BINDER KBF 1600
BINDER KBF 1600 — Climate chambers with Temperature 0…70 °C.
The BINDER KBF 1600 is a 1,610 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, housed in a 1,795 mm wide cabinet. As the largest standard KBF model, it is distinguished from smaller siblings by its 40 kg per-rack load capacity and 640 W maximum heat compensation. It serves as a high-throughput test platform for stability and material testing under defined climatic conditions.
BINDER KBF Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 1610 L (KBF1600-230V / KBF1600UL-120V) |
| Load per rack | 40 kg (KBF1600-230V / KBF1600UL-120V) |
| Maximum heat compensation | 640 W |
| Nominal power | 1.7 kW (KBF1600-230V / KBF1600UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.4 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 3 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 7 min |
| Height | 1951 mm (KBF1600-230V / KBF1600UL-120V) |
| Width | 1795 mm (KBF1600-230V / KBF1600UL-120V) |
| Power frequency | 50/60 Hz (KBF1600-230V / KBF1600UL-120V) |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A (KBF1600-230V / KBF1600UL-120V) |
| Humidity fluctuation | 1.5 % RH |
| Humidity range | 10...80 % RH |
Description
Construction and Series Positioning
The KBF 1600 is the largest chamber in the KBF series, offering 1,610 L of interior volume in a 1,795 mm wide by 1,951 mm high cabinet. Unlike the smaller KBF 130 and 260, which use a triangular instrument panel, the KBF 1600 features a signal triangle and an ergonomically adjustable control terminal for operator access. The chamber is equipped with a temperature safety device according to DIN 12880:2007 class 2/3.3, configurable as a temperature limiter or temperature protection controller. The refrigeration system uses R600A hydrocarbon refrigerant, and the humidification system design depends on the serial number (up to or from 20250000010225).
Behavior Under Load and in Different Media
- Temperature control is stable to 0.1 K fluctuation and 0.4 K uniformity across the 0...70 °C range, with humidity control from 10 to 80 % RH at 1.5 % RH fluctuation.
- Recovery after a 30 s door opening is 3 minutes at 25 °C/60 % RH and 7 minutes at 40 °C/75 % RH, indicating rapid re-stabilization for frequent access.
- The chamber supports up to 40 kg per rack and a maximum heat compensation of 640 W, allowing testing of active or heat-generating samples.
- Humidity sensor tolerance is ±2 % RH; exposure to ammonia, acetone, ethanol, formaldehyde, ozone, hydrochloric acid, hydrogen sulphide, nitrogen oxides, and sulphur dioxide must remain below the listed permanent-load concentrations to avoid sensor damage.
Installation, Connection, and Control
- The chamber requires a 220...240 V, 1~, 50/60 Hz supply with a 16 A unit fuse and a CEE 7/7 plug; nominal power is 1.7 kW. A 120 V UL variant is available separately.
- From size 260 upward, a wall spacer is required to maintain proper airflow; the KBF 1600 falls under this requirement.
- Water connections for humidification and wastewater must be rated for at least 11 bar maximum operating pressure. The humidifying system design differs by serial number, so spare parts and connection methods must be selected accordingly.
- Control is via the MB2 chamber controller with program control, fixed-value mode, time and week programs, and a chart recorder display. Temperature safety devices can be configured as class 2 or 3.3 per DIN 12880.
- The chamber is IP 20 rated, so it must not be sprayed with water. Electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended before initial startup and after maintenance.
Integration & Operating Notes
The KBF 1600 draws 1.7 kW nominal and requires a 16 A unit fuse. Verify site voltage and frequency before connection; use overcurrent protection rated for the chamber. The 120 V UL variant is a separate model and cannot be connected to a 230 V supply.
Permanent exposure to ammonia above 5500 ppm, acetone above 3300 ppm, ethanol above 3500 ppm, formaldehyde above 2400 ppm, ozone above 0.5 ppm, hydrochloric acid above 300 ppm, hydrogen sulphide above 350 ppm, nitrogen oxides above 5 ppm, or sulphur dioxide above 5 ppm will damage the humidity sensor. Account for these limits when testing materials that off-gas such substances.
From chamber size 260 upward, including the KBF 1600, a wall spacer must be installed to maintain the specified distance from the wall. Failure to do so can impair heat dissipation and humidity control, leading to inaccurate results or chamber malfunction.
Regular maintenance must be performed at least once a year, including steam humidifier service. Condensers should be cleaned at least twice a year by a qualified technician; in dusty environments, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and must be reset by BINDER Service.
The chamber's maximum operating pressure is 11 bar. Ensure that freshwater and wastewater connections, hoses, and fittings are rated for at least this pressure to prevent leaks or damage. The humidifying system design varies by serial number, so verify the correct connection method before installation.
The chamber is equipped with a temperature safety device class 2/3.3 per DIN 12880. The user must select between temperature limiter and temperature protection modes and set limit and offset values. Incorrect configuration can lead to either nuisance alarms or inadequate protection for sensitive samples.
Selection & Integration Q&A
Choose the KBF 1600 when you need the largest interior volume (1,610 L) and can accommodate its 1,795 mm width. It supports 40 kg per rack and 640 W heat compensation, making it suitable for high-throughput testing or large batches. Smaller siblings like the KBF 1060 or 470 are preferable when floor space is limited or lower capacity suffices.
Freshwater and wastewater connections must be rated for at least 11 bar maximum operating pressure. The humidifying system design differs by serial number (up to or from 20250000010225), so verify the correct connection method and spare parts for your chamber. Use hoses and fittings that meet or exceed this pressure rating.
The chamber requires a wall spacer to maintain proper distance from walls, and it is IP 20 rated so it must not be sprayed with water. Ambient dust increases condenser cleaning frequency. The chamber must be installed on a level surface and connected to a 220–240 V, 1~, 50/60 Hz supply with a 16 A fuse.
Annual maintenance by authorized personnel is mandatory, including steam humidifier service and twice-yearly condenser cleaning. The base delivery does not include APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, flexible Pt 100 sensor, external freshwater/wastewater cans, or BINDER Pure Aqua Service; these must be ordered separately.