Eppendorf New Brunswick Innova ® 44/44R - Stackable Incubator Shaker
Eppendorf New Brunswick Innova ® 44/44R - Stackable Incubator Shaker — Laboratory shakers.
The New Brunswick Innova 44/44R is a floor-standing, stackable incubator shaker built on a triple-eccentric counterbalanced drive with a solid-state DC brushless motor. It covers 10–80 °C with 0.25 °C uniformity and 25–400 rpm at a 2.5 cm orbit, distinguishing it as the economical workhorse of the Innova stackable line. It serves as a high-capacity incubation and shaking platform for microbial and cell culture in shared laboratory infrastructure.
Full specifications
| Audible And Visual Alarms | yes |
| Available Program Modes | -232 |
| Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) | F, S |
| Door Opening | glide up |
| Drive Mechanism | triple-eccentric counterbalanced drive |
| Gassing Manifold | yes |
| Internal Convenience Light | yes |
| Memory | non volatile with automatic power failure restart |
| Motor Type | Solid state, DC brushless motor |
| Multi-step Programming | yes |
| Photosynthetic Lighting | no |
| Platform Size (cm/in) | 76 cm |
| Refrigerated | − |
| Slide-out Platform | yes |
| Temperature range | 10...80 °C |
| Temperature uniformity | 0.25 °C |
| Timer mfr | 0.1 – 99.9 h, continuous |
| Dimensions (W × D × H) mfr | 134,9 × 84,3 × 66 cm / 53,1 × 33,2 × 26 in |
| Speed range mfr | 25 – 400 rpm with 2.5 cm orbit; 25 – 300 rpm wit |
| Steps per program mfr | 15 |
| Stored programs mfr | 4 |
| External Height (with Lid/door Open) | 113.9 |
| Product Weight | 238 kg |
| Power supply mfr | 230 V ± 10 %, 50 Hz |
| Power Plug | type F/type I |
| Water/humidity Monitoring | no |
Description
Construction and series positioning
- Floor-standing, stackable (F, S) incubator shaker with a 76 cm slide-out platform and a door that glides up; external height with the lid open is 113.9 cm.
- Overall dimensions 134.9 × 84.3 × 66 cm and 238 kg product weight; the unit is designed to be stacked, with speed derating for stacked configurations.
- The triple-eccentric counterbalanced drive and solid-state DC brushless motor are the mechanical core; the 44/44R is positioned in the flyer as the economical workhorse relative to the flagship Innova S44i, which adds incubated/refrigerated variants and photosynthetic lighting.
- This model is not refrigerated (refrigerated: −) and has no photosynthetic lighting; temperature control is heating-only over 10–80 °C.
Behavior under load and in different modes
- Speed envelope depends on orbit and stacking: 25–400 rpm with the 2.5 cm orbit; 25–300 rpm with the 5.1 cm orbit or with two stacked units; 25–250 rpm with three stacked units. The engineer must select the orbit and stack height before fixing the process setpoint.
- Temperature uniformity is 0.25 °C across the 10–80 °C range; there is no humidity monitoring and no water/humidity control.
- Programming supports 4 stored programs of 15 steps each, multi-step programming, a timer of 0.1–99.9 h, and non-volatile memory with automatic power-failure restart.
Installation, control and operation
- Electrical: 230 V supply with type F/type I power plug; the unit is a fixed floor installation at 238 kg, so floor loading and access for the glide-up door (113.9 cm open height) must be planned.
- Control and I/O: audible and visual alarms, internal convenience light, gassing manifold, and a slide-out platform for loading vessels.
- The flyer digest provides no further installation or maintenance detail; operating specifics beyond the above are not documented in the supplied facts.
Integration & Operating Notes
The drive is rated 25–400 rpm only as a single unit at the 2.5 cm orbit; two stacked units are limited to 25–300 rpm and three stacked units to 25–250 rpm. Plan the stack configuration before writing process setpoints, because a method validated at 400 rpm cannot be run in a multi-unit stack.
Memory is non-volatile with automatic power-failure restart, so after a mains interruption the shaker restarts the stored program without operator action. Anyone relying on a manual restart as a safety step must account for this behavior, and alarms should be wired to a remote monitoring point if unattended operation is intended.
Water/humidity monitoring is absent and photosynthetic lighting is not fitted; the gassing manifold is the only atmosphere-related interface. Processes requiring controlled humidity or light cycles must use the flagship S44i variant instead, or provide external conditioning.
The flyer digest contains no maintenance schedule, error-code list, or consumable/accessory list for the 44/44R. The user must obtain the full instruction manual for drive service intervals, platform accessories, and alarm-code handling before commissioning; none of these are included in the base delivery documentation provided here.
Selection & Integration Q&A
Pick the 44/44R for heating-only incubation at 10–80 °C with 25–400 rpm shaking and no humidity or light control. Choose the S44i when you need refrigerated operation below ambient, photosynthetic lighting, or the flagship user interface; the 44/44R is the economical workhorse of the stackable pair.
No. Two stacked units are limited to 25–300 rpm and three stacked units to 25–250 rpm; 400 rpm is available only as a single unit at the 2.5 cm orbit. Revalidate any method above the stacked limit or run it in a single unstacked unit.
The supplied flyer documents no maintenance schedule, error-code list, platform clamps, flask holders, or replacement drive parts; these must be ordered separately and the full instruction manual obtained. Plan for periodic drive and platform inspection per that manual, since no interval is stated in the available facts.