BINDER ED-S 115
BINDER ·Made in Germany · In Active Production

BINDER ED-S 115

BINDER ED-S 115 — Drying ovens with Temperature 7…250 °C.

The ED-S 115 is a 118 L natural-convection drying and heating oven (Solid.Line) with a stainless steel interior and PID controller, operating from +7 to 250 °C. It is the larger 115-size variant of the ED-S series, distinguished from the ED-S 56 by its 118 L volume and 75 kg total load capacity. It serves gentle drying, heating, and thermal treatment of non-combustible materials where forced air is not required.

Temperature
7…250 °C

BINDER ED-S Series — Model Comparison & Selection

Model Temperature, °CDepth, mmHeating up time to 150 c, minHeight, cmInterior depth, mmInterior volume, L
ED-S 115 Current 7–2506165573.5390118
ED-S 56 7–2505776062.535062

Full specifications

Core Performance & Functional Specifications
Temperature 7...250 °C
Heating time to 150°C 55 min
Interior volume 118 L
Load per rack 20 kg
Nominal power 1.25 kW
Permitted load 75 kg (EDS115-230V)
Temperature fluctuation 0.9 K
Temperature uniformity 2.8 K
Rear wall clearance 160 mm (EDS115-230V)
Side wall clearance 100 mm (EDS115-230V)
Article Number mfr 9090
Recovery time after door was opened for 30 s at 150°C mfr 40 min
Dimensions, Weight & Mechanical Design
Depth 616 mm (EDS115-230V)
Height 735 mm (EDS115-230V)
Interior depth 390 mm (EDS115-230V)
Interior height 550 mm (EDS115-230V)
Interior width 550 mm (EDS115-230V)
Width 764 mm (EDS115-230V)
Net weight of the unit (empty) mfr 50 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection, and operation

Integration & Operating Notes

Mounting
Maintain 160 mm rear and 100 mm side clearance

Insufficient clearance prevents heat dispersal and can cause overheating or damage. Do not install the chamber in unventilated recesses; ensure the room has adequate ventilation to remove dissipated heat.

Instruction Manual BD-ED-FD-S (E1) 06-2026_en, chap. 3.4, p. 17
Electrical
Supply voltage must stay within ±10 % of nominal

The 230 V model requires a 1N~ 230 V 50/60 Hz supply with protective earth and a 6.3 A fuse. Verify mains quality before installation; out-of-tolerance voltage can affect controller operation and heating performance.

Instruction Manual BD-ED-FD-S (E1) 06-2026_en, type plate p. 8; chap. 4.3 p. 19-20
Process Limits
Do not process combustible or explosive materials

The chamber has no explosion protection. Operating temperature must remain below the flash point or sublimation point of the charging material. Only harmless materials may be processed; corrosive substances (acids, halides) can damage the stainless steel interior.

Instruction Manual BD-ED-FD-S (E1) 06-2026_en, chap. 1.9, p. 10-11
Control & I/O
Set class 2 safety controller before first use

The ED-S 115 is fitted with a class 2 overtemperature safety controller per DIN 12880. Configure its set-point according to the manual to ensure independent protection against overheating; an unconfigured safety controller may not trip at the intended limit.

Instruction Manual BD-ED-FD-S (E1) 06-2026_en, chap. 9, p. 24
Maintenance
Annual electrical safety test and maintenance required

Perform maintenance at least once a year and repeat electrical inspections per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial start-up and after any repair. Disconnect the power plug before maintenance; do not remove the rear panel. Replace the door gasket only when cold.

Instruction Manual BD-ED-FD-S (E1) 06-2026_en, chap. 12.1-12.2, p. 29-30
Mounting
Lift only at the four lower corners with two people

Never lift or transport the chamber using the door handle or door. After cold storage, wait at least one hour at ambient temperature before start-up to prevent condensation-related electrical hazards.

Instruction Manual BD-ED-FD-S (E1) 06-2026_en, chap. 3.1-3.3, p. 16-17

Selection & Integration Q&A

When should I choose the ED-S 115 over the smaller ED-S 56 or a forced-convection FD-S model?

Choose the ED-S 115 for larger loads up to 75 kg and 118 L where gentle natural convection is acceptable. The ED-S 56 offers only 56 L. Select an FD-S model if faster recovery after door openings or forced air circulation is needed.

What are the electrical supply requirements for the ED-S 115?

The 230 V variant requires a 1N~ 230 V 50/60 Hz supply with protective earth, a CEE 7/7 plug, and a 6.3 A fuse. Voltage must remain within ±10 % of nominal. The 120 V variant uses a NEMA 5-15 plug and 12.5 A fuse.

What maintenance and accessories are required, and what is not included in the base delivery?

Annual maintenance and electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. The base delivery includes the chamber only; racks, exhaust connection parts, and spare door gaskets must be ordered separately. Replace the door gasket only when cold.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model ED S 115 en
Datasheet ·394,5 КБ ·5 pp.
View PDF
PDF
BD ED FD S (E1) 06 2026 en
Manual ·10,8 МБ ·65 pp.
View PDF