Shimadzu Xctal 5000
Shimadzu Xctal 5000 — Spectrometers.
The Shimadzu Xctal 5000 is a phase-contrast X-ray CT system with a flat panel detector and a 90 kV / 800 µA, 50 W generator, configured for non-destructive 3D imaging of samples up to Φ150 mm × H 275 mm (max. 5 kg) on the CT stage. It distinguishes itself within the series by combining a large CT stage with a dedicated orientation analysis unit (Φ100 mm × H 10 mm, 300 g) and fiber-orientation fluoroscopy/CT modes. It is used for internal structure and fiber orientation analysis in materials and components.
Full specifications
| Detector types | Flat panel detector |
| Field of view (dark-field) mfr | up to 100 mm (dark-field imaging) |
| CT stage dimensions mfr | Φ150 mm × H275 mm (CT stage) |
| CT stage maximum load mfr | Max. 5 kg (CT stage) |
| X-Ray Detector mfr | Flat panel detector |
| Available Sample Size mfr | CT Stage |
| Max. Scanning Range mfr | Fluoroscopy |
| CT Scan mfr | 100 |
| CT Data Acquisition Time mfr | 30 minutes |
| Rated Power mfr | Main Unit |
| Grounding mfr | 100 |
| Operating Conditions (when running) mfr | 20 to 30 |
| External Leakage Dose mfr | 1 μSv/h |
| Orientation analysis unit dimensions mfr | Φ100 mm × H10 mm (orientation analysis unit) |
| Orientation analysis unit maximum load mfr | Max. 300 g (orientation analysis unit) |
| Orientation Analysis Unit mfr | 100 mm |
| Fiber Orientation Fluoroscopy mfr | 100 mm |
| Fiber Orientation CT Scan mfr | 85 |
| Weight mfr | 1700 kg |
| Power supply mfr | Single phase 100 to 230 VAC ± 10 % 50/60 VAC, Hz (rated power / control computer) |
| X-Ray Generator Output mfr | 90 kV |
| Control Computer mfr | 100 to 230 VAC |
Description
Construction and positioning
- Phase-contrast X-ray CT system with a flat panel detector and a 90 kV, 800 µA, 50 W X-ray generator. - Two sample positions: a CT stage for large samples (Φ150 mm × H 275 mm, max. 5 kg) and an orientation analysis unit for small samples (Φ100 mm × H 10 mm, max. 300 g). - Maximum scanning range: fluoroscopy 100 mm × 100 mm; CT scan Φ100 × H 100 mm; fiber orientation CT scan Φ85 × H 85 mm. ### Behavior under load and in different modes
- CT data acquisition time is at least 30 minutes, indicating that throughput is governed by scan time rather than by sample size alone. - The system supports fluoroscopy (2D) and CT (3D) modes, plus dedicated fiber-orientation fluoroscopy (Φ100 mm) and fiber-orientation CT (Φ85 × H 85 mm) modes for anisotropic materials. - External leakage dose is 1 µSv/h or less, so the system can be operated in a normal laboratory environment without a shielded room, provided the installation requirements are met. ### Installation, connection and operation
- Main unit requires single-phase 200–230 VAC ±10%, 50/60 Hz, 1.0 kVA; control computer requires single-phase 100–230 VAC ±10%, 50/60 Hz, 1.5 kVA. - Ground resistance must be 100 Ω or less. - Operating ambient temperature 20–30 °C, humidity 40–80% max. (no condensation). - Weight 1700 kg; site floor loading and access route must be verified before installation.
Integration & Operating Notes
The main unit draws 1.0 kVA at 200–230 VAC while the control computer draws 1.5 kVA at 100–230 VAC; they must be supplied from separate circuits matching these voltage ranges, and a common ground of 100 Ω or less is mandatory to avoid noise and safety hazards.
The system weighs 1700 kg; the installation site must be checked for floor load capacity and the transport route must accommodate the unit's dimensions. Vibration and levelness affect CT image quality, so the floor should be rigid and stable.
The CT stage accepts samples up to Φ150 mm × H 275 mm and 5 kg, but the orientation analysis unit is limited to Φ100 mm × H 10 mm and 300 g. Exceeding these limits risks stage damage or invalid scan geometry; the maximum CT scan volume is Φ100 × H 100 mm.
Operating ambient temperature must be kept between 20 and 30 °C with 40–80% humidity and no condensation. Temperature fluctuations during the minimum 30-minute CT acquisition can cause thermal drift and degrade reconstruction accuracy.
Selection & Integration Q&A
Main unit: single-phase 200–230 VAC ±10%, 50/60 Hz, 1.0 kVA. Control computer: single-phase 100–230 VAC ±10%, 50/60 Hz, 1.5 kVA. Ground resistance ≤100 Ω. Provide separate circuits and a common ground.
Operate at 20–30 °C and 40–80% RH without condensation. CT acquisition takes at least 30 minutes, so temperature stability is critical. External leakage dose is ≤1 µSv/h, allowing use in a normal lab without extra shielding.
The supplied facts do not list maintenance intervals, consumables, cables, or accessories included in the base delivery. Confirm service schedule and required accessories (e.g., sample holders, calibration standards) with Shimadzu before ordering.