Shimadzu Tescan Solaris 2
Shimadzu ·Made in Japan · In Active Production

Shimadzu Tescan Solaris 2

Shimadzu Tescan Solaris 2 — Microscopes.

The Tescan Solaris 2 is a field-emission scanning electron microscope (FE-SEM) from Shimadzu, configured as a high-resolution imaging platform. It is distinguished within its class by a fully integrated electron optics column and detection system, optimized for nanoscale surface characterization. Its primary purpose is to provide high-resolution secondary and backscattered electron images for materials science and quality control.

Full specifications

Description

Construction and Architecture

Integration & Operating Notes

Mounting
Requires vibration-isolated floor with low magnetic field

The FE-SEM column is sensitive to mechanical vibration and electromagnetic interference; site survey must confirm floor vibration below 1 µm/s and AC magnetic field below 0.1 µT to avoid resolution loss.

Site Preparation Guide (typical for FE-SEM
Electrical
Stable power with low harmonic distortion is mandatory

The instrument requires a dedicated power line with voltage stability within ±5% and total harmonic distortion below 5% to prevent beam instability and electronics faults.

Installation Manual, Power Requirements
Cooling & Media
Cooling water quality and flow must be maintained

The cooling water circuit must provide flow of at least 2 L/min at 20°C with filtration to 50 µm; failure to maintain this can cause overheating and shutdown.

Facility Requirements Document
Maintenance
Regular filament and aperture cleaning extends source life

The Schottky emitter requires periodic conditioning and the apertures need cleaning every 6 months or after 1000 hours of operation to maintain beam current stability.

Maintenance Schedule, Section 5

Selection & Integration Q&A

What are the requirements for the electrical supply and cooling water for the Solaris 2?

The system requires a dedicated 208-240 V AC single-phase supply with stable voltage and low harmonic distortion, plus cooling water at 20°C with a flow of at least 2 L/min and filtration to 50 µm.

What are the site environment limits for installing the Solaris 2?

Install in a room with temperature stability of ±1°C per hour, relative humidity below 60%, and low vibration and magnetic fields. Avoid locations near heavy machinery or AC power lines.

What maintenance items and accessories are not included with the base instrument?

The base delivery typically excludes consumables such as apertures, filaments, and sample holders. A service contract and spare parts kit are recommended. Refer to the maintenance schedule for periodic replacement intervals.

Dimensional Drawings & Technical Documentation

Dimensional drawings

Tescan Solaris 2 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation