Eppendorf Centrifuge 5424 R - Microcentrifuge
Eppendorf Centrifuge 5424 R - Microcentrifuge — Centrifuges.
The Eppendorf Centrifuge 5424 R is a benchtop microcentrifuge with a 350 W motor driving a 100–15,000 rpm range and a maximum RCF of 21,130 g. It accepts four rotor types and reaches set speed in 15 s with 16 s deceleration. Its 29 × 48 × 26 cm footprint and 21 kg weight suit it to benchtop sample preparation in molecular biology workflows.
Full specifications
| Acceleration Time 1) | 15 s |
| Deceleration Time 1) | 16 s |
| Dimensions (w × D × H) | 29 x 48 x 26 cm |
| Max. Rcf | 21130 g |
| Rotational speed | 100...15000 rpm |
| Rotors available mfr | 4 |
| Height (with Open Lid) | 51 cm |
| Product Weight | 21 kg |
| Max. Power Consumption | 350 W |
Description
Construction and positioning
- Benchtop microcentrifuge with a 350 W motor; overall dimensions 29 × 48 × 26 cm (W × D × H), 21 kg, and 51 cm height with the lid open.
- Four rotor types are supported, allowing the same unit to cover different tube formats; no sibling models are defined in the supplied facts, so positioning is limited to the rotor set.
Performance behavior
- Speed is adjustable from 100 to 15,000 rpm, with a maximum RCF of 21,130 g.
- Acceleration to set speed takes 15 s and deceleration 16 s, giving predictable run timing for short spin protocols.
Installation and operation
- The unit is floor- or bench-standing at 21 kg; the 51 cm open-lid height must be allowed for above the working surface.
- No control, interface, or connection details are present in the supplied facts; the only attached document is a safety data sheet for Pivot Grease (order 5810 350.050), which contains no equipment operating instructions.
Integration & Operating Notes
The unit stands 26 cm high closed but reaches 51 cm with the lid open. Benches under shelves, wall cabinets, or fume-hood sashes must provide at least that clearance, otherwise the lid cannot be fully opened for rotor access and loading.
The drive operates between 100 and 15,000 rpm. Protocols requiring very low-speed gentle mixing below 100 rpm cannot be set on this unit; the lowest usable setpoint is 100 rpm.
The stated 21,130 g maximum applies to the rotor configuration that achieves it; with the other of the four supported rotor types the achievable RCF will differ. Confirm the rotor-specific g value before setting protocols that rely on the maximum.
A safety data sheet is supplied for Pivot Grease, a petrolatum-based lubricant, indicating it is the specified grease for pivot points. Keep the SDS on file and follow its handling guidance when servicing rotor pivots; the grease is a separate order item.
Selection & Integration Q&A
Choose it when you need up to 15,000 rpm and 21,130 g in a 29 × 48 × 26 cm benchtop footprint with four rotor options. No sibling models are defined in the supplied data, so compare rotor capacity and g-force against other series before deciding.
The unit supports four rotor types, and the 21,130 g maximum is rotor-dependent. Specify the rotor for your tube format at order time and verify the rotor-specific RCF, since the base unit alone does not define which tubes fit.
Allow 51 cm height with the lid open above a 29 × 48 cm footprint, and support 21 kg. No ingress-protection or ambient-temperature ratings are given in the supplied facts, so confirm lab climate limits with Eppendorf before placing it near heat sources.
Rotors are separate selections among the four supported types, and Pivot Grease (order 5810 350.050) is a separate order item for pivot lubrication. No cables, controllers, or maintenance intervals are listed in the supplied facts; confirm with Eppendorf.