Avantik EM8H
Avantik ·Made in USA · In Active Production

Avantik EM8H

Avantik EM8H — Embedding centers.

The Avantik EM8H is the dispensing console module of the EM8 Embedding Center, a benchtop paraffin embedding station with a 5 L reservoir and two 1.7 L side trays. It provides a 12 x 14.75 inch working area held at 50–70 °C for tissue orientation and cassette embedding. It is used upstream of the EM8C cryo console in histology sample preparation workflows.

Full specifications

Core Performance & Functional Specifications
Working area 12 x 14.75 inch
Working temperature 50-70 °C
Dimensions, Weight & Mechanical Design
Depth 17 inch
Height 26 inch
Weight 62 lbs
Width 23.25 inch

Description

Construction and positioning

Integration & Operating Notes

Electrical
Universal mains input needs no site transformer

The 100–240 V / 50–60 Hz input means the console can be connected to European or North American laboratory circuits directly; confirm the local plug type and circuit rating at installation, since the documentation digest does not state current draw.

Avantik-EM8-Embedding-Center-Sheet, Extra Specifications
Process Limits
Working temperature must stay within 50–70 °C

The working area is specified only for 50–70 °C. Operating outside this band is not covered by the published specification, and paraffin type should be chosen so its dispensing viscosity is reached inside this window.

Avantik-EM8-Embedding-Center-Sheet, Specifications
Process Limits
Reservoir and tray volumes cap batch size between refills

The 5 L main reservoir and two 1.7 L side trays define how many cassettes can be processed before replenishment; plan refill breaks accordingly, as no low-level alarm or automatic topping-up feature is described in the available documentation.

Avantik-EM8-Embedding-Center-Sheet, Extra Specifications
Maintenance
Heated forceps are separate accessories, not in base delivery

Electric heated forceps EQ6109, EQ6110 and EQ6111 are listed as accessories; order them with the console if heated manipulation is required, and obtain cleaning and paraffin-change intervals from the full manual, which is not summarized in the supplied digest.

Avantik-EM8-Embedding-Center-Sheet, Accessories

Selection & Integration Q&A

When should I choose the EM8H instead of the EM8C cryo console?

Pick the EM8H when the task is heated paraffin dispensing and tissue orientation: 5 L reservoir, two 1.7 L trays, 50–70 °C work area. The EM8C is the cold module (0 to −15 °C) for block solidification; most labs need both.

What does the EM8H need from adjacent equipment and services?

Only a 100–240 V / 50–60 Hz mains outlet and a level bench sized for the 23.25 x 17 in, 62 lb enclosure. Heated forceps (EQ6109–EQ6111) are separate accessories; no controller, drive, or external chiller connection is specified.

What duty-cycle and environment limits apply to the EM8H?

The working area is specified for 50–70 °C only, and the documentation digest gives no duty-cycle rating, warm-up time, or ambient temperature range. Keep the unit on a stable bench and confirm continuous-operation suitability with the full manual.

What maintenance and consumables are not included with the EM8H?

Base delivery covers the console only; electric heated forceps EQ6109, EQ6110 and EQ6111 are optional accessories. Paraffin and cassettes are user-supplied, and the supplied digest lists no service intervals, so follow the full manual for cleaning and reservoir changes.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Avantik-EM8-Embedding-Center-Sheet
Datasheet ·1,8 МБ ·2 pp.
View PDF