Shimadzu Up Series — Laboratory balances.

The Shimadzu Up Series is a top-loading laboratory balance with a 220 g capacity and 0.001 g readability, featuring a 110 mm round pan and external calibration. It delivers stable weighing with approximately 1 s response time and ±3 ppm/°C temperature stability, making it suitable for routine laboratory weighing tasks.

Full specifications

Core Performance & Functional Specifications
Capacity 220 g (UP223X / UP223Y)
UP423X / UP423Y: 420 g · UP623X / UP623Y: 620 g · UP823X / UP823Y: 820 g · UP1023X / UP1023Y: 1020 g · UP2202X / UP2202Y: 2200 g · UP4202X / UP4202Y / UP4201X / UP4201Y: 4200 g · UP6202X / UP6202Y: 6200 g · UP422X / UP422Y: 420 g · UP822X / UP822Y: 820 g · UP8201X / UP8201Y: 8200 g
Linearity ± 0.002 g (UP223X / UP223Y / UP423X / UP423Y / UP623X / UP623Y)
UP823X / UP823Y / UP1023X / UP1023Y: ± 0.003 g
Resolution 0.001 g (small pan model)
Weighing range mfr 100 to 220 g (UP223X / UP223Y)
Weighing range mfr 100 to 420 g (UP423X / UP423Y)
Weighing range mfr 100 to 620 g (UP623X / UP623Y)
Weighing range mfr 400 to 820 g (UP823X / UP823Y)
Weighing range mfr 500 to 1020 g (UP1023X / UP1023Y)
Weighing range mfr 1000 to 2200 g (UP2202X / UP2202Y)
Weighing range mfr 1000 to 4200 g (UP4202X/Y, UP4201X/Y)
Weighing range mfr 1000 to 6200 g (UP6202X / UP6202Y)
Weighing range mfr 100 to 420 g (UP422X / UP422Y)
Weighing range mfr 100 to 820 g (UP822X / UP822Y)
Interfaces, Connections & Control
Minimum display value mfr 0.01 g (UP2202X/Y, UP4202X/Y, UP6202X/Y, UP422X/Y, UP822X/Y)
Minimum display value mfr 0.1 g (UP4201X/Y, UP8201X/Y)

Description

Construction and Positioning

Integration & Operating Notes

Environment
Operate within 10–30 °C for specified sensitivity stability

Sensitivity stability is specified as ±3 ppm/°C only within the 10–30 °C range. Operating outside this range may increase drift and reduce weighing accuracy; ensure the balance is placed away from drafts, heat sources, and direct sunlight.

Factory Specification Sheet

Selection & Integration Q&A

What are the environmental and duty-cycle limits for this balance?

Operate within 10–30 °C to maintain ±3 ppm/°C sensitivity stability. Avoid vibrations, drafts, and rapid temperature changes. No duty-cycle limits are specified; continuous operation is typical for laboratory balances.

Dimensional Drawings & Technical Documentation