Eppendorf Centrifuge 5425/5425 R - 24-place lab standard
Eppendorf ·Made in Germany · In Active Production

Eppendorf Centrifuge 5425/5425 R - 24-place lab standard

Eppendorf Centrifuge 5425/5425 R - 24-place lab standard — Centrifuges.

The Centrifuge 5425 is a non-refrigerated 24-place microcentrifuge with a 280 W motor, reaching 15,060 rpm and 21,300 × g with fixed-angle rotors (3,217 × g with swing-bucket). It is the ambient-temperature member of the 5425/5425 R pair, distinguished from the refrigerated 5425 R by the absence of cooling. It is used for high-speed pelleting of 1.5/2.0 mL tubes and 5.0 mL tubes in molecular biology workflows.

Full specifications

Core Performance & Functional Specifications
Acceleration Time 1) 15 s
Deceleration Time 1) 15 s
Max. Rcf 21300 g
Max. Rcf With Fixed-angle Rotor 21300 g
Max. Rcf With Swing-bucket Rotor 3217 g
Remote Device Monitoring And Notifications (visionize ® ) VisioNize ® box enabled
Rotational speed 100...15060 rpm
Soft Brake Function 10 ramps
Rotors available mfr 6
Max. capacity mfr 10 × 5.0 mL
Timer mfr 10 s to 9:59 h, with continuous run function
Dimensions (W × D × H) mfr 24 × 39 × 24 cm / 9,5 × 15,4 × 9,5 in
Dimensions, Weight & Mechanical Design
Height (with Open Lid) 48 cm
Product Weight 15.6 kg
Electrical, Power & Drive Specifications
Max. Power Consumption 280 W
Power supply mfr 230 V, 50 – 60 Hz
Interfaces, Connections & Control
Power Plug power plug type F
Operating Conditions, Environment & Protection
Cooling non-refrigerated
Noise level 51 dB

Description

Construction and series positioning

Behavior and control

Installation and operation

Integration & Operating Notes

Electrical
230 V type F supply; confirm receptacle compatibility

The unit is rated 230 V with a type F (Schuko) plug. Installations in regions using type B, G or I outlets require an adapter or receptacle change; the 280 W motor draw should be accounted for on shared laboratory circuits.

Factory Specification Sheet
Process Limits
RCF depends on the installed rotor, not the drive alone

Maximum 21,300 × g applies only with fixed-angle rotors; swing-bucket operation is limited to 3,217 × g. Verify the rotor’s own rating before setting speed, since the drive can command 15,060 rpm regardless of rotor limits.

Factory Specification Sheet
Control & I/O
VisioNize box enables remote monitoring

The 5425 is VisioNize box enabled, allowing run status and notifications to be monitored remotely. Network provisioning and the VisioNize box itself are separate from the centrifuge and must be planned with the lab IT setup.

Factory Specification Sheet
Maintenance
Rotors and adapters are separate order items

The base delivery does not include a rotor; select the appropriate rotor (e.g., FA-24×2, FA-10×5, F-32×0.2-PCR) and adapters for the tube formats used. Rotor inspection and cleaning follow the rotor’s own instructions.

Centrifugation Brochure Centrifuge 5425 R

Selection & Integration Q&A

When should I choose the 5425 instead of the refrigerated 5425 R?

Choose the 5425 for ambient-temperature pelleting where 21,300 × g and 15,060 rpm suffice and cooling is unnecessary. Select the 5425 R when sub-ambient temperature control is required; it is the refrigerated member of the same platform.

What rotor and tube-format constraints apply when configuring the 5425?

Six rotors are supported: fixed-angle FA-24×2, FA-24×2-PTFE, FA-10×5 and FA-18×2-Kit; PCR F-32×0.2-PCR and S-96×0.2-PCR. Max capacity is 10 × 5.0 mL; swing-bucket use caps RCF at 3,217 × g versus 21,300 × g fixed-angle.

What is not included in the base delivery and must be ordered separately?