Eppendorf SpinPro ® 6 R - Smart High-Performance Centrifuge
Eppendorf SpinPro ® 6 R - Smart High-Performance Centrifuge — Centrifuges.
The SpinPro 6 R is a refrigerated high-performance laboratory centrifuge with a 1600 W motor, reaching 13,500 rpm and 20,020 × g with fixed-angle rotors. Its −11 to 40 °C temperature range and 99-program memory distinguish it as the refrigerated, high-speed member of the SpinPro line. It is used for high-throughput pelleting, plate centrifugation, and temperature-sensitive separations.
Full specifications
| Acceleration/braking Ramps | 9 |
| Act Label | yes |
| Blood Collection Tubes | 60 mm |
| Capacity | 60 gram |
| Conical Tubes | 34 mL |
| Fixed-angle Rotors For High-speed Applications | yes |
| Max. Rcf | 20020 g |
| Max. Rcf With Fixed-angle Rotor | 20020 g |
| Max. Rcf With Swing-bucket Rotor | 4300 g |
| Number Of Programs | 99 |
| Plate Rotor Options For Mtp, Pcr Or Dwp | yes |
| Plates | 10 |
| Rotational speed | 1...13500 rpm |
| Supporting Gmp/glp | 21 |
| Temperature range | -11...40 °C |
| Universal Adapters For Conical Tubes, Plates And Bottles | yes |
| Rotors available mfr | 5 |
| Timer mfr | 1 s to 99 h 59 min 59 s, with continuous run fun |
| Dimensions (W × D × H) mfr | 70,6 × 74,5 × 38,5 cm / 27,8 × 29,3 × 15,2 in |
| Rotor S-2xUniversal ID mfr | ≤62 dB(A) |
| Rotor FA-6x250 ID mfr | ≤62 dB(A) |
| Rotor FA-10x50 ID mfr | ≤65 dB(A) |
| Rotor FA-48x2 ID mfr | ≤70 dB(A) |
| Rotor FA-30x2 ID mfr | ≤70 dB(A) |
| Height (with Open Lid) | 86.7 cm |
| Product Weight | 100 kg |
| Footprint (dimensions w/o front panel, WxD) mfr | 70 × 70 cm, 27.6 × 27.6 in |
| Max. Power Consumption | 1600 W |
| Power supply mfr | 220 – 240 V, 50 – 60 Hz |
| Power Plug | power plug type EF |
| Cooling | refrigerated |
| Noise level | 70 dB |
Description
Construction and positioning
- Refrigerated floor-standing unit with a 70.6 × 74.5 × 38.5 cm footprint (70 × 70 cm without front panel) and 100 kg product weight; height with open lid is 86.7 cm.
- Accepts five rotor types: fixed-angle (FA-6×250 ID, FA-10×50 ID, FA-48×2 ID, FA-30×2 ID) and swing-bucket (S-2×Universal ID), plus plate rotors for MTP, PCR, and DWP formats.
- Maximum RCF is 20,020 × g with fixed-angle rotors and 4,300 × g with swing-bucket rotors; speed range is 1–13,500 rpm.
Behavior under load and media
- Temperature control spans −11 to 40 °C, allowing runs below ambient for heat-sensitive samples.
- Capacity facts include 60 g, 34 mL conical tubes, 60 mm blood-collection tubes, and 10 plates; universal adapters accept conical tubes, plates, and bottles.
- Noise output is rotor-dependent: ≤62 dB(A) with S-2×Universal ID and FA-6×250 ID, ≤65 dB(A) with FA-10×50 ID, and ≤70 dB(A) with FA-48×2 ID and FA-30×2 ID.
Control and operation
- Nine acceleration/braking ramps and 99 programs; timer ranges from 1 s to 99 h 59 min 59 s with continuous run, short-spin, and timer-programmed start or stop.
- Powered from 220–240 V, 50–60 Hz with a type EF power plug; motor power is 1600 W.
- The brochure covers the centrifuge and listed rotors/adapters but does not print ordering information for the centrifuge itself.
Integration & Operating Notes
The unit is rated only for 220–240 V at 50–60 Hz and draws 1600 W; connecting to a 110–120 V circuit or using an incompatible receptacle will prevent operation and may damage the drive. Confirm the site provides the type EF outlet and a circuit sized for the full motor power.
The bench or floor location must accommodate the 70.6 × 74.5 cm footprint and provide at least 86.7 cm of vertical clearance when the lid is open, plus access for loading and unloading rotors. The 100 kg weight requires a level, load-bearing surface.
Maximum RCF is 20,020 × g only with fixed-angle rotors; swing-bucket operation is limited to 4,300 × g. Exceeding the rotor-specific limit risks rotor failure and invalidates separation results, so the run setpoint must be checked against the installed rotor type.
Temperature control is limited to −11…40 °C; runs requiring lower temperatures or operation outside this band are not supported. Samples and adapters should be equilibrated within this range before starting a temperature-sensitive protocol.
Selection & Integration Q&A
Choose the 6 R when you need both sub-ambient cooling (−11 to 40 °C) and high-speed fixed-angle runs up to 20,020 × g at 13,500 rpm. If runs are ambient-only or limited to swing-bucket work at 4,300 × g, a non-refrigerated or swing-bucket-only unit is sufficient.
Use the S-2×Universal ID swing-bucket rotor with universal adapters 1–3 for conical tubes, plates, and bottles; plate rotors accept MTP, PCR, and DWP formats (10 plates). Fixed-angle rotors FA-6×250 ID, FA-10×50 ID, FA-48×2 ID, and FA-30×2 ID cover high-speed tube work.
Provide a 220–240 V, 50–60 Hz supply with type EF plug and 1600 W capacity, a level surface for the 100 kg unit, and 86.7 cm vertical clearance for the open lid. Keep ambient and sample temperatures within the −11 to 40 °C control range.
The brochure lists rotors S-2×Universal ID, FA-48×2 ID, FA-30×2 ID, FA-6×250 ID, FA-10×50 ID and universal adapters 1–3 as separate items; ordering information for the centrifuge itself is not printed and must be obtained from the local sales representative. No service interval is stated in the supplied documents.