mfr | 30 |
| About this manual mfr | 6 |
| Validity mfr | 6 |
| Definitions and terminology mfr | 7 |
| Definition of Caution, Warning and Note mfr | 7 |
| Warning Symbols mfr | 8 |
| Safety mfr | 9 |
| Proper use mfr | 9 |
| Improper use mfr | 10 |
| Safety guideline for PVG-500 and PVG-502 mfr | 11 |
| Disposal mfr | 12 |
| Service mfr | 13 |
| He mfr | 0.8 |
| Ne mfr | 1.4 |
| Ar mfr | 1.7 |
| Kr mfr | 2.4 |
| Xe mfr | 3.0 |
| Setpoint setting range mfr | 2 mbar |
| ATM adjustment default mfr | 1000 mbar |
| HV adjustment default mfr | 1.2 mbar |
| Cable length | 100 m |
| Weight | 80 g |
| Mounting orientation mfr | any any |
| Power consumption | 1 W |
| Voltage vs. pressure mfr | 1.286 V/decade |
| Supply voltage at gauge mfr | 14 VDC |
| Current consumption mfr | 500 |
| Fuse required mfr | 1 |
| Fuse required 1) mfr | 1 |
| Setpoint voltage range mfr | 2.67 VDC |
| Output signal measurement range mfr | 1.9 VDC |
| DN 16 ISO-KF mfr | 80 |
| DN 16 CF-R mfr | 100 |
| Filament rupture signal mfr | 0.1 V |
| Filament mfr | 110 |
| Gas Type Dependence mfr | 24 |
| Protection rating | 40 |
| Relative humidity mfr | 80 °C |
Description
Construction and series positioning
- The PVG-500 uses a tungsten filament; the PVG-502 is the nickel-filament alternative for applications where tungsten is unsuitable.
- This model is the DN 16 ISO-KF version without switching functions (PVG500KF16). Siblings add either switching functions (PVG500KF16S) or a CF-R flange (PVG500CF16/CF16S).
- Internal volume is approximately 1.5 cm³ and weight 80 g for the KF version, so it adds minimal dead volume and mechanical load to small manifolds.
Measurement behavior
- Output is logarithmic: 1.286 V/decade over +1.9…+10.0 VDC, with a response time of 80 ms.
- Accuracy is ±15% of reading from 1×10⁻³ to 100 mbar, degrading to ±50% at the range extremes (5×10⁻⁴…1×10⁻³ mbar and 100…1000 mbar).
- Readings are gas-dependent; for gases other than air, O₂, CO and N₂ apply a calibration factor C (e.g. Ar 1.7, He 0.8, H₂ 0.5) below 1 mbar.
- The gauge is factory-calibrated; ATM and HV adjustments are made with a tactile switch (max. ø1.1 mm pin) to correct zero drift.
Installation and operation
- Mounting orientation is unrestricted; horizontal to upright is preferred to keep condensates and particles out of the sensor.
- The 8-pin FCC-68/RJ45 connector carries supply, measurement signal, gauge identification (27.0 kΩ) and, on S variants, relay contacts. This model has no relay outputs.
- Supply is +14…+30 VDC, ≤1 W, <500 mA starting current, fused with a 1 A slow-blow fuse; the supply must be SELV and grounded.
- Allow at least 10 minutes stabilization after power-up; continuous operation is recommended.
Integration & Operating Notes
The gauge is designed only for a grounded SELV supply with a 1 A slow-blow fuse in the feed. Connecting to a non-SELV or unfused source removes the protective separation and can damage the electronics or create a shock hazard in fault conditions.
The admissible pressure is 10 bar absolute and is limited to inert gases. The filament reaches 110 °C and can ignite flammable or combustible gas mixtures with oxidants within the explosion range, so such service is prohibited regardless of pressure.
After supply voltage is applied, the measurement signal is available between pins 3 and 5, but at least 10 minutes stabilization is required before readings are meaningful. Interlocks or process starts driven by this signal should be inhibited during that period; continuous operation is advisable irrespective of pressure.
The gauge is factory-calibrated but zero drift can occur from long operation or contamination. ATM and HV adjustments can restore calibration; if contamination is severe or the gauge malfunctions, the sensor must be replaced. Filament wear and contamination are not covered by warranty, and end-user repair voids liability.
Selection & Integration Q&A
Pick the PVG-500 (PVG500KF16) for DN 16 ISO-KF systems needing no relay setpoints and where a tungsten filament is acceptable. Choose PVG-502 for a nickel filament, or an S variant (PVG500KF16S) if you need two adjustable setpoint relays for interlocks or alarms.
It needs a grounded SELV supply of +14…+30 VDC with ≤1 Vpp ripple, fused at 1 A slow-blow, drawing <500 mA starting current and ≤1 W. The controller input must accept 0…+10.3 VDC and decode the 1.286 V/decade logarithmic signal; gauge identification is 27.0 kΩ on pin 4.
Allow 10 minutes stabilization after power-up and adjust ATM/HV with a max. ø1.1 mm pin if drift occurs; replace the sensor if contamination is severe. The 8×0.14 mm² shielded sensor cable, clamping ring and centering ring are not included in base delivery and must be ordered separately.