Shimadzu NOA 7100
Shimadzu NOA 7100 — Gas analyzers.
The Shimadzu NOA 7100 is a transportable, all-in-one NOx/O2 gas analyzer combining atmospheric-pressure chemiluminescence NOx detection with limiting-current zirconia O2 measurement. Built-in pump, filter, and electric cooler enable direct sampling without external pretreatment. It serves continuous emissions monitoring and process gas analysis where portability and multi-range flexibility are required.
Full specifications
| Measurement components mfr | NOx, O2 (NOA-7100) |
| Measurement components (NOA-7100 variant) mfr | NOx, NO, NO2 (*1) (NOA-7100 variant) |
| Measurement range mfr | NOx: 0-25/50/100/250/500/1000/2500/4000 (NOA-7100) |
| Measurement range (NOA-7100 variant) mfr | NOx, NO, NO2: 250/500/1000/2500/4000 ppm ppm (NOA-7100 variant) |
| Gas sample amount mfr | Approx. 2 L/min (NOA-7100) |
| Gas sample amount (NOA-7100 variant) mfr | 100 mL/min (NOA-7100 variant) |
| External drain separator mfr | Fitted (NOA-7100) |
| External drain separator (NOA-7100 variant) mfr | Not fitted (NOA-7100 variant) |
| Small flow rate set mfr | Not fitted (NOA-7100) |
| Small flow rate set (NOA-7100 variant) mfr | Fitted (NOA-7100 variant) |
| Communication set mfr | Not fitted (NOA-7100) |
| Communication set (NOA-7100 variant) mfr | Fitted (NOA-7100 variant) |
| NOx-NO switching solenoid valve set mfr | Not fitted (NOA-7100) |
| NOx-NO switching solenoid valve set (NOA-7100 variant) mfr | Fitted (NOA-7100 variant) |
| Coexisting gas CO (O2 measurement) mfr | Less than 1 vol% (O2 measurement) |
| Coexisting gas CO2 (NOx, NO, NO2 measurement) mfr | 30 vol% or less vol% (NOx, NO, NO2 measurement) |
| Coexisting gas CO2 (O2 measurement) mfr | Less than 30 vol% (O2 measurement) |
| Coexisting gas Hydrocarbon (O2 measurement) mfr | Less than 1 vol% (O2 measurement) |
| Coexisting gas N2O (O2 measurement) mfr | Must not be contained (O2 measurement) |
| Coexisting gas SO3 (NOx, NO, NO2 measurement) mfr | 50 ppm or less ppm (NOx, NO, NO2 measurement) |
| Coexisting gas SO3 (O2 measurement) mfr | Must not be contained (O2 measurement) |
| Coexisting gas Other flammable gases (NOx, NO, NO2 measurement) mfr | Must not be contained (NOx, NO, NO2 measurement) |
Description
Construction and positioning
- All-in-one portable design integrates sample pump, filter, and electric cooler, eliminating external conditioning for field or temporary installations. - Dual-detector architecture: chemiluminescence for NOx (NO/NO2 optional) and zirconia cell for O2, sharing a common sample path. - Compact 420 × 260 × 452 mm, 16 kg housing with LCD showing three parameters simultaneously. ### Measurement behavior
- NOx ranges from 0–25 to 0–4000 ppm, O2 from 0–5 to 0–25 vol%, user-selectable; repeatability within ±0.5% FS, linearity ±2% FS. - Response times configurable: NOx approx. 20/30/60 s, O2 approx. 30/60 s; NOx/NO within 90 s. - Zero and span drift within ±1% and ±1.5% FS per day respectively (25 ppm range: ±1.5% FS/day zero). - Sample gas consumption approx. 2 L/min or 100 mL/min, selectable. ### Installation and operation
- Permissible ambient temperature 5–40°C, humidity ≤90%RH, no direct sunlight or radiant heat. - Power options: 100 V series (410 VA) or 200 V series (700 VA steady 200 VA), 50/60 Hz. - Analog outputs: 0–1 VDC, 3 channels (insulated output, individual channels not insulated); external signal input 0–1 VDC for auxiliary analyzer data. - Data storage to internal memory, USB flash drive (CSV), or via Wi-Fi/LAN to a computer. - Sample inlet accepts pressure -980 to +980 Pa, dust ≤0.1 g/Nm³, ambient temperature.
Integration & Operating Notes
The permissible ambient temperature range of 5–40°C is valid only when the analyzer is shielded from direct sunlight and radiant heat sources; exposure can cause internal overheating and measurement drift, invalidating the specified accuracy.
Power requirements offer 100 V series (410 VA) or 200 V series (700 VA steady 200 VA). Ensure the supply circuit can handle the higher apparent power during startup; undersized wiring may cause voltage drops affecting analyzer stability.
At the NOA-7100 inlet port, sample pressure must remain between -980 and +980 Pa and dust concentration ≤0.1 g/Nm³. Exceeding these limits can clog the built-in filter or damage the pump, leading to measurement errors or downtime.
The 0–1 VDC transmission outputs are insulated as a group but individual channels are not isolated from each other. When connecting to a data logger or PLC, ensure that all channel returns are referenced to the same potential to avoid ground loops and signal errors.
Selection & Integration Q&A
Choose the NOA-7100 when you need a transportable, all-in-one unit with built-in pump, filter, and cooler for field or temporary monitoring. For fixed installations with external conditioning, a rack-mount model may be more suitable.
The NOA-7100 provides 0–1 VDC, 3-channel outputs with a common ground (individual channels not isolated). Ensure the receiving device has a high-impedance input and is referenced to the same ground to prevent signal errors.
Calibration gases, external filters, and replacement parts are not included. Regular zero/span checks are required to meet drift specifications. USB flash drives for data export and LAN cables are also user-supplied.