Peter Huber Kältemaschinenbau CC-200BX Brückenthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-200BX Brückenthermostat

Peter Huber Kältemaschinenbau CC-200BX Brückenthermostat — Thermostatic baths.

The CC-200BX is a bridge thermostat with a telescopic arm (up to 884 mm) and 128 mm immersion depth, designed to clamp onto an open bath. Its 200–240 V single- or two-phase unit delivers 1.5–2.1 kW heating and reaches 28–200 °C uncooled, 20–200 °C with water cooling, or -20–200 °C with a refrigerator. It circulates thermofluid through external circuits via a variable-speed pressure/suction pump.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Brueckenthermostate Series — Model Comparison & Selection

Model Current consumption, ADimensions w d hMax delivery flow rate, liter/minuteTemperature range, °CWeight, kgAmbient temperature limit, °C
CC-200BX Brückenthermostat Current 10345 x 200 x 3262228–20075–40
CC-300BX Brückenthermostat 16345 x 190 x 3922528–300135–40

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 345 x 200 x 326 mm
Maximum delivery flow rate 27 l/min (pressure pump max.; pressure 0, 7 bar)
suction pump max.; pressure 0, 4 bar: 22 l/min
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 28...200 °C
with cooling unit: -20...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 5 - 2, 1 kW
Dimensions, Weight & Mechanical Design
Weight 7 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 200-240V 1~/2~ 50/60Hz V; Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Provide UL 489 branch protection and a 10–16 A fuse

The unit draws up to 10 A at 200–240 V and requires external branch-circuit protection according to UL 489 with a fuse rated between 10 A and 16 A. Verify the site supply and protection device before installation; the country-specific plug must be specified at ordering.

CC-200BX Cooling Thermostat Datasheet, page 2
Electrical
Do not combine -10% voltage with +3% frequency

Voltage and frequency tolerances are ±10% only if they do not run in the opposite direction. A supply at -10% voltage together with +3% frequency is outside the permitted envelope and must be corrected before commissioning.

CC-200BX Cooling Thermostat Datasheet, page 2
Process Limits
Sub-ambient operation needs water cooling or a refrigerator

Without cooling the operating range starts at 28 °C; water cooling extends it to 20 °C and a refrigerator to -20 °C. Select the cooling configuration against the lowest required process temperature before specifying the unit.

CC-200BX Cooling Thermostat Datasheet, page 1
Cooling & Media
Keep thermofluid viscosity at or below 50 mm²/s

The maximum permissible kinematic viscosity is 50 mm²/s. At low temperatures viscosity rises, so the fluid must be chosen so that it stays within this limit across the whole operating range; otherwise pump delivery and heat transfer degrade.

CC-200BX Cooling Thermostat Datasheet, page 1
Mounting
Check bath depth against 128 mm immersion depth

The immersion depth is 128 mm and the telescopic arm reaches up to 884 mm. Confirm that the bath geometry allows the heater and sensor to be immersed to this depth and that the bridge arm spans the bath opening.

CC-200BX Cooling Thermostat Datasheet, page 1
Process Limits
External circuit must fit 27 l/min at 0.7 bar

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar. Calculate the pressure drop of the connected external loop; if it exceeds the pump capability, flow falls below the required rate.

CC-200BX Cooling Thermostat Datasheet, page 1
Environment
Install only in a dry indoor location, 5–40 °C

IP20 protection and an ambient limit of 5–40 °C rule out wet or dusty locations. Class A Group 1 to EN55011 means additional EMC measures may be needed in residential or light-industrial environments.

CC-200BX Cooling Thermostat Datasheet, pages 1–2
Maintenance
Set overtemperature protection to DIN 12876 before use

The unit has adjustable overtemperature protection according to DIN 12876 and safety class III/FL. Adjust the limit to the bath fluid and process before unattended operation; the stainless-steel wetted parts and plastic pumps define the cleaning and fluid-compatibility practice.

CC-200BX Cooling Thermostat Datasheet, page 1

Selection & Integration Q&A

When should I choose the CC-200BX rather than the CC-300BX?

Pick the CC-200BX when the existing bath matches its 128 mm immersion depth and 884 mm telescopic arm reach. The CC-300BX is the neighbor for larger bath geometries; confirm the arm span and immersion depth against your bath before deciding.

What upstream power and protection does the CC-200BX require?

A 200–240 V, 1~ or 2~, 50/60 Hz supply with max. 10 A, external branch-circuit protection to UL 489, and a fuse of 10–16 A. The country-specific plug must be specified at ordering.

What site and duty limits apply to installation?

Ambient 5–40 °C, IP20, dry indoor location only. Voltage and frequency tolerances of ±10% apply only if they do not run in opposite directions. Class A Group 1 to EN55011 may require extra EMC measures in light-industrial settings.

What is not included in the base delivery that I must plan for?

The power cable and country-specific plug must be specified at ordering. External branch-circuit protection to UL 489 is site-provided. The external Pt100 sensor, thermofluid, and cooling water or refrigerator for sub-ambient operation are separate.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-200BX Cooling Thermostat Datasheet
Datasheet ·54,6 КБ ·2 pp.
View PDF