Pfeiffer Vacuum MVP 003-2 DC
Pfeiffer Vacuum MVP 003-2 DC — Diaphragm / Membrane Vacuum Pumps with Pumping speed 0.25 m³/h, Ultimate pressure 7 mbar.
The MVP 003-2 DC is an oil-free, two-head/two-stage diaphragm vacuum pump with 24 V DC drive, 0.25 m³/h pumping speed and 7.0 hPa ultimate pressure. As the smallest entry of the MVP 003-015 series it has no I/O interface and fixed 3000 min⁻¹ speed, serving compact OEM and benchtop evacuation or backing duty.
Full specifications
| Pumping speed | 0.25 m3/h |
| Ultimate pressure | 7.0 hPa |
| Rotational speed | 3000.0 min-1 |
| Gas ballast mfr | Yes |
| Nominal rotation speed mfr | 3000.0 min-1 |
| Intake pressure max. mfr | 1000.0 hPa |
| Exhaust pressure max. mfr | 1300.0 hPa |
| Weight | 0.7 kg |
| Dimensions (L x W x H) mfr | 116.0 x 116 x 62 mm |
| Power supply | 24.0 V DC |
| Current max. mfr | 0.5 A |
| Inlet flange | G 1/8" |
| Outlet flange | G 1/8" with silencer |
| Noise level | 48.0 dB |
| Operating temperature | 5 - 40 °C |
| Cooling type | Natural convection |
| Protection class mfr | IP30 |
Description
Architecture and position in the series
- Dry-compressing diaphragm pump with 2 diaphragm heads and 2 pumping stages, DC motor at fixed 3000 min⁻¹, natural convection cooling and integrated gas ballast solenoid valve (closed without current).
- At 0.7 kg and 116×116×62 mm it is the most compact variant of the series. Unlike the MVP 003-2 DC Vario (0–5 V speed control and rotation speed output), the MVP 010-3 DC Neo (0.6 m³/h, 1.0 hPa, RS-485, cTUVus) or the mains-powered MVP 015-2 AC, this model has no I/O interface, no certification listing and runs unregulated on 24 V DC.
- Oil- and particle-free operation throughout the series means no oil management in system design and no contamination of pumped media.
Behavior in operation
- Ultimate pressure of 7.0 hPa and 0.25 m³/h throughput place it at the bottom of the series performance range — adequate for small-volume evacuation or backing of smaller turbo pumps, not for high-throughput duty.
- The pump starts only against a maximum 1 bar differential pressure; shut-off valves must be open before or at start. Specified throughput and final pressure are reached only after at least 30 minutes of warm-up with the vacuum connection closed.
- Gas ballast is switched via a solenoid valve; for condensable vapor processes, run 30 min with gas ballast open after the process, and 5–10 min with open vacuum connection before shutdown to purge the suction chamber.
- Surface temperature can exceed 70 °C; sound pressure is 48 dB, suitable for noise-sensitive benchtop/instrument integration without acoustic enclosures.
Installation, connection and servicing
- G 1/8" inlet and outlet (outlet with silencer); use shortest possible lines at full nominal diameter. Exhaust line must be routed downward and without shut-off units to prevent overpressure and condensate return.
- Supply is 24 V DC, max. 0.5 A; the PSU must be doubly isolated per IEC 61010-1/60950-1/62368-1 and the supply fused — the motor has no thermal cut-out and restarts automatically after power failure.
- Wear parts (diaphragms, valves) have a typical life of 8000 operating hours; head covers are retorqued to 0.8 Nm ±0.1 Nm with Torx T10, and per-head spacer disks (three sizes) must be reinstalled identically during diaphragm replacement.
Selection & Integration Q&A
Choose it for simple fixed-speed duty (3000 min⁻¹) with no remote control — it has no I/O interface, unlike the Vario (0–5 V analog speed input, speed feedback) and the DC Neo (RS-485). If you need lower ultimate pressure (1.0 hPa) or higher speed (0.6 m³/h), step up to MVP 010-3 DC Neo.
24 V DC, max. 0.5 A, fused, with double isolation per IEC 61010-1/60950-1/62368-1. The MVP 003-2 DC/Vario tolerate only ±5% supply tolerance (tighter than the Neo models' ±10%), so size and regulate the PSU accordingly; there is no mains switch or thermal cut-out on the motor.
Yes — 0.7 kg, 116×116×62 mm, IP30, 48 dB — but the enclosure must allow natural-convection airflow since cooling is passive and surfaces exceed 70 °C. Ambient 5–40 °C, max. 2000 m altitude, pollution degree 2. After vapor-laden processes run 30 min with gas ballast open; before shutdown run 5–10 min with open vacuum connection.
The factory-sealed pump only — the G 1/8" silencer, gas-ballast valve kits, purge gas nozzle with filter, RS-485/USB cables, OmniControl/handheld unit and service/overhaul kits are accessories. Plan diaphragm/valve overhaul using original kits (~8000 h wear life); the ex-factory closure seal must not be damaged or warranty is void.