Labconco Kjeldahl Distillation Apparatus
Labconco · In Active Production

Labconco Kjeldahl Distillation Apparatus

Labconco Kjeldahl Distillation Apparatus — Kjeldahl & extraction apparatus.

The Labconco Kjeldahl Distillation Apparatus is a steam-distillation unit for ammonia determination in digested samples. This 18-place configuration is the largest hooded combination variant in the series, integrating digestion and distillation in one bench unit. It serves laboratories running high-throughput nitrogen/protein analysis where fume containment and multi-sample parallelism are required.

Full specifications

Core Performance & Functional Specifications
Number of places mfr 18 places
CFM mfr 1000 CFM
Electrical requirements mfr 115 phase
Dimensions, Weight & Mechanical Design
Shipping weight mfr 361 lbs
Electrical, Power & Drive Specifications
Amps mfr 32 A
KW mfr 21 kW

Description

Series positioning and construction

Capacity and utility envelope

Operation

Integration & Operating Notes

Electrical
Dedicated 32 A / 21 kW service required; verify panel capacity

The unit draws 32 A at 21 kW, which exceeds typical 20 A bench circuits. A dedicated branch with correctly sized conductors and overcurrent protection must be confirmed before installation; sharing the circuit with other laboratory equipment risks nuisance tripping during simultaneous heating of all 18 places.

Factory Specification Sheet
Mounting
Plan 1000 CFM exhaust capacity at the installation point

The hooded configuration requires 1000 CFM of ventilation to capture digestion fumes. The laboratory exhaust system must be sized for this airflow at the unit's location, and duct routing/static pressure should be checked so the hood face velocity is maintained when the unit is in operation.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose this 18-place hooded combination unit over the 6- or 12-place hooded variants?

Choose the 18-place unit when batch throughput justifies the 32 A / 21 kW service and 1000 CFM exhaust. The 6- and 12-place hooded units fit smaller workloads and lower utility capacity; the open combination units suit labs with existing fume-hood space instead of an integrated hood.

What electrical and ventilation infrastructure must be in place before the unit can be commissioned?

A dedicated 32 A, 21 kW electrical supply and an exhaust system capable of 1000 CFM at the unit's hood. Confirm panel spare capacity, conductor sizing, and duct static pressure before delivery; the catalog excerpt lists no other interface requirements.

What maintenance items and accessories are not included in the base delivery?

The catalog excerpt covers ordering tables only and does not list included accessories, spare parts, or service intervals. Reagent/steam components, glassware, and exhaust connections should be confirmed against the unit's instruction manual and order configuration before purchase.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Classical and Rapid Kjeldahl, Fat and Crude Fiber Apparatus Catalog
Brochure ·9,0 МБ ·28 pp.
View PDF