Shimadzu Nexera Prep
Shimadzu Nexera Prep — Chromatographs.
The Nexera Prep is a parallel-type double-plunger solvent delivery pump for preparative chromatography, with a maximum flow rate of 150.0 mL/min and a maximum discharge pressure of 44 MPa (10 µL plunger) or 49 MPa (47 µL plunger). It is distinguished within the series by its high-flow capability and optional plunger rinsing mechanism, making it suitable for scaling up purification processes from analytical to preparative scale.
Direct drop-in replacements
Full specifications
| Detector types | UV-VIS SPD-40, PDA SPD-M40, RID-20A, ELSD-LT III, LCMS-2050 |
| Plunger capacity (LC-40D) mfr | 10 μL (LC-40D) |
| Plunger capacity (LC-20AR) mfr | 47 μL (LC-20AR) |
| Plunger capacity (LC-20AP) mfr | 250 μL (LC-20AP) |
| Maximum discharge pressure (LC-40D) mfr | 44 MPa (LC-40D) |
| Maximum discharge pressure (LC-20AR) mfr | 49 MPa (LC-20AR) |
| Maximum discharge pressure (LC-20AP) mfr | 42 MPa (LC-20AP) |
| Flow rate setting range (LC-40D) mfr | 0.01 to 20.00 mL/min (LC-40D) |
| Flow rate setting range (LC-20AR / LC-20AP) mfr | 0.01 to 150.00 mL/min (LC-20AR / LC-20AP) |
| Flow rate (SIL-40C autosampler) mfr | Up to 10 mL/min (SIL-40C autosampler) |
| Size (Column Holder) mfr | W 250 × H 465 × D 400 mm (Column Holder) |
| Size (Column Holder SLIM) mfr | W 110 × H 625 × D 500 mm (Column Holder SLIM) |
| Installable columns (Column Holder) mfr | I.D. 20 to 50 mm columns, 2 pc. (Column Holder) |
| Pore size (C18 / C8 / PFPP columns) mfr | 12 nm (C18 / C8 / PFPP columns) |
| Pore size (HD-C18 column) mfr | 8 nm (HD-C18 column) |
| Column I.D. (conventional scale) (Conventional) mfr | 3 to 8 mm (Conventional) |
| Flowrate (conventional scale) (Conventional) mfr | 0.5 to 3 mL/min (Conventional) |
| Column I.D. (semi-preparative scale) (Semi-preparative) mfr | 10 to 20 mm (Semi-preparative) |
| Flowrate (semi-preparative scale) (Semi-preparative) mfr | 3 to 10 mL/min (Semi-preparative) |
| Column I.D. (preparative scale) (Preparative) mfr | 30 to 50 mm (Preparative) |
| Flowrate (preparative scale) (Preparative) mfr | 10 to 100 mL/min (Preparative) |
| Maximum flowrate capacity (LC-20AR) mfr | 20 mL/min (LC-20AR) |
| Maximum flowrate capacity (LC-40D) mfr | 10 mL/min (LC-40D) |
| Maximum sample injection volume (SIL-40 setup) mfr | 2 mL (SIL-40 setup) |
| Maximum number of LH-40 and/or FRC-40 units (system expansion) mfr | six units (system expansion) |
| Detector signal input channels mfr | up to four channels |
| Size and weight (LC-40D) mfr | W260 × H140 × D500 mm, 10 kg (LC-40D) |
| Size and weight (LC-20AR) mfr | W260 × H140 × D500 mm, 16 kg (LC-20AR) |
| Size and weight (LC-20AP) mfr | W260 × H210 × D500 mm, 19 kg (LC-20AP) |
| Flow rate (LH-40 liquid handler) mfr | Up to 150 mL/min (LH-40 liquid handler) |
| pH range (C18 / C8 columns) mfr | 1 – 12 (C18 / C8 columns) |
| pH range (HD-C18 column) mfr | 1 – 10 (HD-C18 column) |
| pH range (Phenyl / PFPP columns) mfr | 1 – 8 (Phenyl / PFPP columns) |
Description
Construction and Positioning
The Nexera Prep is a solvent delivery module built around a parallel-type double-plunger pump. It is available with three plunger capacities—10 µL, 47 µL, and 250 µL—each corresponding to a different maximum discharge pressure (44 MPa, 49 MPa, and 42 MPa, respectively). The 47 µL variant offers the highest pressure tolerance, while the 250 µL variant is optimized for maximum flow rate. The pump supports constant-pressure solvent delivery, which is essential for maintaining stable flow under varying backpressures typical in preparative separations. The optional plunger rinsing mechanism (syringe or rinsing pump, part number 228-39625-41) helps extend seal life when using buffers or aggressive solvents. ### Performance Characteristics
Flow rate accuracy is specified as no more than ±1% or ±2 µL/min (whichever is greater) under defined conditions, and flow rate precision is ≤0.06% RSD or 0.02 min SD. These specifications ensure reliable delivery for both analytical and preparative workflows. The pump can deliver up to 150.0 mL/min, making it suitable for high-throughput purification. The parallel plunger design minimizes pulsation, and the constant-pressure mode is particularly useful when backpressure fluctuates, such as during gradient elution or column switching. The pump is compact (W260 × H140 × D500 mm for 10 µL and 47 µL variants; W260 × H210 × D500 mm for 250 µL) and weighs between 10 kg and 19 kg depending on configuration. It is intended for integration into a chromatography system, requiring connection to solvent reservoirs, a sample injector, and a column. The plunger rinsing mechanism, if used, requires an additional rinsing pump or syringe.
Integration & Operating Notes
The 10 µL plunger supports up to 44 MPa, the 47 µL up to 49 MPa, and the 250 µL up to 42 MPa. Exceeding these limits can damage the plunger seals and affect flow accuracy. Ensure system backpressure remains below the applicable limit for the installed plunger.
When using buffer solutions, the optional plunger rinsing mechanism (syringe or rinsing pump, part number 228-39625-41) should be employed to prevent salt crystallization and prolong seal life. Without rinsing, maintenance intervals may be shortened.
Flow rate accuracy is specified as ±1% or ±2 µL/min (whichever is greater) under defined conditions, but for flow rates between 0.1 and 5.0 mL/min, the tolerance is ±1% or ±10 µL/min. This must be considered when precise low-flow delivery is required.
The pump supports constant-pressure solvent delivery, which can be advantageous when system backpressure varies, such as during gradient methods or when using columns with changing permeability. This mode helps maintain stable flow.
Selection & Integration Q&A
Choose the 47 µL plunger when you need the highest pressure tolerance (49 MPa) and moderate flow rates. The 10 µL is for lower flow rates with slightly lower pressure (44 MPa), while the 250 µL is for maximum flow (150 mL/min) but lower pressure (42 MPa). Select based on required flow and backpressure.