Shimadzu NSA 308
Shimadzu NSA 308 — Gas analyzers.
The Shimadzu NSA 308 is a stationary gas analyzer using nondispersive infrared absorption for NOx, SO2, CO, and CO2, and a magnetic wind method for O2. It is a floor-standing unit measuring approximately 830 mm wide, 765 mm deep, and 1805 mm high, weighing 300 kg, and operating on 100 VAC at about 1 kVA. It is intended for continuous emission monitoring or process gas analysis where multiple combustion gases must be measured simultaneously.
Full specifications
Description
Construction and positioning
The NSA 308 is a floor-standing gas analyzer with dimensions of approximately 830 mm (W) × 765 mm (D) × 1805 mm (H) and a mass of 300 kg. It combines two measurement principles: nondispersive infrared absorption (NDIR) for NOx, SO2, CO, and CO2, and a magnetic wind method for O2. This dual-technology architecture allows simultaneous measurement of five gas components in a single unit. ### Measurement behavior
The analyzer provides selectable measurement ranges: NOx from 0–200 to 0–2500 ppm, SO2 from 0–100 to 0–2500 ppm, CO from 0–100 to 0–5000 ppm, CO2 from 0–5 to 0–20 vol%, and O2 from 0–10 to 0–25 vol%. No digitized curve points or dynamic response data are available, so behavior under load or in different media cannot be characterized beyond these ranges. The unit requires a 100 VAC power supply with approximately 1 kVA consumption. Given its size and weight, it should be installed on a level floor with adequate clearance for service access.
Integration & Operating Notes
The analyzer weighs 300 kg and stands 1805 mm tall on a footprint of approximately 830 × 765 mm. The installation surface must be level and capable of supporting this load without deflection, and the unit should be secured against tipping, especially in areas with seismic activity or vibration.
The power supply requirement is 100 VAC with approximately 1 kVA consumption. A dedicated circuit with appropriate overcurrent protection should be provided. Voltage fluctuations outside the specified range may affect measurement stability or damage internal electronics.