Eppendorf Centrifuge CR22N - High-Speed Centrifuge
Eppendorf Centrifuge CR22N - High-Speed Centrifuge — Centrifuges.
The Eppendorf CR22N is a refrigerated floor-standing high-speed centrifuge driven by a 3.2 kW induction motor, reaching 22,000 rpm and 58,700 × g with the R27A fixed-angle rotor. Its rotary inverter compressor holds −20 to 40 °C across 6 × 1,000 mL or 4 × 1,500 mL capacity, with nine acceleration/deceleration profiles and 30 program memories.
Full specifications
| Acceleration/deceleration Mode | 9 |
| Continuous Flow Rotor | no |
| Drive Unit | Induction Motor |
| Economy Mode | yes |
| Interfaces | USB port |
| Max. Rcf | 58700 g |
| Number Of Programs | 30 |
| Operating Log Management | 5 |
| Refrigerator | Rotary inverter driven compressor |
| Rotational speed | 300...22000 rpm |
| Safety Devices | Door interlock, overspeed detection, imbalance detection, rotor cover detection, over-temperature detection |
| Temperature range | -20...40 °C |
| User Lockout Function | yes |
| Vacuum System | no |
| Variable Deceleration Slope | yes |
| Max. capacity mfr | 6 × 1,000 mL/4 × 1,500 mL |
| Separation example mfr | High centrifugation speed of up to 58,700 x g wi |
| Power requirement mfr | Single phase: AC200/208/220/230/240 V ±10%, 30 A |
| Footprint (dimensions W/o Front Panel, Wxd) | 700 mm |
| Height (with Open Lid) | 1510 mm |
| Product Weight | 260 kg |
| Max. Power Consumption | 3200 W |
| Power supply mfr | 230 V, 50/60 Hz |
| Keypad | Touch sensitive LCD panel |
| Cooling | refrigerated |
| Noise level | 64 dB |
Description
Construction and positioning
- Floor-standing refrigerated unit with a 700 × 760 × 915 mm footprint (1510 mm with lid open) and 260 kg product weight; the 700 mm width excludes the front panel.
- Induction motor drive (3.2 kW) with a rotary inverter-driven refrigeration compressor; no vacuum system and no continuous-flow rotor capability, distinguishing it from ultracentrifuge-class machines.
- Maximum capacity 6 × 1,000 mL or 4 × 1,500 mL; maximum RCF 58,700 × g at 22,000 rpm with the R27A fixed-angle rotor.
Operating behavior
- Speed range 300–22,000 rpm with nine acceleration/deceleration modes and a variable deceleration slope, allowing separation protocols to be tuned for pellet integrity.
- Temperature control from −20 to 40 °C via the inverter compressor; economy mode is available for reduced refrigeration demand during idle or low-load periods.
- Safety chain includes door interlock, overspeed detection, imbalance detection, rotor cover detection and over-temperature detection; user lockout is provided.
Control and operation
- Touch-sensitive LCD keypad with USB port; 30 program memories and five operating-log management entries.
- Power requirement: single phase AC 200/208/220/230/240 V ±10%, 30 A, 50/60 Hz; nominal supply cited as 230 V, 50/60 Hz.
- Noise level 64 dB.
Integration & Operating Notes
The unit is specified for AC 200/208/220/230/240 V ±10% at 30 A, 50/60 Hz. A 230 V branch sized only for the 3.2 kW motor power will be undersized once refrigeration and acceleration transients are included; verify breaker and cable rating against the 30 A figure before installation.
The 58,700 × g maximum is reached only with the R27A fixed-angle rotor at 22,000 rpm. Other rotors will have lower rated RCF and lower permissible speeds, so protocols must be validated against the specific rotor's rating rather than the instrument maximum.
The safety chain includes door interlock, overspeed, imbalance, rotor cover and over-temperature detection. A run will not start with the lid open or an improperly seated rotor cover, and imbalance detection will abort acceleration; balance tubes by weight before loading to avoid mid-run stops.
Eppendorf epServices offers AdvancedCare, PremiumCare, Warranty 1 and Warranty plus 2 agreements plus Installation Service, IQ/OQ GxP, OQ GxP and Maintenance as on-demand items. These, and any USB cable or rotor, are not part of the base instrument and must be ordered and scheduled separately.
Selection & Integration Q&A
Pick the CR22N for 6 × 1,000 mL / 4 × 1,500 mL volumes at up to 58,700 × g and 22,000 rpm with the R27A rotor, with −20 to 40 °C refrigeration. Ultracentrifuge-class machines are needed only when higher g-force or vacuum operation is required; the CR22N has no vacuum system and no continuous-flow rotor.
A single-phase AC 200/208/220/230/240 V ±10%, 30 A, 50/60 Hz circuit is required; the 3.2 kW motor rating alone does not size the branch. A USB port is the only listed interface, so any logging or LIMS connection needs a host PC and cable supplied separately.
Rotors such as the R27A, USB cable and any GxP documentation are not part of base delivery. Eppendorf epServices supplies Installation Service, IQ/OQ GxP, OQ GxP and Maintenance on demand, plus AdvancedCare, PremiumCare, Warranty 1 and Warranty plus 2 agreements, which must be ordered and scheduled separately.