Elmo Rietschle V-DTR
Elmo Rietschle V-DTR — Central Industrial & Medical Vacuum Systems with Max differential pressure 1000 mbar, Motor power 6.5 kW, Noise level 79 dB.
The V-DTR is a single-stage, dry-running rotary vane compressor with an eccentrically mounted rotor and vanes, driven by a flanged three-phase motor via a coupling. This modification delivers 100 m³/h at 50 Hz (120 m³/h at 60 Hz) and up to 0.8 bar differential pressure (1.0 bar at 60 Hz), positioning it as the mid-capacity unit in the V-DTR 100/140 series. It is intended for continuous oil-free air supply in central vacuum/pressure systems where 30–90% relative humidity air is conveyed.
Full specifications
| Max differential pressure | 0.8 bar (50 Hz) 60 Hz: 1.0 bar |
| Motor power | 4.0 kW (50 Hz) 60 Hz: 6.5 kW |
| Noise level | 75 dB (50 Hz) 60 Hz: 79 dB |
| Pumping speed | 100 m³/hour (50 Hz) 60 Hz: 120 m³/hour |
| Rotational speed | 1455 rpm (50 Hz) 60 Hz: 1740 rpm |
| Capacity mfr | 100 m3/h (50 Hz) |
| Capacity mfr | 100 m3/h (60 Hz) |
| Overpressure mfr | 0.8 bar (50 Hz) |
| Overpressure mfr | 1.0 bar (60 Hz) |
| Motor efficiency mfr | 88.6 % (50 Hz) |
| Motor efficiency mfr | 91.7 % (60 Hz) |
| Volumenstrom | Capacity (alt 50 Hz) mfr | 130 m**3/hour (50 Hz) |
| Volumenstrom | Capacity (alt 60 Hz) mfr | 155 m**3/hour (60 Hz) |
| Überdruck | Overpressure (alt 50 Hz) mfr | 1,4 bar (50 Hz) |
| Überdruck | Overpressure (alt 60 Hz) mfr | 1,2 bar (60 Hz) |
| speed | Speed (alt 50 Hz) mfr | 1450 rpm (50 Hz) |
| Weight | 148 kg |
| Max. weight | Max. weight (alt var) mfr | 174 kg |
| Current | 13.9 / 8.0 A (50 Hz) 60 Hz: 12.8 / 7.4 A |
| Voltage | 230 / 400 V ± 10 % V (50 Hz) 60 Hz: 380 / 660 V ± 10 % V |
| Motor rating mfr | 6.5 kW (60 Hz) |
| Rated voltage mfr | 380/660 V (60 Hz) |
| Voltage tolerance mfr | ±10 % |
| Current drawn mfr | 12.8/7.4 A (60 Hz) |
| Bemessungsspannung | Rated voltage (alt 50 Hz) mfr | 400 / 690 V ± 10 % V (50 Hz) |
| Motorleistung | Motor rating (alt 50 Hz) mfr | 7,5 kW (50 Hz) |
| Motorleistung | Motor rating (alt 60 Hz) mfr | 9,0 kW (60 Hz) |
| Stromaufnahme | Current drawn (alt 50 Hz) mfr | 14,4 / 8,36 A (50 Hz) |
| Stromaufnahme | Current drawn (alt 60 Hz) mfr | 17,8 / 10,2 A (60 Hz) |
| Protection rating | IP 55 |
| Operating temperature | F |
| Average noise level mfr | 79 dB(A) (60 Hz) |
| Mittlerer Schalldruckpegel | Average noise level (alt 50 Hz) mfr | 77 dB (50 Hz) |
Description
Construction and series positioning
- Single-stage, dry-running rotary vane compressor: eccentric rotor with vanes creates periodically changing compression chambers.
- Standard flanged three-phase motor (4.0 kW at 50 Hz, 6.5 kW at 60 Hz) connected via coupling; IP 55 protection, insulation class F.
- Integrated pressure regulating valve, fine micro intake filter, pressure-side carbon dust filter, noise-reducing plastic cover, and fan cooling.
- Within the V-DTR 100/140 series, this model is the 100 m³/h (50 Hz) variant; the V-DTR 140 offers higher flow (140 m³/h at 50 Hz) and higher motor power (7.5 kW at 50 Hz).
Behavior under load and media
- Delivers oil-free air at pressures from 0 to the data-plate maximum (0.8 bar at 50 Hz, 1.0 bar at 60 Hz).
- Continuous operation (S1) is permissible; flow depends on pressure per data sheet D361.
- Conveyed air relative humidity must be 30–90%; ambient and inlet temperature range 5–40°C.
- Back pressure on outlet side must not exceed +0.1 mbar; higher back pressure constitutes inadmissible operating mode.
- Output reduction noticeable when installed above 1000 m above sea level.
Installation, connection and control
- Requires horizontal mounting with max. 10° inclination, secured against tipping; at least 40 cm clearance around machine for maintenance.
- Cooling air inlets/outlets must be at least 10 cm from walls; recirculation of exhaust air must be avoided.
- Pipes screwed in by hand only; flexible connections recommended. For pipe lengths >3 m with same cross-section, a non-return valve (accessory ZRK) is recommended to prevent reverse running after stop.
- For parallel operation of multiple machines, an external non-return valve must be installed in the inlet line upstream of each machine.
- Electrical installation by qualified electrician per EN 60204; main switch and motor protection switch to be provided by operating company. Voltage tolerance ±10%, frequency tolerance ±2%.
- Rotation direction: anti-clockwise (viewed at motor fan); check with phase sequence indicator. Brief start (max. 2 s) with outlet line disconnected for verification.
- Pressure control valve adjusted per symbol plate; exceeding permissible final compression pressure can damage compressor. Operation without standard control valve is not intended use.
Integration & Operating Notes
The machine must be set up horizontally on a level floor able to carry its weight, with no external vibrations, and secured against tipping or falling. At least 40 cm free space around the machine is required for pipe installation and maintenance access. Cooling air inlets and outlets must be at least 10 cm from adjacent walls, and recirculation of hot exhaust air must be prevented to avoid overheating.
Selection & Integration Q&A
Choose the V-DTR 100 when your required flow is around 100 m³/h at 50 Hz and differential pressure up to 0.8 bar. The V-DTR 140 delivers 140 m³/h at 50 Hz and 7.5 kW motor power, so select it for higher flow or power demands. The V-DTR 100 is the more compact, lower-power option for mid-range duties.
Pipes must be screwed in by hand only; use flexible connections if necessary. For connection pipes with the same cross-section as the machine connection and longer than 3 m, a non-return valve (accessory ZRK) is recommended to prevent reverse running after stop. For parallel operation of multiple machines, an external non-return valve must be installed in the inlet line upstream of each machine.
Mount horizontally on a level floor with max. 10° inclination in all directions, secured against tipping. Allow at least 40 cm clearance around the machine for maintenance, and keep cooling air inlets/outlets at least 10 cm from walls. Ambient and inlet temperature must be 5–40°C; avoid recirculation of hot exhaust air. Above 1000 m altitude, output reduction is noticeable—consult manufacturer.
The base delivery does not include the main switch and motor protection switch (to be provided by the operating company), nor the non-return valve ZRK (accessory) recommended for long outlet pipes. Spare parts such as filter cartridges (replace every 6 months), vanes (check at 3,000 h), and bearings (A bearing at 20,000 h) are consumables. A phase sequence indicator is needed for rotation check.