Shimadzu TOC 1000E S
Shimadzu ·Made in Japan · In Active Production

Shimadzu TOC 1000E S

Shimadzu TOC 1000E S — TOC analyzers.

The TOC 1000E S is a UV oxidation-conductivity TOC analyzer with a 0–500 µg/L range and 0.02 µg/L detection limit, packaged in a 270×180×140 mm, 3.0 kg enclosure. It is distinguished within the series by its 5-minute startup and selectable 5/10/15/30-minute measurement cycles, making it suited to continuous low-level organic carbon monitoring in pure-water systems.

Full specifications

Core Performance & Functional Specifications
Detection limit 0.02 μg/L
Measurement range 0 to 500 μg/L
Repeatability (when concentration is no greater than 1 (unit not printed)) mfr ±0.05 (when concentration is no greater than 1 (unit not printed))
Accuracy (500 sucrose) mfr ±5% (500 sucrose)
Linearity (when measuring 250 in the 0-500 μg/L range) mfr R2 = 0.98 or greater (when measuring 250 in the 0-500 μg/L range)
TOC measurement principle mfr UV oxidation-conductivity measurement method
Repeatability *1 mfr 0.05 μg/L
Accuracy *1 mfr 5 %
Linearity *1 mfr 2
Startup time mfr 5 minutes
Measurement cycle mfr 5 minutes
External dimensions mfr 270
Protection from dust and water spray mfr 33
Dimensions, Weight & Mechanical Design
Weight 3.0 kg
TOC-1000e S: approximately 3 kg
Wiring length mfr 10 m
Electrical, Power & Drive Specifications
Required power supply mfr 100 to 240 V

Description

Construction and positioning

Integration & Operating Notes

Mounting
IP33 enclosure suits sheltered, spray-free locations

IP33 protects against solid objects above 2.5 mm and water spray at up to 60° from vertical, but not against immersion or directed jets. Install where the enclosure is not submerged or hosed directly, consistent with the stated protection rating.

Factory Specification Sheet
Process Limits
Measurements above 500 µg/L fall outside the verified range

The specified measurement range is 0–500 µg/L TOC. Samples or process upsets exceeding this upper bound are outside the verified linearity and accuracy figures, so upstream dilution or an alternative range must be considered.

Factory Specification Sheet
Process Limits
Cycle time selection trades update rate for stability

Measurement cycles of 5, 10, 15 or 30 minutes are selectable. Shorter cycles give faster updates; longer cycles may be preferred where the process varies slowly. The appropriate choice depends on the required response time of the monitoring loop.

Factory Specification Sheet

Selection & Integration Q&A

What are the requirements for controllers and cabling connected to this analyzer?

LAN, analog and contact wiring must be 10 m or less. The unit runs on 100–240 V AC, 50/60 Hz at 100 VA. Plan the controller or data logger location within that wiring distance and confirm the AC circuit can supply 100 VA.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c391-e120
Datasheet ·1,9 МБ ·12 pp.
View PDF