Shimadzu TOC VWP
Shimadzu TOC VWP — TOC analyzers.
The TOC VWP is a wet oxidation TOC analyzer using UV illumination and persulfate oxidant with NDIR detection, configured as a PC-controlled model. It measures TC, IC, TOC, and NPOC, and supports up to three calibration curves. It is intended for laboratory water quality monitoring where differentiated carbon measurements are required.
Full specifications
| Vial volume (Shimadzu 40 mL Vial LE) mfr | 40 mL (Shimadzu 40 mL Vial LE) |
| Vial quantity mfr | 72 PCs. (227-34200-01, 227-34200-02) |
| Cap quantity mfr | 72 PCs. (227-34200-01, 227-34200-03) |
| Septum quantity mfr | 72 PCs. (227-34200-01, 227-34200-04) |
| TOC guarantee (pre-cleaned vials, on request) mfr | <10 μgC/L (pre-cleaned vials, on request) |
Description
Construction and positioning
- The TOC VWP is a wet oxidation analyzer that combines UV illumination with persulfate oxidant to oxidize carbon species, followed by NDIR detection of the resulting CO2. - It is a PC-control model, meaning operation and data handling are managed through an external computer rather than a standalone interface. ### Measurement behavior
- The instrument supports four measurement modes: TC (total carbon), IC (inorganic carbon), TOC (total organic carbon), and NPOC (non-purgeable organic carbon). - Up to three calibration curves can be stored, allowing flexibility for different concentration ranges or matrices.
Integration & Operating Notes
Because this is a PC-control model, the analyzer does not have a built-in user interface for routine operation. A compatible computer must be provided and connected, and the control software must be installed and configured before measurements can be taken.
The wet oxidation principle relies on persulfate reagent and UV illumination. Both are consumables that must be replenished or replaced periodically. Their condition directly affects oxidation efficiency and therefore measurement accuracy.
Up to three calibration curves can be stored. To maintain accuracy, these curves should be checked against fresh standards at regular intervals, especially after maintenance or reagent replacement, and updated if necessary.
Selection & Integration Q&A
The base delivery does not include consumables such as persulfate oxidant or UV lamps, nor the external PC required for control. These must be sourced separately. Regular maintenance includes reagent replacement and calibration verification.