Agilent Technologies Transformer oil gas, transformer oil analysis
Agilent Technologies Transformer oil gas, transformer oil analysis — Gas chromatography.
The Agilent TOGA is a transformer oil gas analyzer built on the 8890 GC and 8697 headspace sampler, configured with FID/TCD detection and 2 valves/2 columns per ASTM D3612. It quantifies dissolved acetylene, ethane and methane in a 22.5-minute run, distinguishing the headspace (Method C) variant from vacuum-extraction (Method A) siblings.
Full specifications
| Analysis time | 22.5 min |
| Run Time | 22.5 min |
| Software | OpenLab CDS |
| GC platform mfr | 8890 |
| Headspace sampler mfr | 8697 |
| Detectors mfr | FID/TCD FID/TCD |
| Analyzed compounds mfr | acetylene, ethane, methane methane |
| Valve/column configuration mfr | 2 columns/TCD/FID |
| Detector configuration mfr | TCD/FID TCD/FID |
| Enhanced method mfr | 3612 C |
| Pneumatics control mfr | microchannel-based electronic pneumatics control |
| Measured components mfr | 2 |
| Maximum sample sequence mfr | 120 samples |
| Sampler mfr | 8697 sampler |
| Standard configuration 1 mfr | 3612 A |
| Standard configuration 2 mfr | 3612 columns |
| Standard configuration 3 mfr | 3612 opt |
| Standard method mfr | 3612 A |
| Compliance standard mfr | 3612 |
Description
Construction and series positioning
- The system pairs an Agilent 8890 gas chromatograph with an 8697 headspace sampler; the analyzed compounds listed are acetylene, ethane and methane.
- The standard configuration is a 2-valve/2-column arrangement with FID/TCD detectors, referenced to ASTM D3612. The document digests describe three standard TOGA configurations: FID/TCD with packed columns per ASTM D3612 Method A (vacuum extraction); FID/TCD per Method C (headspace injection) with methanizer and PLOT columns; and a Method C variant with pre-column backflush, PLOT columns and methanizer bypass for trace olefins. This card covers the headspace-injection FID/TCD configuration; select the Method A packed-column variant when vacuum extraction is mandated, and the backflush/PLOT variant when trace olefins are the target.
- Pneumatics use microchannel-based electronic pneumatics control.
Behavior and run characteristics
- The stated analysis/run time is 22.5 min per sample. No digitized curve points were supplied, so no elution-order or response-factor behavior is asserted here.
- The maximum sample sequence is 120 samples, which sets the unattended batch horizon for a headspace run at 22.5 min each.
Control and operation
- Instrument control and data handling run under OpenLab CDS. The 8697 sampler handles headspace injection; the 8890 GC carries the 2-valve/2-column/FID/TCD configuration.
- The document digests contain no section-level installation or connection facts, so wiring, gas-supply and utility specifics must be taken from the full manuals rather than inferred here.
Integration & Operating Notes
The TOGA family spans ASTM D3612 Method A (vacuum extraction, packed columns) and Method C (headspace injection, PLOT columns with methanizer). This FID/TCD headspace configuration is not interchangeable with the Method A packed-column variant; the extraction procedure used in the laboratory must be fixed first, because column set, valving and methanizer presence follow from it.
At 22.5 min per run, a full 120-sample sequence represents roughly 45 hours of continuous operation. Plan sampler vial capacity, tray loading and any intermediate calibration or check standards around that horizon so that a batch is not interrupted mid-sequence.
Columns, methanizer, sampler vials and septa, calibration gas standards and carrier/detector gases are separate items; the base system comprises the 8890 GC, 8697 sampler and OpenLab CDS. Budget these consumables and their replacement intervals into commissioning, since the supplied documentation digests give no maintenance schedule.
The available document digests list no voltage, power, gas-purity or bench-clearance data for this configuration. Confirm electrical supply, carrier and fuel gas plumbing, and exhaust/vent routing against the 8890 GC and 8697 site-preparation manuals before installation; do not size utilities from this card.
Selection & Integration Q&A
Pick this configuration when the laboratory follows ASTM D3612 Method C headspace injection and needs acetylene, ethane and methane by FID/TCD in a 22.5-min run. Choose the Method A packed-column variant if vacuum extraction is mandated, or the backflush/PLOT variant with methanizer bypass when trace olefins are the analytical target.
The base system is the 8890 GC, 8697 headspace sampler and OpenLab CDS; columns, methanizer (where fitted), sampler vials and septa, calibration standards and carrier/fuel gases are separate. Confirm gas plumbing and electrical supply against the 8890 and 8697 site-preparation manuals, since the brochure digests give no utility data.
The maximum sample sequence is 120 samples at 22.5 min per run, about 45 hours of continuous operation. Plan vial tray loading, calibration checks and operator intervention within that window; the available digests give no further environmental or duty-cycle limits, so consult the full manuals for bench conditions.
Columns, methanizer, sampler vials and septa, calibration gas standards and carrier/detector gases are not part of the base 8890 GC + 8697 + OpenLab CDS package. The supplied document digests contain no maintenance schedule or service intervals, so replacement intervals must be taken from the full instrument manuals.