Eppendorf Centrifuge 5702/ 5702 R/5702 RH - Low-Speed Centrifugation
Eppendorf Centrifuge 5702/ 5702 R/5702 RH - Low-Speed Centrifugation — Centrifuges.
The Centrifuge 5702 is the non-refrigerated base model of the 5702 family, a low-speed benchtop centrifuge with a 100–4,400 rpm range and 3,052 × g maximum RCF using swing-bucket rotors. It accepts six rotor types and up to 4 × 100 mL, and is distinguished from the R and RH siblings by the absence of active temperature control.
Full specifications
| Acceleration Time 1) | 19 s |
| Blood Collection Tubes | 36 mm |
| Conical Tubes | 20 mL |
| Deceleration Time 1) | 19 s |
| Dimensions (w × D × H) | 32 × 40 × 24 cm |
| Max. Rcf | 3052 g |
| Max. Rcf With Fixed-angle Rotor | 1900 g |
| Max. Rcf With Swing-bucket Rotor | 3052 g |
| Rotational speed | 100...4400 rpm |
| Soft Brake Function | yes |
| Temperature range | − |
| Rotors available mfr | 6 |
| Max. capacity mfr | 4 × 100 mL |
| Timer mfr | 0 – 99 min, with continuous run function |
| Footprint (dimensions W/o Front Panel, Wxd) | 32 cm |
| Height (with Open Lid) | 53 cm |
| Product Weight | 19 kg |
| Max. Power Consumption | 200 W |
| Voltage | 230 V |
| Power Plug | power plug type F |
| Cooling | non-refrigerated |
| Noise level | 52 dB |
Description
Construction and series positioning
- The 5702 is the ambient-temperature member of the 5702/R/RH family; the R adds refrigeration and the RH adds heating and refrigeration. This model has no temperature-control system (temperature range listed as '−').
- Benchtop footprint 32 × 40 cm, height 24 cm closed and 53 cm with the lid open; unit weight 19 kg.
- Six rotors are supported. Maximum RCF is rotor-dependent: 3,052 × g with swing-bucket rotors and 1,900 × g with fixed-angle rotors.
- Maximum capacity is 4 × 100 mL; the unit accommodates 20 mL conical tubes and blood-collection tubes up to 36 mm.
Run behavior
- Speed is settable from 100 to 4,400 rpm. Acceleration and deceleration each take 19 s to/from set speed under the stated conditions.
- A soft-brake function is provided, and the timer covers 0–99 min plus a continuous-run mode.
- Noise is stated as 52 dB overall and <52 dB(A) with Rotor A-4-38.
Electrical and installation
- Powered from 230 V, 50–60 Hz with a type F (Schuko) power plug; maximum power consumption 200 W.
- No control interface, communication protocol, or external I/O is listed in the supplied facts; operation is via the unit's own panel and timer.
- The supplied documentation digest contains no additional installation, connection, or operating instructions beyond the brochure scope statement.
Integration & Operating Notes
The unit is specified for 230 V, 50–60 Hz with a type F (Schuko) plug and 200 W maximum consumption. Installations on 115 V mains or with non-type-F receptacles require an adapter or a different supply arrangement; the supplied facts list no alternative voltage.
Swing-bucket rotors reach 3,052 × g while fixed-angle rotors are limited to 1,900 × g. Programming a setpoint above the installed rotor's limit is not supported by the stated data; verify the rotor in use before setting speed.
The 5702 is the non-refrigerated variant; the temperature range field is listed as '−'. Runs whose protocol depends on holding a set temperature (e.g. 4 °C) require the 5702 R or 5702 RH sibling rather than this unit.
Selection & Integration Q&A
Choose the 5702 when runs are at ambient temperature and no setpoint cooling or heating is needed; it is the non-refrigerated member. If the protocol requires holding a temperature, the R (refrigerated) or RH (heated and refrigerated) sibling is required, since this model's temperature range is listed as '−'.
Six rotors are supported, including swing-bucket (up to 3,052 × g) and fixed-angle (up to 1,900 × g) types. Maximum capacity is 4 × 100 mL, and the unit accepts 20 mL conical tubes and blood-collection tubes up to 36 mm. Select the rotor before setting speed, because the RCF ceiling follows the rotor.
The supplied documentation digest does not specify a maintenance schedule, service intervals, or which rotors, buckets, adapters, and cables ship with the base unit. Rotors and buckets are catalogued separately in the brochure, so confirm the rotor/bucket configuration and any adapters at order time; no service-interval data is available in the provided facts.