BINDER E 28 — Drying ovens with Temperature 60…230 °C.

The BINDER E 28 is a naturally ventilated, electrically heated drying and heating oven with a 28 L stainless steel (V2A 1.4301) interior chamber and a 0–120 minute mechanical timer. It operates from 60 °C to 230 °C, distinguishing it from the B 28 incubator sibling (30–70 °C) within the same 28 L chassis. It is intended for drying, heat treatment and hot-air sterilization of laboratory loads up to 25 kg total.

Temperature
60…230 °C

Full specifications

Core Performance & Functional Specifications
Temperature 60...230 °C
Heating time to 150°C 36 min
Interior volume 28 L
Load per rack 10 kg
Nominal power 0.8 kW
Permitted load 25 kg
Rear wall clearance 160 mm
Side wall clearance 100 mm
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 19 min
Unit doors mfr 1
Dimensions, Weight & Mechanical Design
Depth 425 mm
Height 405 mm
Interior depth 250 mm
Interior height 280 mm
Interior width 400 mm
Width 580 mm
Net weight of the unit (empty) mfr 22 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior and load

Installation and operation

Integration & Operating Notes

Electrical
Provide a grounded 230 V outlet within reach of the 1800 mm cable

The fixed power cable is at least 1800 mm long and terminates in a CEE 7/7 grounded plug. The outlet must remain accessible for emergency disconnection, and supply tolerance is ±10 % at 50 or 60 Hz. The 230 V variant carries two chamber fuses.

Operating Manual, Section 4.1 Electrical Connection
Process Limits
Do not set below 60 °C or load beyond 25 kg total

The E 28 control range begins at 60 °C; lower set-points are outside the specified range. Total charge must not exceed 25 kg, with no more than 10 kg per rack, to preserve airflow and thermal uniformity. Avoid corrosive charges such as acids and halides that attack the stainless steel.

Operating Manual, Section 2 Technical Description
Environment
Install indoors at 18–40 °C and max 70 % RH non-condensing

Performance data are referenced to +22 °C ± 3 °C ambient. Operation outside +18 to +40 °C, above 70 % relative humidity, above 2000 m altitude, or in damp areas voids the specified behavior and risks condensation damage given IP 20 protection.

Operating Manual, Section 3.4 Ambient Conditions

Selection & Integration Q&A

When should I choose the E 28 rather than the B 28 incubator of the same 28 L size?

Choose the E 28 for drying, heat treatment or hot-air sterilization from 60 °C to 230 °C. Choose the B 28 when the process requires incubation between 30 °C and 70 °C with a main power switch and thermostat-brake overtemperature setting.

What supply and protection must be provided for the E 28, and can it share a circuit with other equipment?

It needs a grounded 230 V, 50/60 Hz supply within ±10 %, nominal current 3.5 A and 0.8 kW. The 230 V variant has two chamber fuses. Size the upstream circuit for at least the nominal current and keep the plug accessible for emergency disconnection.

What maintenance and consumables are required beyond the base delivery?

Plan annual maintenance and electrical repeat inspection by BINDER-authorized personnel, plus door gasket replacement when cold. The base delivery includes only the fixed cable and plug; racks, spare fuses and gaskets are accessories to be ordered separately.