Shimadzu Quantasep Adept 300
Shimadzu ·Made in Japan · In Active Production

Shimadzu Quantasep Adept 300

Shimadzu Quantasep Adept 300 — Laboratory balances.

The Quantasep Adept 300 is a single-use, automated chromatography system with a 20 mL hold-up volume and two feed/delivery pumps, each supporting four lines (expandable to six). It operates at flow rates of 10–100 mL/min or 30–300 mL/min (±3%) and system pressure up to 50 psi, with integrated UV, conductivity, and pH monitoring. This modification is distinguished by its single-use flow path and absence of inline buffer dilution/mixing, positioning it for process development and small-scale production where cross-contamination risk must be minimized.

Full specifications

Description

Construction and Positioning

Integration & Operating Notes

Process Limits
Do not exceed 50 psi system pressure

Exceeding the 50 psi limit may damage single-use flow path components or cause leaks. The pressure sensor range is 0–100 psi, but the system is rated for 50 psi; ensure pressure alarms are set accordingly.

Factory Specification Sheet
Maintenance
Replace single-use flow path components per batch

The flow path is intended for single use; replace tubing, gaskets, and other wetted parts according to the manufacturer's recommendations or after each batch to prevent cross-contamination and ensure consistent performance.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the Quantasep Adept 300 over a higher-end model in the series?

Choose the Adept 300 when your process uses pre-prepared buffers and does not require inline dilution or mixing. Higher-end models include those features but may have a larger footprint or higher cost. The Adept 300 is ideal for process development and small-scale production with single-use requirements.

What are the requirements for connecting a column to the Adept 300?

The system supports one column with forward, reverse, or bypass flow. Ensure the column's pressure rating exceeds the system's 50 psi limit. Use appropriate fittings compatible with the single-use flow path; the hold-up volume is 20 mL, so consider delay volume in method development.

What maintenance items and accessories are not included in the base delivery?

The base system includes the instrument and standard flow path components. Optional items such as the flow meter, pre-column pH and conductivity sensors, and additional pump lines (up to six per pump) or fractionation lines (up to ten) are not included. Replacement single-use tubing and sensors are consumables and must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation