Eppendorf Centrifuge 5425/5425 R - 24-place lab standard
Eppendorf Centrifuge 5425/5425 R - 24-place lab standard — Centrifuges.
The Centrifuge 5425 is a non-refrigerated 24-place microcentrifuge with a 280 W motor, reaching 15,060 rpm and 21,300 × g with fixed-angle rotors (3,217 × g with swing-bucket). It is the ambient-temperature member of the 5425/5425 R pair, distinguished from the refrigerated 5425 R by the absence of cooling. It is used for high-speed pelleting of 1.5/2.0 mL tubes and 5.0 mL tubes in molecular biology workflows.
Full specifications
| Acceleration Time 1) | 15 s |
| Deceleration Time 1) | 15 s |
| Max. Rcf | 21300 g |
| Max. Rcf With Fixed-angle Rotor | 21300 g |
| Max. Rcf With Swing-bucket Rotor | 3217 g |
| Remote Device Monitoring And Notifications (visionize ® ) | VisioNize ® box enabled |
| Rotational speed | 100...15060 rpm |
| Soft Brake Function | 10 ramps |
| Rotors available mfr | 6 |
| Max. capacity mfr | 10 × 5.0 mL |
| Timer mfr | 10 s to 9:59 h, with continuous run function |
| Dimensions (W × D × H) mfr | 24 × 39 × 24 cm / 9,5 × 15,4 × 9,5 in |
| Height (with Open Lid) | 48 cm |
| Product Weight | 15.6 kg |
| Max. Power Consumption | 280 W |
| Power supply mfr | 230 V, 50 – 60 Hz |
| Power Plug | power plug type F |
| Cooling | non-refrigerated |
| Noise level | 51 dB |
Description
Construction and series positioning
- The 5425 is the non-refrigerated variant of the 5425/5425 R platform; the 5425 R adds refrigeration and is documented separately (230 V / 50–60 Hz, keypad or rotary knob, supplied without rotor or with FA-24×2).
- This model is a 24-place, 10 × 5.0 mL-capacity ambient microcentrifuge with a 280 W motor, 230 V supply and type F power plug.
- Six rotor options are supported, spanning fixed-angle (FA-24×2, FA-24×2-PTFE, FA-10×5, FA-18×2-Kit), PCR strip (F-32×0.2-PCR, S-96×0.2-PCR) and swing-bucket formats; maximum RCF is rotor-dependent (21,300 × g fixed-angle, 3,217 × g swing-bucket).
Behavior and control
- Speed is adjustable from 100 to 15,060 rpm; acceleration and deceleration each take 15 s on ramp 1, with 10 soft-brake ramps available for delicate samples.
- The timer spans 10 s to 9:59 h; noise is 51 dB.
- The unit is VisioNize box enabled for remote monitoring and notifications.
Installation and operation
- Footprint is 24 × 39 × 24 cm (W × D × H), 48 cm with the lid open, and weight is 15.6 kg.
- No temperature control is provided (temperature range “−”); the 5425 R should be selected where sub-ambient running is required.
Integration & Operating Notes
The unit is rated 230 V with a type F (Schuko) plug. Installations in regions using type B, G or I outlets require an adapter or receptacle change; the 280 W motor draw should be accounted for on shared laboratory circuits.
Maximum 21,300 × g applies only with fixed-angle rotors; swing-bucket operation is limited to 3,217 × g. Verify the rotor’s own rating before setting speed, since the drive can command 15,060 rpm regardless of rotor limits.
The 5425 is VisioNize box enabled, allowing run status and notifications to be monitored remotely. Network provisioning and the VisioNize box itself are separate from the centrifuge and must be planned with the lab IT setup.
The base delivery does not include a rotor; select the appropriate rotor (e.g., FA-24×2, FA-10×5, F-32×0.2-PCR) and adapters for the tube formats used. Rotor inspection and cleaning follow the rotor’s own instructions.
Selection & Integration Q&A
Choose the 5425 for ambient-temperature pelleting where 21,300 × g and 15,060 rpm suffice and cooling is unnecessary. Select the 5425 R when sub-ambient temperature control is required; it is the refrigerated member of the same platform.
Six rotors are supported: fixed-angle FA-24×2, FA-24×2-PTFE, FA-10×5 and FA-18×2-Kit; PCR F-32×0.2-PCR and S-96×0.2-PCR. Max capacity is 10 × 5.0 mL; swing-bucket use caps RCF at 3,217 × g versus 21,300 × g fixed-angle.
The centrifuge is supplied without a rotor; rotors (FA-24×2, FA-10×5, F-32×0.2-PCR, etc.) and adapters are separate items. The VisioNize box for remote monitoring is also a separate accessory.