Eppendorf Mastercycler ® X50 - PCR Thermocycler
Eppendorf Mastercycler ® X50 - PCR Thermocycler — Thermal cyclers (PCR).
The Mastercycler X50 is a 96-well Peltier thermal cycler with a silver thermoblock, six Peltier elements, and a 4–99 °C block range. It offers a 2D gradient over 12 columns and 8 rows (1–30 °C) and four control modes (Fast, Intermediate, Standard, Safe). It is used for PCR amplification and thermal cycling protocols requiring precise temperature uniformity.
Full specifications
| Block Homogeneity 20 – 72 °c | 0.2 °C |
| Control Accuracy | 0.15 °C |
| Gradient Range | 1...30 °C |
| Gradient Temperature Control Range | 30...99 °C |
| Heating Rate | 10 °C/s |
| Heating Technology | Six peltier elements |
| Sample Capacity | 96 |
| Temperature Control Range Of The Block | 4...99 °C |
| Thermoblock | Silver |
| 2D-Gradient block mfr | over 12 columns and 8 rows |
| Temperature Control Mode | Fast, Intermediate, Standard, Safe |
Description
Construction and positioning
- The Mastercycler X50 is a 96-well thermal cycler with a silver thermoblock heated and cooled by six Peltier elements. It accepts 96-well plates and 0.1/0.2 mL tubes; 384-well plates are not supported.
- The block provides a 2D gradient spanning 12 columns and 8 rows, enabling column- and row-wise temperature optimization within a single run.
- No sibling models are listed for this series, so the X50 stands as the sole 96-well configuration in the provided facts.
Thermal behavior and control modes
- Block temperature control range is 4–99 °C; gradient temperature control range is 30–99 °C with a gradient span of 1–30 °C.
- Heating rate is 10 °C/s. Control accuracy is 0.15 °C, and block homogeneity at 20–72 °C is 0.2 °C; at 95 °C homogeneity is ≤±0.3 °C.
- Four temperature-control modes are available: Fast, Intermediate, Standard, and Safe. These trade speed against uniformity and should be selected according to assay sensitivity.
Networking and operation
- The unit can control up to 9 other units in a network; it cannot itself be controlled remotely (the 'unit can only be controlled' row is marked '–').
- No installation, connection, or control details are provided in the attached document digest, which covers consumables only.
Integration & Operating Notes
The block control range tops out at 99 °C and the gradient control range at 99 °C. Setting higher targets is outside the verified operating envelope and may invalidate results or stress the Peltier elements; keep protocols within 4–99 °C (block) and 30–99 °C (gradient).
The 2D gradient covers 12 columns and 8 rows with a span of 1–30 °C, and gradient control is only available between 30 and 99 °C. Annealing optimization steps below 30 °C cannot use the gradient function; plan column temperatures accordingly.
The X50 can control 9 other units in a network but cannot itself be controlled by another unit. In a multi-instrument setup, designate this unit as the master or verify that the controlling unit supports the required network role before wiring.
The attached consumables brochure lists twin.tec PCR plates (96/384, skirted, semi-skirted, unskirted, divisible), films, foils, mats, and plate lids as separate catalog items. Base delivery does not include plates or sealing consumables; specify them separately for the 96-well format.
Selection & Integration Q&A
Choose the X50 for 96-well plates and 0.1/0.2 mL tubes; it does not support 384-well plates. If your throughput requires 384-well format, this model is not suitable and a 384-well-capable instrument must be selected instead.
The 96-well block accepts twin.tec PCR plates (skirted, semi-skirted, unskirted, divisible) and 0.1/0.2 mL tubes. Sealing films, foils, mats, and plate lids are separate catalog items and are not part of base delivery; order them alongside the instrument.
Yes. The X50 can control up to 9 other units in a network, but it cannot itself be controlled by another unit. It must therefore serve as the master in any networked configuration; verify cabling and master assignment before commissioning.
No maintenance schedule or service-kit details are provided in the supplied documents. The only accessory information available concerns consumables (plates, films, foils, mats, lids), which are catalog items separate from the instrument; confirm service intervals with the manufacturer.