Eppendorf New Brunswick Innova ® 40/40R - Benchtop Orbital Shaker
Eppendorf ·Made in Germany · In Active Production

Eppendorf New Brunswick Innova ® 40/40R - Benchtop Orbital Shaker

Eppendorf New Brunswick Innova ® 40/40R - Benchtop Orbital Shaker — Laboratory shakers.

Benchtop orbital incubator shaker with a triple-eccentric counterbalanced drive and a 46 cm platform, available in refrigerated (40R) and non-refrigerated (40) versions. The 40R holds 15–80 °C with 0.25 °C uniformity and runs 25–500 rpm depending on load. It serves as a stand-alone benchtop incubation and shaking station for cultures and assays.

Full specifications

Core Performance & Functional Specifications
Audible And Visual Alarms yes
Available Program Modes -232
Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) B
Dimensions (w × D × H) 55,9 × 76,2 × 61 cm
Door Opening top opening hood
Drive Mechanism triple-eccentric counterbalanced drive
Memory non volatile with automatic power failure restart
Motor Type Solid state, DC brushless motor
Multi-step Programming yes
Platform Size (cm/in) 46 cm
Programming Steps / Storable Programs 15
Refrigerated yes
Remote Device Monitoring And Notifications (visionize ® ) VisioNize ® box enabled
Slide-out Platform no
Temperature range 15...80 °C
Temperature uniformity 0.25 °C
Timer 0.01 – 99.59 h
Speed range mfr 25 up to 500 rpm (depending on load)
Steps per program mfr 15
Stored programs mfr 4
Capacity mfr medium
Dimensions, Weight & Mechanical Design
External Height (with Lid/door Open) 101.6 cm
Product Weight 79.4 kg
Electrical, Power & Drive Specifications
Voltage 230 V
Interfaces, Connections & Control
Power Plug type F/type I
Standards, Safety & General Data
Cetlus And Ce Marked yes

Description

Construction and positioning

Behavior and operating envelope

Control, connection and operation

Integration & Operating Notes

Mounting
Allow 101.6 cm clear height for the open hood

The lid swings to an external height of 101.6 cm, so benchtop or under-shelf placement must leave that clearance above the 61 cm cabinet; the unit is benchtop-only (not stackable or under-bench) and weighs 79.4 kg, so the bench must carry that load.

Factory Specification Sheet
Electrical
230 V supply with type F or type I plug

The unit is specified at 230 V and ships with a type F/type I power plug; confirm the site socket and circuit match before installation, since the plug type is fixed at delivery and no other voltage is stated for this model.

Factory Specification Sheet
Process Limits
Top speed depends on the load on the platform

The 25–500 rpm range is load-dependent: the 500 rpm ceiling applies only within the load the drive can counterbalance, so heavy or unevenly distributed vessel loads reduce the achievable maximum speed and must be planned for in protocol design.

Factory Specification Sheet
Control & I/O
Power-failure restart resumes the stored program

Non-volatile memory with automatic power-failure restart means the shaker returns to its programmed run after a mains interruption; account for this when the process must not resume unattended, and use the audible/visual alarms to detect fault states.

Factory Specification Sheet
Control & I/O
VisioNize box enables remote monitoring

Remote device monitoring and notifications are provided through the VisioNize box; the box and its connectivity are a separate item from the shaker itself, so specify it explicitly if remote alarm and status visibility are required.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the Innova 40R rather than the non-refrigerated 40?

Pick the 40R when you need controlled temperatures below ambient down to 15 °C; the non-refrigerated 40 cannot cool. Both share the 46 cm platform, 25–500 rpm load-dependent range and 15-step programming, so the choice is purely the cooling requirement.

What supply and connectivity must be provided at the installation site?

A 230 V supply with a matching type F or type I socket; the plug type is fixed at delivery. For remote monitoring and notifications, a VisioNize box must be specified separately, as it is not part of the base shaker.

Can the unit sit on a bench under a shelf, and what load can the drive take?

It is benchtop-only and needs 101.6 cm clear height for the open hood above the 61 cm cabinet, so a low shelf blocks loading. The 79.4 kg unit needs a bench rated for it, and the 500 rpm ceiling applies only within the load the counterbalanced drive can handle.

What maintenance and accessories are not included in the base delivery?

The base unit ships without the VisioNize box for remote monitoring; specify it if needed. No service intervals are stated in the supplied documents, so maintenance scheduling must come from the operating manual rather than this card.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Shakers Brochure Incubator Shakers Incubator Shaker Portfolio
Brochure ·673,8 КБ ·1 pp.
View PDF